Comparison of the standards for absorbed dose to water of the IAEA and the KRISS, Korea in accelerator photon beams

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류원자력공학

초록

A bilateral comparison was conducted between the International Atomic Energy Agency (IAEA) and the Korea Research Institute of Standards and Science (KRISS) to measure the absorbed dose to water in accelerator photon beams. KRISS served as a linking laboratory to compare the IAEA standard with the key comparison reference value (KCRV) of the BIPM.RI(I)–K6 program, in which KRISS participated in 2017. Two ionization chambers from the IAEA were used as transfer instruments for the comparison. Both laboratories measured the calibration coefficients of these instruments and calculated the ratios. The ratio of the KRISS standard to the KCRV was applied to obtain the degree of equivalence of the IAEA, along with its uncertainty. The largest deviation of the IAEA measurement from the KCRV was 3.4 mGy/Gy, significantly smaller than the expanded uncertainty of 10.7 mGy/Gy (k = 2, 95% level of confidence). This study demonstrates the equivalence of IAEA’s measurement standard for accelerator photon beams to other primary standard dosimetry laboratories. It provides evidence for the satisfactory operation of IAEA’s quality management system and enhances the international credibility of the IAEA SSDL network, particularly in high-energy accelerator photon beams from linear accelerators.

Abstract

A bilateral comparison was conducted between the International Atomic Energy Agency (IAEA) and the Korea Research Institute of Standards and Science (KRISS) to measure the absorbed dose to water in accelerator photon beams. KRISS served as a linking laboratory to compare the IAEA standard with the key comparison reference value (KCRV) of the BIPM.RI(I)–K6 program, in which KRISS participated in 2017. Two ionization chambers from the IAEA were used as transfer instruments for the comparison. Both laboratories measured the calibration coefficients of these instruments and calculated the ratios. The ratio of the KRISS standard to the KCRV was applied to obtain the degree of equivalence of the IAEA, along with its uncertainty. The largest deviation of the IAEA measurement from the KCRV was 3.4 mGy/Gy, significantly smaller than the expanded uncertainty of 10.7 mGy/Gy (k = 2, 95% level of confidence). This study demonstrates the equivalence of IAEA’s measurement standard for accelerator photon beams to other primary standard dosimetry laboratories. It provides evidence for the satisfactory operation of IAEA’s quality management system and enhances the international credibility of the IAEA SSDL network, particularly in high-energy accelerator photon beams from linear accelerators.

참고문헌 (8)

  1. "Key Comparison Database (KCDB)"[URL]
  2. C. Kessler, "Comparison BIPM.RI(I)-K6 of the standards forabsorbed dose to water of the KRISS, Korea and the BIPM in accelerator photon beams", Metrologia, 57, pp.06001~ (2020)
  3. Updated report on the evaluation of degree of equivalence in regional dosimetry comparisons, "D. T. Burns", ~, pp.CCRI(I) (2017), 17-09
  4. S. R. Domen, "A heat-loss-compensated calorimeter: theory, design, and performance", J. Res. NatL. Stand., Sect A, 78A, pp.595~610 (1974)
  5. I. J. Kim, "Building a graphite calorimetry system for the dosimetry of therapeutic X-ray beams", Nucl. Eng. Technol, 49, pp.810~816 (2017)
  6. C. M. Sutton, "Analysis and linking of international measurement comparisons", Metrologia, 41, pp.272~277 (2004)
  7. Absorbed Dose Determination in External Beam Radiotherapy: an International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water, "IAEA", ~, pp.vol. 398, TRS (2000)
  8. D. T. Burns, "Ion recombination corrections for the NACP parallel-plate chamber in a pulsed electron beam", Phys. Med. Biol, 43, pp.2033~2045 (1998)

관련 논문AI추천

국제표준규격의 137Cs 감마선장에서 전리함선량계 종류에 따른 공기커마율 측정 및 평가결과의 유효성 검증안현, 강영록, 계용욱, 김효진, 목성윤, 백정화, 이지은, 이창근, 장희진, 조월순(사)한국방사선산업학회 · 방사선산업학회지2025 · 제19권 제3호Uncertainty evaluation of absorbed dose measurement based on temperature rise using a water calorimeter for 60Co gamma rays김윤호, 김인중, 박병건, 박종인, 성영민, 이철영, 정성문, Solomon Rukundo한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제11호Beam Energy Determination for a 10 kW LINAC through Monte Carlo Calculation권호제(사)한국방사선산업학회 · 방사선산업학회지2020 · 제14권 제4호기준 방사선장 구축을 위한 방사선 조사장치의 특성 및선질 평가에 대한 연구권다영, 노성진, 민제호, 한기택(사)한국방사선산업학회 · 방사선산업학회지2022 · 제16권 제4호Performance Evaluation of Stealth Chamber as a Novel Reference Chamber for Measuring Percentage Depth Dose and Profile of VitalBeam Linear Accelerator강상원, 강성희, 김경현, 김연래, 서태석, 신영주, 이정우, 정재용, 정진범대한방사선과학회 · Journal of Radiological Science and Technology2018 · 제41권 제3호Assessment Results to Determine Proficiency in Irradiation Procedures이윤종(사)한국방사선산업학회 · 방사선산업학회지2024 · 제18권 제4호

이 간행물의 다른 논문 (49건)