본문 바로가기
KERD
논문

환경 바이오모니터링에서의 혈중 중금속 분석 숙련도 시험을 위한 표준물질 제조 및 적용성 평가

Development and Validation of Reference Material for Proficiency Testing of Blood Heavy Metals in Environmental Biomonitoring

2

조회

0

다운로드

1

인용

0

참고문헌

키워드

초록

Objectives: This study developed and validated reference materials (RMs) to analyze metal compounds in blood. Methods: We referred to KoNEHS (Korea National Environmental Health Survey) to estimate concentrations of blood metals (cadmium, Cd; lead, Pb; mercury, Hg) and applied analytical methods (inductively coupled plasma – mass spectroscopy, ICP-MS, for Cd and Pb; graphite furnace – atomic absorption spectrometry, GFAAS, for Cd and Pb; and direct mercury analyzer, DMA, for Hg). Homogeneity and stability tests were carried out. In addition, certified values and uncertainties of RMs were calculated through internal and external experiments. All RMs were developed and assessed in various forms according to element, analytical method, and two types of concentration levels high concentration for occupational exposure and low concentration for environmental exposure. Results: All samples showed acceptable homogeneity, except for low concentration of Cd in the GF-AAS method. Short- and long-term stabilities were satisfied by ANOVA testing. In the inter-laboratory comparison, robust medians were lower than the certified values of all RMs (robust median/reference value; 1.301/1.327 µg/L for Cd, ICP-MS, low concentration; 3.152/3.388 µg/L for Cd, ICP-MS, high concentration; 1.219/1.301 µg/L for Cd, GF-AAS, low concentration; 3.074/3.321 µg/L for Cd, GF-AAS, high concentration; 14.473/14.516 µg/L for Pb, ICP-MS, low concentration; 50.069/50.114 µg/L for Pb, ICP-MS, high concentration; 12.881/14.147 µg/L for Pb, GF-AAS, low concentration; 47.015/47.591 µg/L for Pb, GF-AAS, high concentration; 4.059/4.218 µg/L for Hg, DMA, low concentration; 11.474/11.181 µg/L for Hg, DMA, high concentration). Conclusion: This study demonstrates procedures for developing and validating RMs for biomonitoring in the field of the environmental health.

Abstract

Objectives: This study developed and validated reference materials (RMs) to analyze metal compounds in blood. Methods: We referred to KoNEHS (Korea National Environmental Health Survey) to estimate concentrations of blood metals (cadmium, Cd; lead, Pb; mercury, Hg) and applied analytical methods (inductively coupled plasma – mass spectroscopy, ICP-MS, for Cd and Pb; graphite furnace – atomic absorption spectrometry, GFAAS, for Cd and Pb; and direct mercury analyzer, DMA, for Hg). Homogeneity and stability tests were carried out. In addition, certified values and uncertainties of RMs were calculated through internal and external experiments. All RMs were developed and assessed in various forms according to element, analytical method, and two types of concentration levels high concentration for occupational exposure and low concentration for environmental exposure. Results: All samples showed acceptable homogeneity, except for low concentration of Cd in the GF-AAS method. Short- and long-term stabilities were satisfied by ANOVA testing. In the inter-laboratory comparison, robust medians were lower than the certified values of all RMs (robust median/reference value; 1.301/1.327 µg/L for Cd, ICP-MS, low concentration; 3.152/3.388 µg/L for Cd, ICP-MS, high concentration; 1.219/1.301 µg/L for Cd, GF-AAS, low concentration; 3.074/3.321 µg/L for Cd, GF-AAS, high concentration; 14.473/14.516 µg/L for Pb, ICP-MS, low concentration; 50.069/50.114 µg/L for Pb, ICP-MS, high concentration; 12.881/14.147 µg/L for Pb, GF-AAS, low concentration; 47.015/47.591 µg/L for Pb, GF-AAS, high concentration; 4.059/4.218 µg/L for Hg, DMA, low concentration; 11.474/11.181 µg/L for Hg, DMA, high concentration). Conclusion: This study demonstrates procedures for developing and validating RMs for biomonitoring in the field of the environmental health.

참고문헌 (24)

