Purpose: The objective of this study is to determine the effect of physical changes on MR temperature imaging at 7.0T and to examine proton-resonance-frequency related changes of MR phase images and T1 related changes of MR magnitude images, which are obtained for MR thermometry at various magnetic field strengths. Materials and Methods: An MR-compatible capacitive-coupled radio-frequency hyperthermia system was implemented for heating a phantom and swine muscle tissue, which can be used for both 7.0T and 3.0T MRI. To determine the effect of flip angle correction on T1-based MR thermometry, proton resonance frequency, apparent T1, actual flip angle, and T1 images were obtained. For this purpose, three types of imaging sequences are used, namely, T1-weighted fast field echo with variable flip angle method, dual repetition time method, and variable flip angle method with radio-frequency field nonuniformity correction. Results: Signal-to-noise ratio of the proton resonance frequency shift-based temperature images obtained at 7.0T was five-fold higher than that at 3.0T. The T1 value increases with increasing temperature at both 3.0T and 7.0T. However, temperature measurement using apparent T1-based MR thermometry results in bias and error because B1 varies with temperature. After correcting for the effect of B1 changes, our experimental results confirmed that the calculated T1 increases with increasing temperature both at 3.0T and 7.0T. Conclusion: This study suggests that the temperature-induced flip angle variations need to be considered for accurate temperature measurements in T1-based MR thermometry.

Abstract

Purpose: The objective of this study is to determine the effect of physical changes on MR temperature imaging at 7.0T and to examine proton-resonance-frequency related changes of MR phase images and T1 related changes of MR magnitude images, which are obtained for MR thermometry at various magnetic field strengths. Materials and Methods: An MR-compatible capacitive-coupled radio-frequency hyperthermia system was implemented for heating a phantom and swine muscle tissue, which can be used for both 7.0T and 3.0T MRI. To determine the effect of flip angle correction on T1-based MR thermometry, proton resonance frequency, apparent T1, actual flip angle, and T1 images were obtained. For this purpose, three types of imaging sequences are used, namely, T1-weighted fast field echo with variable flip angle method, dual repetition time method, and variable flip angle method with radio-frequency field nonuniformity correction. Results: Signal-to-noise ratio of the proton resonance frequency shift-based temperature images obtained at 7.0T was five-fold higher than that at 3.0T. The T1 value increases with increasing temperature at both 3.0T and 7.0T. However, temperature measurement using apparent T1-based MR thermometry results in bias and error because B1 varies with temperature. After correcting for the effect of B1 changes, our experimental results confirmed that the calculated T1 increases with increasing temperature both at 3.0T and 7.0T. Conclusion: This study suggests that the temperature-induced flip angle variations need to be considered for accurate temperature measurements in T1-based MR thermometry.

참고문헌 (24)

  1. Lewa CJ, "Temperature relationships of proton spin-lattice relaxation time T1 in biological tissues", Bull Cancer, 67, pp.525~530 (1980)
  2. Graham SJ, "Time and temperature dependence of MR parameters during thermal coagulation of ex vivo rabbit muscle", Magn Reson Med, 39, pp.198~203 (1998)
  3. Ishihara Y, "A precise and fast temperature mapping using water proton chemical shift", Magn Reson Med, 34, pp.814~823 (1995)
  4. Kim JM, "Monitoring and guidance on high-intensity focused ultrasound treatment by multiple fast field echo at 3.0 T MRI: ex-vivo studies with multiparametric mapping", Proceedings of the 26th Annual Meeting ISMRM, pp.1485~ (2018)
  5. Moseley ME, "Early detection of regional cerebral ischemia in cats: comparison of diffusion- and T2-weighted MRI and spectroscopy", Magn Reson Med, 14, pp.330~346 (1990)
  6. Young IR, "Modeling and observation of temperature changes in vivo using MRI", Magn Reson Med, 32, pp.358~369 (1994)
  7. Langereis S, "A temperature-sensitive liposomal 1H CEST and 19F contrast agent for MR image-guided drug delivery", J Am Chem Soc, 131, pp.1380~1381 (2009)
  8. Schwan HP, "The conductivity of living tissues", Ann N Y Acad Sci, 65, pp.1007~1013 (1957)
  9. Yeo SY, "Conductivity imaging for assessing the treatment outcome of MR-HIFU ablation of uterine fibroids", Proceedings of the 25th Annual Meeting ISMRM, pp.2102~ (2017)
  10. Kwon DH, "Seven-Tesla magnetic resonance images of the substantia nigra in Parkinson disease", Ann Neurol, 71, pp.267~277 (2012)
  11. Goelman G, "Chemical-shift artifact reduction in Hadamardencoded MR spectroscopic imaging at high (3T and 7T)magnetic fields", Magn Reson Med, 58, pp.167~173 (2007)
  12. Rieke V, "MR thermometry", J Magn Reson Imaging, 27, pp.376~390 (2008)
  13. Wang P., "Evaluation of MR thermometry with proton resonance frequency method at 7T", Quant Imaging Med Surg, 7, pp.259~266 (2017)
  14. Fite BZ, "Magnetic resonance thermometry at 7T for real-time monitoring and correction of ultrasound induced mild hyperthermia", PLoS One, 7, pp.e35509~ (2012)
  15. Henderson E, "A fast 3D looklocker method for volumetric T1 mapping", Magn Reson Imaging, 17, pp.1163~1171 (1999)
  16. Zhu DC, "Full-brain T1 mapping through inversion recovery fast spin echo imaging with time-efficient slice ordering", Magn Reson Med, 54, pp.725~731 (2005)
  17. Cheng HL, "Rapid high-resolution T(1) mapping by variable flip angles: accurate and precise measurements in the presence of radiofrequency field inhomogeneity", Magn Reson Med, 55, pp.566~574 (2006)
  18. Liberman G, "T(1) mapping using variable flip angle SPGR data with flip angle correction", J Magn Reson Imaging, 40, pp.171~180 (2014)
  19. Lee Y, "Analysis of the precision of variable flip angle T1 mapping with emphasis on the noise propagated from RF transmit field maps", Front Neurosci, 11, pp.106~ (2017)
  20. Yarnykh VL., "Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field", Magn Reson Med, 57, pp.192~200 (2007)
  21. Helms G, "Quantitative FLASH MRI at 3T using a rational approximation of the Ernst equation", Magn Reson Med, 59, pp.667~672 (2008)
  22. Kim HJ, "Experiment study of MR compatible RF hyperthermia system", Proceedings of the 24th Annual Meeting ISMRM, pp.1644~ (2016)
  23. 김종민, "온도범위에 따른 비침습적 자기공명 온도측정방법의 최적화", 의공학회지, 36(6), pp.241~250 (2015), 대한의용생체공학회
  24. Kim JM, "Comparison of temperature mapping methods using proton resonance frequency shift and T1 in 3-T and 7-T MRI", Proceedings of the 25th Annual Meeting ISMRM, pp.2588~ (2017)