Development of 1H-31P Animal RF Coil for pH Measurement Using a Clinical MR Scanner

1

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류방사선과학

초록

Purpose : To establish a pH measurement system for a mouse tumor study using a clinical scanner, to develop the 1H and 31P radio frequency (RF) coil system and to test pH accuracy with phantoms. Materials and Methods: The 1H and the 31P surface coils were designed to acquire signals from mouse tumors. Two coils were positioned orthogonally for geometric decoupling. The pH values of various pH phantoms were calculated using the 1H decoupled 31P MR spectrum with the Henderson-Hasselbalch equation. The calculated pH value was compared to that of a pH meter. Results: The mutual coil coupling was shown in a standard S12. Coil coupling (S12) were -73.0 and -62.3 ㏈ respectively. The signal-to-noise ratio (SNR) obtained from the homogeneous phantom 1H image was greater than 300. The high resolution in vivo mice images were acquired using a 31P-decoupled 1H coil. The pH values calculated from the 1H-decoupled 31P spectrum correlated well with the values measured by pH meter (R2=0.97). Conclusion: Accurate pH values can be acquired using a 1H-decoupled 31P RF coil with a clinical scanner. This two-surface coil system could be applied to other nuclear MRS or MRI.

Abstract

Purpose : To establish a pH measurement system for a mouse tumor study using a clinical scanner, to develop the 1H and 31P radio frequency (RF) coil system and to test pH accuracy with phantoms. Materials and Methods: The 1H and the 31P surface coils were designed to acquire signals from mouse tumors. Two coils were positioned orthogonally for geometric decoupling. The pH values of various pH phantoms were calculated using the 1H decoupled 31P MR spectrum with the Henderson-Hasselbalch equation. The calculated pH value was compared to that of a pH meter. Results: The mutual coil coupling was shown in a standard S12. Coil coupling (S12) were -73.0 and -62.3 ㏈ respectively. The signal-to-noise ratio (SNR) obtained from the homogeneous phantom 1H image was greater than 300. The high resolution in vivo mice images were acquired using a 31P-decoupled 1H coil. The pH values calculated from the 1H-decoupled 31P spectrum correlated well with the values measured by pH meter (R2=0.97). Conclusion: Accurate pH values can be acquired using a 1H-decoupled 31P RF coil with a clinical scanner. This two-surface coil system could be applied to other nuclear MRS or MRI.

참고문헌 (21)

  1. Glunde K, "Metabolic tumor imaging using magnetic resonance spectroscopy", Semin Oncol, 38, pp.26~41 (2011)
  2. Stubbs M, "An assessment of 31P MRS as a method of measuring pH in rat tumours", NMR Biomed, 5, pp.351~359 (1992)
  3. Zhou R, "Intracellular acidification of human melanoma xenografts by the respiratory inhibitor m-iodobenzylguanidine plus hyperglycemia: a 31P magnetic resonance spectroscopy study", Cancer Res, 60, pp.3532~3236 (2000)
  4. Gallagher FA, "Imaging pH with hyperpolarized 13C", NMR Biomed, 24, pp.1006~1015 (2011)
  5. Mason RP, "Perfluorocarbon imaging in vivo: a 19F MRI study in tumorbearing mice", Magn Reson Imaging, 7, pp.475~485 (1989)
  6. McSheehy PM, "A pharmacokinetic and pharmacodynamic study in vivo of human HT29tumours using 19F and 31P magnetic resonance spectroscopy", Eur J Cancer, 33, pp.2418~2427 (1997)
  7. Gillies RJ, "pH imaging. A review of pH measurement methods and applications in cancers", IEEE Eng Med Biol Mag, 23, pp.57~64 (2004)
  8. Iessi E, "Tumor acidity and malignancy: novel aspects in the design of antitumor therapy", Cancer Therapy, 6, pp.55~66 (2008)
  9. Hashim AI, "Imaging pH and metastasis", NMR Biomed, 24, pp.582~591 (2011)
  10. Gillies RJ, "MRI of the tumor microenvironment", J Magn Reson Imaging, 16, pp.430~450 (2002)
  11. Negendank W, "Studies of human tumors by MRS: a review", NMR Biomed, 5, pp.303~324 (1992)
  12. Bottomley PA, "Protondecoupled, Overhauser-enhanced, spatially localized carbon-13spectroscopy in humans", Magn Reson Med, 12, pp.348~363 (1989)
  13. Merkle H, "B1-insensitive heteronuclear adiabatic polarization transfer for signal enhancement", J Magn Reson, 99, pp.480~494 (1992)
  14. Adriany G, "A half-volume coil for efficient proton decoupling in humans at 4 Tesla", J Magn Reson, 125, pp.178~184 (1997)
  15. "WMIC 2013 P266 New Design of dual Tuned RF coil for Fluorine MR Molecular Imaging", ~ (2013)
  16. Barberi EA, "A transmitonly", receive-only (TORO) RF system for high-field MRI, Magn Reson Med, pp.2000, 43[DOI]
  17. Cassidy PJ, "Determining the tuning and matching requirements of RF coils using electromagnetic simulation and electric circuit analysis", Concepts in Magn Reson Part B Magn Reson Eng, 25, pp.27~41 (2005)
  18. De Graaf RA, "In vivo NMR Spectroscopy - Principles and Techniques", pp.80~82 (2007), John Wiley & Sons Ltd
  19. Larcombe-McDouall J, "In vivo pH and metabolite changes during a single contraction in rat uterine smooth muscle", J Physiol, 518, pp.783~790 (1999)
  20. Babic SI, "Calculating mutual inductance between circular coils with inclined axes in air", IEEE Trans. Magn, 44, pp.1743~1750 (2008)
  21. Grover FW, "Inductance Calculations", pp.193~208 (1964), Dover

관련 논문AI추천

이 간행물의 다른 논문 (9건)