  1. Kim S, "Application of biomonitoring to activities on environmental health and recommendations for Korean National Environmental Health Survey", The Korean Journal of Public Health, 52(1), pp.59~74 (2015)
  2. Becker K, "German health-related environmental monitoring : assessing time trends of the general population's exposure to heavy metals", Int J Hyg Environ Health, 216(3), pp.250~254 (2013)
  3. "National Health and Nutrition Examination Survey (NHANES)" (2016)[URL]
  4. Cerná M, "Human biomonitoring in the Czech Republic : an overview", Int J Hyg Environ Health, 215(2), pp.109~119 (2012)
  5. Lee JW, "Korea National Survey for Environmental Pollutants in the Human Body 2008 : heavy metals in the blood or urine of the Korean population", Int J Hyg Environ Health, 215(4), pp.449~457 (2012)
  6. Ha M, "Korean Environmental Health Survey in Children and Adolescents (KorEHS-C): survey design and pilot study results on selected exposure biomarkers", Int J Hyg Environ Health, 217(2), pp.260~270 (2014)
  7. Kim NS, "Blood lead levels of residents living around 350 abandoned metal mines in Korea", Environ Monit Assess, 184(7), pp.4139~4149 (2012)
  8. 홍석영, "토양 숙련도 시험용 금속 표준시료(PTMs) 제조방법 확립 및 적용성 평가", 한국환경분석학회지, 19(2), pp.126~134 (2016), 한국환경분석학회
  9. 황병훈, "한약재중 중금속 분석의 정도관리에 관한 연구", 한국환경분석학회지, 15(4), pp.275~281 (2012), 한국환경분석학회
  10. 김영희, "중금속분석용 폐수표준물질 제조 및 실험시간 비교평가", 분석과학, 23(3), pp.295~303 (2010), 한국분석과학회
  11. Maier EA, "Interlaboratory studies and certified reference materials for environmental analysis: the BCR approach. Vol. 22", pp.115~164 (1999), Elsevier
  12. "SRM 955C–Toxic metals in caprine blood"[URL]
  13. Haines DA, "Human biomonitoring of environmental chemicals-early results of the 2007-2009 Canadian Health Measures Survey for males and females", Int J Hyg Environ Health, 215(2), pp.133~137 (2012)
  14. "Fourth National Report on Human Exposure to Environmental Chemicals" (2015)[URL]
  15. "Results on the blood levels of heavy metal concentration in Korean" (2005)[URL]
  16. International Organization for Standardization, "Statistical methods for use in proficiency testing by interlaboratory comparisons", ~ (2005)
  17. International Organization for Standardization, "Guideline 35: reference materials – general and statistical principles for certification", ~ (2006)
  18. Fearn T, "A new test for ‘sufficient homogeneity’", Analyst, 126(8), pp.1414~1417 (2001)
  19. Thompson M, "The international harmonized protocol for the proficiency testing of(chemical)analytical laboratories(Technical report)", Pure Appl Chem, 65(9), pp.2123~2144 (1993)
  20. González AG, "A practical guide to analytical method validation, including measurement uncertainty and accuracy profiles", TrAC Trends in Analytical Chemistry, 26(3), pp.227~238 (2007)
  21. 고범준, "인증표준물질(CRM)을 이용한 소변 중 대마 대사체 분석법 평가", 분석과학, 23(1), pp.83~88 (2010), 한국분석과학회
  22. Yang D, "Development of a new ferulic acid certified reference material for use in clinical chemistry and pharmaceutical analysis", Acta Pharm Sin B, 5(3), pp.231~237 (2015)
  23. Nogueira R, "Development of a new sodium diclofenac certified reference material using the mass balance approach and ¹H qNMR to determine the certified property value", Eur J Pharm Sci, 48(3), pp.502~513 (2013)
  24. The amazing Horwitz function, "Thompson M.", Royal Society of Chemistry, 1~2, pp.AMC Technical Brief No. 17 (2004)

관련 논문AI추천

바이오모니터링 프로그램을 위한 혈중 금속류 동시분석법 개발 및 확인 평가오세림, 오은하, 임호섭, 차상원, 한상범한국환경보건학회 · 한국환경보건학회지2021 · 제47권 제1호흑연로 원자 흡광 광도기와 유도 결합 플라즈마 질량 분석기를 이용한 인체 혈중 카드뮴 농도 비교강민경, 권정연, 김병권, 김유미, 서정욱, 임현주, 홍영습한국환경보건학회 · 한국환경보건학회지2018 · 제44권 제5호국가 바이오모니터링 프로그램을 위한 소변 중 금속류 동시분석법 개발 및 검증김기훈, 양민호, 이재익, 임호섭, 조용민, 차상원, 한상범한국환경보건학회 · 한국환경보건학회지2019 · 제45권 제6호흑연로 원자 흡광 광도법과 유도 결합 플라즈마 질량 분석법을 이용한 혈중 납 농도 비교강민경, 권정연, 김병권, 김유미, 서정욱, 임현주, 홍영습한국환경보건학회 · 한국환경보건학회지2019 · 제45권 제3호혈장 중 극미량 납 분석을 위한 ICP-MS 분석법 검증김남수, 김용순, 신호상, 안병준, 이병국, 이성배, 이용훈한국분석과학회 · 분석과학2015 · 제28권 제5호Analysis of Methylmercury Concentration in the Blood of Koreans by Using Cold Vapor Atomic Fluorescence Spectrophotometry김경연, 김대선, 김록범, 김병권, 김유미, 서병성, 조은미, 홍영습대한진단검사의학회 · Annals of Laboratory Medicine2012 · 제32권 제1호

이 간행물의 다른 논문 (10건)