Evaluation of a Digital Image Analyzer UIMD PBIA for Determining White Blood Cell Differential Count

2

조회

0

다운로드

1

인용

0

참고문헌

키워드

초록

Background: The UIMD PBIA (ANI CO., Suwon, Korea) is a newly developed automated digital image analyzer using innovative algorithms for the analysis of peripheral blood smears. We evaluated the accuracy and throughput of UIMD PBIA for the classification of white blood cells (WBCs). Methods: A total of 29,605 cells in 242 clinical samples (192 samples with abnormal findings and 50 normal samples) were used to evaluate the classification accuracy and throughput of the UIMD PBIA. In addition, the total processing time for WBC classification by UIMD PBIA was measured to calculate the throughput. Results: UIMD PBIA revealed outstanding performance for the identification of normal samples (99.0% accuracy) and five-part differentials (neutrophil, lymphocyte, monocyte, eosinophil, basophil, 99.2% accuracy). Misclassifications frequently occurred for immature granulocytes (83.6-93.9% accuracy), blasts (93.5% accuracy), and abnormal lymphocytes (81% accuracy). The pathogenic cells were likely to be misclassified into other classes of the same lineage. The average throughput was approximately 42 slides per hour. In cases with pancytopenia, the throughput was approximately 29 slides per hour. Conclusions: The UIMD PBIA offers the most accurate results for WBC classification and the highest throughput, thereby reducing the technical workload, especially in cases with normal findings and pancytopenia. Accordingly, this study revealed the feasibility of using a digital switch in CBC analysis.

Abstract

Background: The UIMD PBIA (ANI CO., Suwon, Korea) is a newly developed automated digital image analyzer using innovative algorithms for the analysis of peripheral blood smears. We evaluated the accuracy and throughput of UIMD PBIA for the classification of white blood cells (WBCs). Methods: A total of 29,605 cells in 242 clinical samples (192 samples with abnormal findings and 50 normal samples) were used to evaluate the classification accuracy and throughput of the UIMD PBIA. In addition, the total processing time for WBC classification by UIMD PBIA was measured to calculate the throughput. Results: UIMD PBIA revealed outstanding performance for the identification of normal samples (99.0% accuracy) and five-part differentials (neutrophil, lymphocyte, monocyte, eosinophil, basophil, 99.2% accuracy). Misclassifications frequently occurred for immature granulocytes (83.6-93.9% accuracy), blasts (93.5% accuracy), and abnormal lymphocytes (81% accuracy). The pathogenic cells were likely to be misclassified into other classes of the same lineage. The average throughput was approximately 42 slides per hour. In cases with pancytopenia, the throughput was approximately 29 slides per hour. Conclusions: The UIMD PBIA offers the most accurate results for WBC classification and the highest throughput, thereby reducing the technical workload, especially in cases with normal findings and pancytopenia. Accordingly, this study revealed the feasibility of using a digital switch in CBC analysis.

참고문헌 (16)

  1. Clinical and Laboratory Standards Institute, "Reference leukocytes (WBC) differential count (proportional) and evaluation of instrumental methods; Approved standard" (2007), Clinical and Laboratory Standards Institute
  2. Da Costa L, "Digital image analysis of blood cells", Clin Lab Med, 35, pp.105~122 (2015)
  3. Tatsumi N, "Automated image processing : past, present, and future of blood cell morphology identification", Clin Lab Med, 22, pp.299~315 (2002)
  4. Kratz A, "Performance evaluation of the CellaVision DM96 system : WBC differentials by automated digital image analysis supported by an artificial neural network", Am J Clin Pathol, 124, pp.770~781 (2005)
  5. Ceelie H, "Examination of peripheral blood films using automated microscopy; evaluation of Diffmaster Octavia and Cellavision DM96", J Clin Pathol, 60, pp.72~79 (2007)
  6. Briggs C, "Can automated blood film analysis replace the manual differential? An evaluation of the CellaVision DM96 automated image analysis system", Int J Lab Hematol, 31, pp.48~60 (2009)
  7. Tabe Y, "Performance evaluation of the digital cell imaging analyzer DI-60 integrated into the fully automated Sysmex XN hematology analyzer system", Clin Chem Lab Med, 53, pp.281~289 (2015)
  8. Kim HN, "Performance of automated digital cell imaging analyzer Sysmex DI-60", Clin Chem Lab Med, 56, pp.94~102 (2017)
  9. Kratz A, "Digital morphology analyzers in hematology : ICSH review and recommendations", Int J Lab Hematol, 41, pp.437~447 (2019)
  10. Fu X, "Morphogo : an automatic bone marrow cell classification system on digital images analyzed by artificial intelligence", Acta Cytol, 64, pp.588~596 (2020)
  11. Palmer L, "ICSH recommendations for the standardization of nomenclature and grading of peripheral blood cell morphological features", Int J Lab Hematol, 37, pp.287~303 (2015)
  12. Alférez E, "Morphological features using digital image processing in lymphoid neoplasias", Int J Lab Hematol, 33(suppl 1), pp.53~ (2011)
  13. Alférez S, "Automatic recognition of atypical lymphoid cells from peripheral blood by digital image analysis", Am J Clin Pathol, 143, pp.168~176 (2015)
  14. Alférez S, "Automatic classification of atypical lymphoid B cells using digital blood image processing", Int J Lab Hematol, 36, pp.472~480 (2014)
  15. Bengtsson HI, "Digital morphology analyzers in hematology : Comments on the ICSH review and recommendations", Int J Lab Hematol, 42, pp.e213~e215 (2020)
  16. 김아현, "White blood cell differential counts in severely leukopenic samples: a comparative analysis of different solutions available in modern laboratory hematology", Blood Research, 49(2), pp.120~126 (2014), 대한혈액학회

관련 논문AI추천

백혈구감소증 검체에서 디지털 세포형태 분석기 UIMD PBIA의 임상적 유용성 평가유은경, 이자영, Kim Hyun Woo대한진단검사의학회 · Laboratory Medicine Online2024 · 제14권 제4호Digital Morphology Analyzer Sysmex DI-60 vs. Manual Counting for White Blood Cell Differentials in Leukopenic Samples: A Comparative Assessment of Risk and Turnaround TimeHur Mina, Kim Hanah, Kim Hyeong Nyeon, Lee Gun Hyuk, Nam Minjeong, Park Mikyoung, Yoon Sumi대한진단검사의학회 · Annals of Laboratory Medicine2022 · 제42권 제4호Mindray MC-80 자동디지털세포형태분석기와 수동 검사법 비교: 말초혈액 도말의 비정상 세포 분석이종호, Yi Tae Hyeon대한진단검사의학회 · Laboratory Medicine Online2024 · 제14권 제3호Reliable, Accurate Determination of the Leukocyte Differential of Leukopenic Samples by Using Hematoflow Method고광상, 김수화, 김용구, 박연준, 박종문, 심양보, 임지향, 조용준, 한경자대한진단검사의학회 · Annals of Laboratory Medicine2011 · 제31권 제3호Validation of Hematology Rapid Reporting System for Complete Blood Cell Count with Differential김지용, 박지영, 윤수영, 이정민대한진단검사의학회 · Laboratory Medicine Online2022 · 제12권 제1호Performance Evaluation of the Mindray BC-6200 Hematology Analyzer; Comparison with Sysmex XE-2100 and Manual Microscopy김한아, 박미경, 이창훈, 이태환, 허미나대한진단검사의학회 · Laboratory Medicine Online2022 · 제12권 제4호

이 간행물의 다른 논문 (9건)