Background: Complete blood cell count (CBC) test, peripheral blood smears (PBS), and automatic hematology analyzers are used to screen pa- tients with high-risk hematology malignancies. In this study, the workflow of urgent and routine samples and methods to improve the detection rate of blasts were evaluated. The purpose of this study was to validate the hematology rapid reporting system (HRS) for CBC and establish a strategy for the effective use of an automated hematology analyzer in clinical laboratories. Methods: The flag performances of UniCel DxH 800 (Beckman Coulter, USA) and XE-2100 (Sysmex, Japan) systems used for blood tests were analyzed for urgent and routine samples. The results were compared with PBS results. Each test method’s sensitivity and specificity were deter- mined to identify the most efficient blast detection method. In addition, receiver operating characteristic (ROC) curves of various parameters such as CBC and age were analyzed. Results: The sensitivity and specificity of HRS were 66.80% and 99.85%, respectively, with urgent flags (four flags) and 92.95% and 98.49%, respectively, with routine flags (15 flags). The sensitivity of HRS for routine samples was significantly different from that for urgent samples (66.80% vs. 92.95%; < 0.0001), but no significant difference was observed in the specificity for HRS between routine and urgent samples (99.85% vs. 98.49%; = 1.00). Conclusions: HRS using flags and blast history can be effectively used in clinical laboratories. HRS can be used as an effective reporting work- flow for differential CBC in clinical laboratories.

Abstract

Background: Complete blood cell count (CBC) test, peripheral blood smears (PBS), and automatic hematology analyzers are used to screen pa- tients with high-risk hematology malignancies. In this study, the workflow of urgent and routine samples and methods to improve the detection rate of blasts were evaluated. The purpose of this study was to validate the hematology rapid reporting system (HRS) for CBC and establish a strategy for the effective use of an automated hematology analyzer in clinical laboratories. Methods: The flag performances of UniCel DxH 800 (Beckman Coulter, USA) and XE-2100 (Sysmex, Japan) systems used for blood tests were analyzed for urgent and routine samples. The results were compared with PBS results. Each test method’s sensitivity and specificity were deter- mined to identify the most efficient blast detection method. In addition, receiver operating characteristic (ROC) curves of various parameters such as CBC and age were analyzed. Results: The sensitivity and specificity of HRS were 66.80% and 99.85%, respectively, with urgent flags (four flags) and 92.95% and 98.49%, respectively, with routine flags (15 flags). The sensitivity of HRS for routine samples was significantly different from that for urgent samples (66.80% vs. 92.95%; < 0.0001), but no significant difference was observed in the specificity for HRS between routine and urgent samples (99.85% vs. 98.49%; = 1.00). Conclusions: HRS using flags and blast history can be effectively used in clinical laboratories. HRS can be used as an effective reporting work- flow for differential CBC in clinical laboratories.

참고문헌 (9)

  1. 조윤아, "Mindray BC-6800 자동혈구분석기의 유용성 평가", Laboratory Medicine Online, 3(3), pp.131~137 (2013), 대한진단검사의학회
  2. Briggs CJ, "Improved flagging rates on the Sysmex XE-5000 compared with the XE-2100 reduce the number of manual film reviews and increase laboratory productivity", Am J Clin Pathol, 136, pp.309~316 (2011)
  3. Barnes PW, "The international consensus group for hematology review : suggested criteria for action following automated CBC and WBC differential analysis", Lab Hematol, 11, pp.83~90 (2005)
  4. Bain BJ, "Diagnosis from the blood smear", N Engl J Med, 353, pp.498~507 (2005)
  5. Ohsaka A, "Artificial intelligence(AI)and hematological diseases : establishment of a peripheral blood convolutional neural network(CNN)-based digital morphology analysis system", Rinsho Ketsueki, 61, pp.564~569 (2020)
  6. Craig FE, "The utility of peripheral blood smear review for identifying specimens for flow cytometric immunophenotyping", Int J Lab Hematol, 39(1S), pp.41~46 (2017)
  7. Rabizadeh E, "Acute leukemia detection rate by automated blood count parameters and peripheral smear review", Int J Lab Hematol, 37, pp.44~49 (2015)
  8. Houwen B, "The differential cell count", Lab Hematol, 7, pp.89~100 (2001)
  9. Novis DA, "Louis M, Ben-Ezra J. Laboratory productivity and the rate of manual peripheral blood smear review : a College of American Pathologists Q-Probes study of 95141 complete blood count determinations performed in 263 institutions", Arch Pathol Lab Med, 130, pp.596~601 (2006)

관련 논문AI추천

Differential Blast Counts Obtained by Automated Blood Cell Analyzers김용구, 박연준, 오은지, 임지향, 정승원, 채효진, 한경자대한진단검사의학회 · Annals of Laboratory Medicine2010 · 제30권 제6호Performance Evaluation of the Mindray BC-6200 Hematology Analyzer; Comparison with Sysmex XE-2100 and Manual Microscopy김한아, 박미경, 이창훈, 이태환, 허미나대한진단검사의학회 · Laboratory Medicine Online2022 · 제12권 제4호Flow Cytometric White Blood Cell Differential Using CytoDiff is Excellent for Counting Blasts강지민, 김명신, 김용구, 박연준, 오은지, 한경자대한진단검사의학회 · Annals of Laboratory Medicine2015 · 제35권 제1호Reliable, Accurate Determination of the Leukocyte Differential of Leukopenic Samples by Using Hematoflow Method고광상, 김수화, 김용구, 박연준, 박종문, 심양보, 임지향, 조용준, 한경자대한진단검사의학회 · Annals of Laboratory Medicine2011 · 제31권 제3호Current Status and Proposal of a Guideline for Manual Slide Review of Automated Complete Blood Cell Count and White Blood Cell Dfferential권계철, 김영재, 김희진, 박효순, 신상용, 우희연, 이영경, 장윤환, 조성란, 채석래, 최종락, 한경자대한진단검사의학회 · Annals of Laboratory Medicine2010 · 제30권 제6호Evaluation of a Digital Image Analyzer UIMD PBIA for Determining White Blood Cell Differential Count김명신, 김용구, 이종미, 한경자대한진단검사의학회 · Laboratory Medicine Online2023 · 제13권 제1호

이 간행물의 다른 논문 (12건)