본문 바로가기
KERD
논문KCI 등재

Experiment of proof-of-principle on prompt gamma-positron emission tomography (PG-PET) system for in-vivo dose distribution verification in proton therapy

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류원자력공학

초록

In our previous study, we proposed an integrated PG-PET-based imaging method to increase the prediction accuracy for patient dose distributions. The purpose of the present study is to experimentally validate the feasibility of the PG-PET system. Based on the detector geometry optimized in the previous study, we constructed a dual-head PG-PET system consisting of a 16  16 GAGG scintillator and KETEK SiPM arrays, BaSO4 reflectors, and an 8  8 parallel-hole tungsten collimator. The performance of this system as equipped with a proof of principle, we measured the PG and positron emission (PE) distributions from a 3  6  10 cm3 PMMA phantom for a 45 MeV proton beam. The measured depth was about 17 mm and the expected depth was 16 mm in the computation simulation under the same conditions as the measurements. In the comparison result, we can find a 1 mm difference between computation simulation and measurement. In this study, our results show the feasibility of the PG-PET system for in-vivo range verification. However, further study should be followed with the consideration of the typical measurement conditions in the clinic application.

Abstract

In our previous study, we proposed an integrated PG-PET-based imaging method to increase the prediction accuracy for patient dose distributions. The purpose of the present study is to experimentally validate the feasibility of the PG-PET system. Based on the detector geometry optimized in the previous study, we constructed a dual-head PG-PET system consisting of a 16  16 GAGG scintillator and KETEK SiPM arrays, BaSO4 reflectors, and an 8  8 parallel-hole tungsten collimator. The performance of this system as equipped with a proof of principle, we measured the PG and positron emission (PE) distributions from a 3  6  10 cm3 PMMA phantom for a 45 MeV proton beam. The measured depth was about 17 mm and the expected depth was 16 mm in the computation simulation under the same conditions as the measurements. In the comparison result, we can find a 1 mm difference between computation simulation and measurement. In this study, our results show the feasibility of the PG-PET system for in-vivo range verification. However, further study should be followed with the consideration of the typical measurement conditions in the clinic application.

참고문헌 (32)

