Pattern of change in force and moment according to amounts of labial tipping and material types of clear aligner: In vitro experiment with a six-axis mechanical sensor apparatus

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류치의학

초록

Objective: To investigate the pattern of change in force and moment according to the amounts of labial tipping and the material types of clear aligners (CA). Methods: A 3D-printed maxillary typodont system, including crown, root, and periodontal ligament space, was constructed. Six groups were established based on the amount of labial tipping of #12 (0.3 mm and 0.6 mm), and the material types (Zendura [thermoplastic, one-layer, 0.75 mm], Trioclear [thermoplastic, three-layer, 0.5 mm], and Graphy TA-28 [direct printing, 0.5 mm]) (n = 6/group). A six-axis mechanical sensor apparatus was connected to the roots. Force and moment were measured after seating the CA in a 37°C water bath. Results: When comparing 0.3 mm and 0.6 mm tipping, #12 showed an increase in labial inclination moment (Lab-inclinationmoment) in Zendura (P < 0.05); a decrease in labial force and an increase in Labinclination- moment in Trioclear (P < 0.05, P < 0.01); while a lingual force and Labinclination moment without significant change in Graphy (all P > 0.05). For 0.3 mm tipping, the amounts of labio-lingual forces and Lab-inclination moment did not differ among the three materials (all P > 0.05). However, at 0.6 mm tipping, Zendura exerted a greater labial force than Trioclear, while Graphy produced a lingual force (P < 0.01). Zendura and Trioclear generated a stronger Lab-inclinationmoment compared to Graphy (P < 0.01). Despite displaying unexpected force and moment directions, Graphy consistently exhibited the lowest absolute values among the groups. Conclusions: Clinicians should select the appropriate aligner material based on the specific biomechanical requirements to prevent unexpected forces and moments.

Abstract

Objective: To investigate the pattern of change in force and moment according to the amounts of labial tipping and the material types of clear aligners (CA). Methods: A 3D-printed maxillary typodont system, including crown, root, and periodontal ligament space, was constructed. Six groups were established based on the amount of labial tipping of #12 (0.3 mm and 0.6 mm), and the material types (Zendura [thermoplastic, one-layer, 0.75 mm], Trioclear [thermoplastic, three-layer, 0.5 mm], and Graphy TA-28 [direct printing, 0.5 mm]) (n = 6/group). A six-axis mechanical sensor apparatus was connected to the roots. Force and moment were measured after seating the CA in a 37°C water bath. Results: When comparing 0.3 mm and 0.6 mm tipping, #12 showed an increase in labial inclination moment (Lab-inclinationmoment) in Zendura (P < 0.05); a decrease in labial force and an increase in Labinclination- moment in Trioclear (P < 0.05, P < 0.01); while a lingual force and Labinclination moment without significant change in Graphy (all P > 0.05). For 0.3 mm tipping, the amounts of labio-lingual forces and Lab-inclination moment did not differ among the three materials (all P > 0.05). However, at 0.6 mm tipping, Zendura exerted a greater labial force than Trioclear, while Graphy produced a lingual force (P < 0.01). Zendura and Trioclear generated a stronger Lab-inclinationmoment compared to Graphy (P < 0.01). Despite displaying unexpected force and moment directions, Graphy consistently exhibited the lowest absolute values among the groups. Conclusions: Clinicians should select the appropriate aligner material based on the specific biomechanical requirements to prevent unexpected forces and moments.

관련 논문AI추천

Force and moment analysis of clear aligners: Impact of material properties and design on premolar rotationDong-Woo Kim, Hyun-Jun Lee, Ki Beom Kim, Seong-Sik Kim, Soo-Byung Park, Sung-Hun Kim, Yong-Il Kim, Youn-Kyung Choi대한치과교정학회 · 대한치과교정학회지2025 · 제55권 제3호Evaluation of different designs of 3D printed clear aligners on mandibular premolar extrusion using force/moment measurement devices and digital image correlation methodJong-Chan Baik, Ki Beom Kim, Seong-Sik Kim, Soo-Byung Park, Sung-Hun Kim, Yonghun Cho, Yong-Il Kim, Youn-Kyung Choi, Yunju Baek대한치과교정학회 · 대한치과교정학회지2024 · 제54권 제6호Force changes associated with differential activation of en-masse retraction and/or intrusion with clear alignersShuo Li, Wei Hu, Ye Zhu대한치과교정학회 · 대한치과교정학회지2021 · 제51권 제1호Changes in force associated with the amount of aligner activation and lingual bodily movement of the maxillary central incisorChaochao Ren, Hongmei Wang, Pai Zhao, Xiaowei Li, Yu-Xing Bai, Zheyao Wang대한치과교정학회 · 대한치과교정학회지2016 · 제46권 제2호In vitro analysis of mandibular arch response to canine distalization with aligners at varying initial angulationsAn Le, Fan Yongjie, Guo Baoxiong, Liu Gebing, Lv Xin, Wang Bingle, Yang Sensen대한치과교정학회 · 대한치과교정학회지2025 · 제55권 제4호Evaluation of the effect of attachments on torque control of palatally positioned maxillary lateral teeth with clear aligners: Finite element analysisIrmak Ocak, Nurver Karsli, Ömür Polat Özsoy, Sevil Gökcek대한치과교정학회 · 대한치과교정학회지2025 · 제55권 제1호

이 간행물의 다른 논문 (8건)