치과용 임플란트 적용을 위한 항균력을 가진 티타늄 표면의 평가

An Evaluation of Antibacterial Titanium Surface For Dental Implant

2

조회

0

다운로드

2

인용

0

참고문헌

키워드

주제분류치의학

초록

The aim of this study was to evaluate antibacterial effect of Cl coated titanium. To coat the Cl on the titanium, first, the titanium was modified by blasting treatment with hydroxyapatite and alumina powder. Anodization process was completed using electrolyte solution of 0.04 M β-glycerol phosphate disodium salt n-hydrate, 0.4 M calcium acetate n-hydrate and 1 M NaCl on the condition of 250 voltages for 3 min. Surface morphology and elements'observation were performed with scanning electron microscopy and energy dispersive spectroscopy and surface profiler was used to analyze the surface roughness. Antibacterial effect was evaluated by film adhesion method. The anodized titanium after blasting showed dimpled surface contained the Cl. Surface average roughness of these surfaces had significantly higher compared to polished titanium. Result of antibacterial test showed that anodized titanium after blasting had an enhanced antibacterial effect compared to the polished titanium. Therefore, these results suggested that titanium contained Cl by anodization after blasting had a rough surface as well as antibacterial effect.

Abstract

The aim of this study was to evaluate antibacterial effect of Cl coated titanium. To coat the Cl on the titanium, first, the titanium was modified by blasting treatment with hydroxyapatite and alumina powder. Anodization process was completed using electrolyte solution of 0.04 M β-glycerol phosphate disodium salt n-hydrate, 0.4 M calcium acetate n-hydrate and 1 M NaCl on the condition of 250 voltages for 3 min. Surface morphology and elements'observation were performed with scanning electron microscopy and energy dispersive spectroscopy and surface profiler was used to analyze the surface roughness. Antibacterial effect was evaluated by film adhesion method. The anodized titanium after blasting showed dimpled surface contained the Cl. Surface average roughness of these surfaces had significantly higher compared to polished titanium. Result of antibacterial test showed that anodized titanium after blasting had an enhanced antibacterial effect compared to the polished titanium. Therefore, these results suggested that titanium contained Cl by anodization after blasting had a rough surface as well as antibacterial effect.

참고문헌 (29)