  1. Choi, "Development of a Geant4-based Independent Patient Dose Validation System with an", J Applied Clin Med Phys (2019)
  2. W P Levin, "Proton beam therapy", British Journal of Cancer, 93(8), pp.849~854 (2005), Springer Science and Business Media LLC[DOI]
  3. Raymond Miralbell, "Potential reduction of the incidence of radiation-induced second cancers by using proton beams in the treatment of pediatric tumors", International Journal of Radiation Oncology*Biology*Physics, 54(3), pp.824~829 (2002), Elsevier BV[DOI]
  4. Karl Otto, "Volumetric modulated arc therapy: IMRT in a single gantry arc", Medical Physics, 35(1), pp.310~317 (2007), Wiley[DOI]
  5. Robert R. Wilson, "Radiological Use of Fast Protons", Radiology, 47(5), pp.487~491 (1946), Radiological Society of North America (RSNA)[DOI]
  6. Antje-Christin Knopf, "In vivo proton range verification: a review", Physics in Medicine and Biology, 58(15), pp.R131~R160 (2013), IOP Publishing[DOI]
  7. F Albertini, "The influence of the optimization starting conditions on the robustness of intensity-modulated proton therapy plans", Physics in Medicine and Biology, 55(10), pp.2863~2878 (2010), IOP Publishing[DOI]
  8. Hsin-Hon Lin, "A comparison of two prompt gamma imaging techniques with collimator-based cameras for range verification in proton therapy", Radiation Physics and Chemistry, 137, pp.144~150 (2017), Elsevier BV[DOI]
  9. Christian Golnik, "Range assessment in particle therapy based on prompt<i>γ<", i>-ray timing measurements, 2014, pp.59(18), 5399~5422[DOI]
  10. J. Krimmer, "Prompt-gamma monitoring in hadrontherapy: A review", Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 878, pp.58~73 (2018), Elsevier BV[DOI]
  11. Patricia Cambraia Lopes, "Optimization of the Signal-to-Background Ratio in Prompt Gamma Imaging Using Energy and Shifting Time-of-Flight Discrimination: Experiments With a Scanning Parallel-Slit Collimator", IEEE Transactions on Radiation and Plasma Medical Sciences, 2(5), pp.510~519 (2018), Institute of Electrical and Electronics Engineers (IEEE)[DOI]
  12. Chul-Hee Min, "Prompt gamma measurements for locating the dose falloff region in the proton therapy", Applied Physics Letters, 89(18), pp.2~5 (2006), AIP Publishing[DOI]
  13. Christian Richter, "First clinical application of a prompt gamma based in vivo proton range verification system", Radiotherapy and Oncology, 118(2), pp.232~237 (2016), Elsevier BV[DOI]
  14. Joost M Verburg, "Energy- and time-resolved detection of prompt gamma-rays for proton range verification", Physics in Medicine and Biology, 58(20), pp.L37~L49 (2013), IOP Publishing[DOI]
  15. Yunhe Xie, "Prompt Gamma Imaging for In Vivo Range Verification of Pencil Beam Scanning Proton Therapy", International Journal of Radiation Oncology*Biology*Physics, 99(1), pp.210~218 (2017), Elsevier BV[DOI]
  16. W Enghardt, "Charged hadron tumour therapy monitoring by means of PET", Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 525(1-2), pp.284~288 (2004), Elsevier BV[DOI]
  17. Veronica Ferrero, "Online proton therapy monitoring: clinical test of a Silicon-photodetector-based in-beam PET", Scientific Reports, 8(1), pp.1~8 (2018), Springer Science and Business Media LLC[DOI]
  18. P Cambraia Lopes, "First in situ TOF-PET study using digital photon counters for proton range verification", Physics in Medicine and Biology, 61(16), pp.6203~6230 (2016), IOP Publishing[DOI]
  19. S.F. Kry, "New small-intestine modeling method for surface-based computational human phantoms", ArXiv, 36, pp.e391~e429 (2017)[DOI]
  20. M Moteabbed, "Monte Carlo patient study on the comparison of prompt gamma and PET imaging for range verification in proton therapy", Physics in Medicine and Biology, 56(4), pp.1063~1082 (2011), IOP Publishing[DOI]
  21. Teiji Nishio, "The Development and Clinical Use of a Beam ON-LINE PET System Mounted on a Rotating Gantry Port in Proton Therapy", International Journal of Radiation Oncology*Biology*Physics, 76(1), pp.277~286 (2010), Elsevier BV[DOI]
  22. Hideaki Tashima, "A single-ring OpenPET enabling PET imaging during radiotherapy", Physics in Medicine and Biology, 57(14), pp.4705~4718 (2012), IOP Publishing[DOI]
  23. Hideaki Tashima, "Development of a small single-ring OpenPET prototype with a novel transformable architecture", Physics in Medicine and Biology, 61(4), pp.1795~1809 (2016), IOP Publishing[DOI]
  24. Xuping Zhu, "Proton Therapy Verification with PET Imaging", Theranostics, 3(10), pp.731~740 (2013), Ivyspring International Publisher[DOI]
  25. Hyun Joon Choi, "Development of integrated prompt gamma imaging and positron emission tomography system for in vivo 3-D dose verification: a Monte Carlo study", Phys. Med. Biol, 65(10), pp.105005~ (2020), IOP Publishing[DOI]
  26. 김찬호, "Digital n-γ Pulse Shape Discrimination in Organic Scintillators with a High-Speed Digitizer", 방사선방어학회지, 44(2), pp.53~63 (2019), 테스트학회[DOI]
  27. Ricardo Bugalho, "Experimental results with TOFPET2 ASIC for time-of-flight applications", Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 912, pp.195~198 (2018), Elsevier BV[DOI]
  28. Thibaut Merlin, "CASToR: a generic data organization and processing code framework for multi-modal and multi-dimensional tomographic reconstruction", Physics in Medicine &amp; Biology, 63(18), pp.185005~ (2018), IOP Publishing[DOI]
  29. S. Agostinelli, "Geant4—a simulation toolkit", Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 506(3), pp.250~303 (2003), Elsevier BV[DOI]
  30. M. C. Battaglia, "Dosimetric response of radiochromic films to protons of low energies in the Bragg peak region", Physical Review Accelerators and Beams, 19(6), pp.1~7 (2016), American Physical Society (APS)[DOI]
  31. J. Sorriaux, "Evaluation of Gafchromic® EBT3 films characteristics in therapy photon, electron and proton beams", Physica Medica, 29(6), pp.599~606 (2013), Elsevier BV[DOI]
  32. J. Sorriaux, "Evaluation of Gafchromic® EBT3 films characteristics in therapy photon, electron and proton beams", Physica Medica, 29(6), pp.599~606 (2013), Elsevier BV[DOI]

관련 논문AI추천

Upgrade of gamma electron vertex imaging system for high-performance range verifi cation in pencil beam scanning proton therapy구영모, 김성훈, 김찬형, 정재린, 정종휘, 조광현, 조성구한국원자력학회 · Nuclear Engineering and Technology2022 · 제54권 제3호New algorithm to estimate proton beam range for multi-slit prompt-gamma camera구영모, 김찬형, 정재린한국원자력학회 · Nuclear Engineering and Technology2022 · 제54권 제9호Tackling range uncertainty in proton therapy: Development and evaluation of a new multi-slit prompt-gamma camera (MSPGC) system구영모, 김성훈, 김찬형, 장세현, 정종휘, 조성구, 조재호, 최세훈한국원자력학회 · Nuclear Engineering and Technology2023 · 제55권 제9호Performance prediction of gamma electron vertex imaging (GEVI) system for interfractional range shift detection in spot scanning proton therapy구영모, 김성훈, 김찬형, 정재린, 정종휘한국원자력학회 · Nuclear Engineering and Technology2022 · 제54권 제6호Strategies to improve the range verification of stochastic origin ensembles for low-count prompt gamma imagingHuang Hsuan-Ming한국원자력학회 · Nuclear Engineering and Technology2023 · 제55권 제10호실리콘 광증폭기와 반응깊이 측정방법을 이용한 수술용 베타 영상/감마 프로브 가능성 연구강한규, 곽인석, 손정환, 이재성, 홍성종대한의용생체공학회 · 의공학회지2016 · 제37권 제1호

이 간행물의 다른 논문 (39건)