Radiofrequency Coil Design for in vivo Sodium Magnetic Resonance Imaging of Mouse Kidney at 9.4T

1

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류방사선과학

초록

The objective of this study was to describe a radiofrequency (RF) coil design for in vivo sodium magnetic resonance imaging (MRI) for use in small animals. Accumulating evidence has indicated the importance and potential of sodium imaging with improved magnet strength (> 7T), faster gradient, better hardware, multi-nucleus imaging methods, and optimal coil design for patient and animal studies. Thus, we developed a saddle-shaped sodium volume coil with a diameter/ length of 30/30 mm. To evaluate the efficiency of this coil, bench-level measurement was performed. Unloaded Q value, loaded Q value, and ratio of these two values were estimated to be 352.8, 211.18, and 1.67, respectively. Thereafter, in vivo acquisition of sodium images was performed using normal mice (12 weeks old; n = 5) with a twodimensional gradient echo sequence and minimized echo time to increase spatial resolution of images. Sodium signal-to-noise ratio in mouse kidneys (renal cortex, medulla, and pelvis) was measured. We successfully acquired sodium MR images of the mouse kidney with high spatial resolution (approximately 0.625 mm) through a combination of sodium-proton coils.

Abstract

The objective of this study was to describe a radiofrequency (RF) coil design for in vivo sodium magnetic resonance imaging (MRI) for use in small animals. Accumulating evidence has indicated the importance and potential of sodium imaging with improved magnet strength (> 7T), faster gradient, better hardware, multi-nucleus imaging methods, and optimal coil design for patient and animal studies. Thus, we developed a saddle-shaped sodium volume coil with a diameter/ length of 30/30 mm. To evaluate the efficiency of this coil, bench-level measurement was performed. Unloaded Q value, loaded Q value, and ratio of these two values were estimated to be 352.8, 211.18, and 1.67, respectively. Thereafter, in vivo acquisition of sodium images was performed using normal mice (12 weeks old; n = 5) with a twodimensional gradient echo sequence and minimized echo time to increase spatial resolution of images. Sodium signal-to-noise ratio in mouse kidneys (renal cortex, medulla, and pelvis) was measured. We successfully acquired sodium MR images of the mouse kidney with high spatial resolution (approximately 0.625 mm) through a combination of sodium-proton coils.

참고문헌 (17)

  1. Maiti AK, "Enhancement in cellular Na+K+ATPase activity by low doses of peroxynitrite in mouse renal tissue and in cultured HK2 cells", Physiol Rep, 4 (2016)
  2. Wetterling F, "A double-tuned (1)H", (23)Na dual resonator system for tissue sodium concentration measurements in the rat brain via Na-MRI, 2010, pp.55, 7681~7695
  3. Haneder S, "In vivo sodium (23Na) imaging of the human kidneys at 7 T: preliminary results", Eur Radiol, 24, pp.494~501 (2014)[DOI]
  4. Wetterling F, "The design of a double-tuned two-port surface resonator and its application to in vivo hydrogen- and sodium-MRI", J Magn Reson, 217, pp.10~18 (2012)[DOI]
  5. Brown R, "A flexible nested sodium and proton coil array with wideband matching for knee cartilage MRI at 3T", Magn Reson Med, 76, pp.1325~1334 (2016)[DOI]
  6. Lykowsky G, "Optimization and comparison of two practical dual-tuned birdcage configurations for quantitative assessment of articular cartilage with sodium magnetic resonance imaging", Quant Imaging Med Surg, 5, pp.799~805 (2015)
  7. Hattori K, "Development of MRI phantom equivalent to human tissues for 3.0-T MRI", Med Phys, 40, pp.032303~ (2013)
  8. Constantinides CD, "Human skeletal muscle: sodium MR imaging and quantification-potential applications in exercise and disease", Radiology, 216, pp.559~568 (2000)
  9. Moon CH, "Quantitative sodium MR imaging of native versus transplanted kidneys using a dual-tuned proton", sodium (1H, Eur Radiol, pp.2014, 24[DOI]
  10. Roemer PB, "The NMR phased array", Magn Reson Med, 16, pp.192~225 (1990)[DOI]
  11. Buist RJ, "23Na and flame photometric studies of the NMR visibility of sodium in rat muscle", Can J Physiol Pharmacol, 69, pp.1663~1669 (1991)[DOI]
  12. Zhang X, "Microstrip RF surface coil design for extremely high-field MRI and spectroscopy", Magn Reson Med, 46, pp.443~450 (2001)[DOI]
  13. Jasinski K, "A volume microstrip RF coil for MRI microscopy", Magn Reson Imaging, 30, pp.70~77 (2012)
  14. Fujita H, "New horizons in MR technology: RF coil designs and trends", Magn Reson Med Sci, 6, pp.29~42 (2007)[DOI]
  15. Rutledge O, "Design and test of a double-nuclear RF coil for (1)H MRI and (13)C MRSI at 7T", J Magn Reson, 267, pp.15~21 (2016)
  16. Nielles-Vallespin S, "3D radial projection technique with ultrashort echo times for sodium MRI: clinical applications in human brain and skeletal muscle", Magn Reson Med, 57, pp.74~81 (2007)[DOI]
  17. Kharrazian R, "Dynamics of 23Na during completely balanced steady-state free precession", J Magn Reson, 179, pp.73~84 (2006)

관련 논문AI추천

이 간행물의 다른 논문 (9건)