  1. Albrektsson T, "The impact of oral implants-past and future,1966-2042", J Can Dent Assoc, 71(5), pp.327~327 (2005)
  2. Le Guéhennec L, "Surface treatments of titanium dental implants for rapid osseointegration", Dent Mater, 23(7), pp.844~854 (2007)
  3. Becker W, "A prospective multicenter clinical trial comparing one-and two-stage titanium screw-shaped fixtures with one-stage plasma-sprayed solid-screw fixtures", Clinic Implant Dent Relat Res, 2(3), pp.159~165 (2000)
  4. Conner KA, "Guided bone regeneration around titanium plasma-sprayed,acid-etched,and hydroxyapatite-coated implants in the canine model", J Periodont, 74(5), pp.658~668 (2003)
  5. O'Hare P, "Biological responses to hydroxyapatite surfaces deposited via a co-incident microblasting technique", Biomaterials, 31(3), pp.515~522 (2010)
  6. Ishizawa H, "Formation and characterization of anodic titanium oxide films containing Ca and P", J Biomed Mater Res, 29(1), pp.65~72 (1995)
  7. Snauwaert K, "Time dependent failure rate and marginal bone loss of implant supported prostheses:a 15-year followup study", Clinical Oral Investigations, 4(1), pp.13~20 (2004)
  8. Berglundh T, "A systematic review of the incidence of biological and technical complications in implant dentistry reported in prospective longitudinal studies of at least 5 years", J Clin Periodontol, 29(3), pp.197~212 (2002)
  9. Quirynen M, "Infectious risks for oral implants:a review of the literature", Clin Oral Implant Res, 13(1), pp.1~19 (2002)
  10. Klinge B, "Peri-implantitis", Dent Clin North Am, 49(3), pp.661~676 (2005)
  11. Adams CS, "Controlled release of vancomycin from thin sol-gel films on implant surfaces successfully controls osteomyelitis", J Orthop Res, 27(6), pp.701~709 (2009)
  12. Brohede U, "Multifunctional implant coatings providing possibilities for fast antibiotics loading with subsequent slow release", J Mater Sci:Mater Med, 20(9), pp.1859~1867 (2009)
  13. Radin S, "Controlled release of vancomycin from thin sol-gel films on titanium alloy fracture plate material", Biomaterials, 28(9), pp.1721~1729 (2007)
  14. Amoroso PF, "Titanium surface modification and its effect on the adherence of Porphyromonas gingivalis:an in vitro study", Clin Oral Implant Res, 17(6), pp.633~637 (2006)
  15. Yoshinari M, "Influence of surface modifications to titanium on oral bacterial adhesion in vitro", J Biomed Mater Res, 52(2), pp.388~394 (2000)
  16. Yoshinari M, "Influence of surface modifications to titanium on antibacterial activity in vitro", Biomaterials, 22(14), pp.2043~2048 (2001)
  17. Shibata Y, "Antibacterial titanium plate anodized by being discharging in NaCl solution exhibits cell compatibility", J Dent Res, 83(2), pp.115~119 (2004)
  18. Ishikawa K, "Blast coating method:new method of coating titanium surface with hydroxyapatite at room temperature", J Biomed Mater Res, 38(2), pp.129~134 (1997)
  19. 설영택, "Optimum surface properties of oxidized implants for reinforcement of osseointegration: Surface chemistry, oxide thickness, porosity, roughness, and crystal structure", INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 20(3), pp.349~359, QUINTESSENCE PUBLISHING CO INC
  20. Novaes Jr AB, "Histomorphometric analysis of the boneimplant contact obtained with 4 different implant surface treatments placed side by side in the dog mandible", Int J Oral Maxillofac Implants, 17(3), pp.377~383 (2002)
  21. Heitz-Mayfield LJ, "Peri-implant diseases:diagnosis and risk indicators", J Clin Periodontol, 35(8), pp.292~304 (2008)
  22. Kang M-K, "Antibacterial effects and cytocompatibility of titanium anodized in sodium chloride, calcium acetate, and ß-glycerol phosphate disodium salt pentahydrate mixed solution", Thin Solid Films, 517(17), pp.5390~5393, Elsevier BV
  23. Aparicio C, "Corrosion behaviour of commercially pure titanium shot blasted with different materials and sizes of shot particles for dental implant applications", Biomaterials, 24(2), pp.263~273 (2003)
  24. Ivanoff CJ, "Histologic evaluation of bone response to oxidized and turned titanium micro-implants in human jawbone", Int J Oral Maxillofac Implants, 18(3), pp.341~348 (2003)
  25. Nakagawara S, "Spectroscopic characterization and pH dependence of bactericidal activity of aqueous chlorine solution", Anal Sci, 14(4), pp.691~697 (1998)
  26. Nakajima N, "Evaiuation of disinfective potential of reactivated free chlorine in pooled tap water by electrolysis", J Microbiol Methods, 57(2), pp.163~173 (2004)
  27. 이명주, "임플란트 시술에 대한 환자의 기능별 만족도", 치위생과학회지, 8(2), pp.97~101 (2008), 한국치위생과학회
  28. 유은미, "치과의료소비자의 임플란트에 대한 지식과 기대도의 관련성", 치위생과학회지, 10(4), pp.219~225 (2010), 한국치위생과학회
  29. 한지형, "임플란트 시술환자 시술 전·후의 기대와 만족도 비교", 치위생과학회지, 11(2), pp.121~127 (2011), 한국치위생과학회

관련 논문AI추천

이 간행물의 다른 논문 (11건)