Quantitative Conductivity Estimation Error due to Statistical Noise in Complex B1 + Map

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

In-vivo conductivity reconstruction using transmit field (B1+) information of MRI was proposed. We assessed theaccuracy of conductivity reconstruction in the presence of statistical noise in complex B1+ map and provided a parametricmodel of the conductivity-to-noise ratio value. Materials and Methods: The B1+ distribution was simulated for a cylindrical phantom model. By adding complex Gaussiannoise to the simulated B1+ map, quantitative conductivity estimation error was evaluated. The quantitative evaluationprocess was repeated over several different parameters such as Larmor frequency, object radius and SNR of B1+ map. Aparametric model for the conductivity-to-noise ratio was developed according to these various parameters. Results: According to the simulation results, conductivity estimation is more sensitive to statistical noise in B1+ phase thanto noise in B1+ magnitude. The conductivity estimate of the object of interest does not depend on the external object surroundingit. The conductivity-to-noise ratio is proportional to the signal-to-noise ratio of the B1+ map, Larmor frequency,the conductivity value itself and the number of averaged pixels. To estimate accurate conductivity value of the targetedtissue, SNR of B1+ map and adequate filtering size have to be taken into account for conductivity reconstruction process. In addition, the simulation result was verified at 3T conventional MRI scanner. Conclusion: Through all these relationships, quantitative conductivity estimation error due to statistical noise in B1+ map ismodeled. By using this model, further issues regarding filtering and reconstruction algorithms can be investigated for MREPT.

Abstract

In-vivo conductivity reconstruction using transmit field (B1+) information of MRI was proposed. We assessed theaccuracy of conductivity reconstruction in the presence of statistical noise in complex B1+ map and provided a parametricmodel of the conductivity-to-noise ratio value. Materials and Methods: The B1+ distribution was simulated for a cylindrical phantom model. By adding complex Gaussiannoise to the simulated B1+ map, quantitative conductivity estimation error was evaluated. The quantitative evaluationprocess was repeated over several different parameters such as Larmor frequency, object radius and SNR of B1+ map. Aparametric model for the conductivity-to-noise ratio was developed according to these various parameters. Results: According to the simulation results, conductivity estimation is more sensitive to statistical noise in B1+ phase thanto noise in B1+ magnitude. The conductivity estimate of the object of interest does not depend on the external object surroundingit. The conductivity-to-noise ratio is proportional to the signal-to-noise ratio of the B1+ map, Larmor frequency,the conductivity value itself and the number of averaged pixels. To estimate accurate conductivity value of the targetedtissue, SNR of B1+ map and adequate filtering size have to be taken into account for conductivity reconstruction process. In addition, the simulation result was verified at 3T conventional MRI scanner. Conclusion: Through all these relationships, quantitative conductivity estimation error due to statistical noise in B1+ map ismodeled. By using this model, further issues regarding filtering and reconstruction algorithms can be investigated for MREPT.

참고문헌 (24)

  1. Haacke EM, "Extraction of conductivity and permittivity using magnetic resonance imaging", Phys Med Biol, 36, pp.723~734 (1991)
  2. Katscher U, "Determination of electric conductivity and local SAR via B1 mapping", IEEE Trans Med Imaging, 28, pp.1365~1374 (2009)
  3. Joines WT, "The measured electrical properties of normal and malignant human tissues from 50 to 900 MHz", Med Phys, 21, pp.547~550 (1994)
  4. Schaefer M, "The complex dielectric spectrum of heart tissue during ischemia", Bioelectro-chemistry, 58, pp.171~180 (2002)
  5. Haemmerich D, "In vivo electrical conductivity of hepatic tumours", Physiol Meas, 24, pp.251~260 (2003)
  6. Fallert MA, "Myocardial electrical-impedance mapping of iscemic sheep hearts and healing aneurysm", Circulation, 87, pp.199~207 (1993)
  7. Voigt T, "Patient-individual local SAR determination: in vivo measurements and numerical validation", Magn Reson Med, 68, pp.1117~1126 (2012)
  8. Yarnykh VL, "Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field", Magn Reson Med, 57, pp.192~200 (2007)
  9. Voigt T, "T1 corrected B1mapping using multi-TR gradient echo sequences", Magn Reson Med, 64, pp.725~733 (2010)
  10. Stollberger R, "Imaging of the active B1 field in vivo", Magn Reson Med, 35, pp.246~251 (1996)
  11. Sacolick LI, "B1 mapping by Bloch-Siegert shift", Magn Reson Med, 63, pp.1315~1322 (2010)
  12. Kim DH, "Simultaneous Electromagnetic Property Imaging using multiecho gradient echo", Proceedings of the 20th Annual Meeting of ISMRM, pp.3464~ (2012)
  13. Voigt T, "Quantitative conductivity and permittivity imaging of the human brain using electric properties tomography", Magn Reson Med, 66, pp.456~466 (2011)
  14. Bulumulla SB, "Conductivity and permittivity imaging at 3.0 T", Concept Magn Reson B, 41B(1), pp.13~21 (2012)
  15. van Lier AL, "B1(+) phase mapping at 7 T and its application for in vivo electrical conductivity mapping", Magn Reson Med, 67, pp.552~561 (2012)
  16. Seo JK, "Error analysis of nonconstant admittivity for MR-based electric property imaging", IEEE Trans Med Imaging, 31, pp.430~437 (2012)
  17. van Lier AL, "Comparing Electric Properties Tomography at 1.5, 3 and 7 T", Proceedings of the 19th Annual Meeting of ISMRM, pp.125~ (2011)
  18. van den Bergen B, "Ultra fast electromagnetic field computations for RF multi-transmit techniques in high field MRI", Phys Med Biol, 54, pp.1253~1264 (2009)
  19. Gudbjartsson H, "The Rician distribution of noisy MRI data", Magn Reson Med, 34, pp.910~914 (1995)
  20. Morrell GR, "An analysis of the accuracy of magnetic resonance flip angle measurement methods", Phys Med Biol, 55, pp.6157~6174 (2010)
  21. Stogryn A, "Equations for calculating the dielectric constant of saline water", IEEE Trans Microwave Theory Tech, 19, pp.733~736 (1971)
  22. Gabriel S, "The dielectric properties of biological tissues: II. Measurements in the frequency range 10Hz to 20 GHz", Phys Med Biol, 41, pp.2251~2269 (1996)
  23. Zypman FR, "MRI electromagnetic field penetration in cylindrical objects", Comput Biol Med, 26, pp.161~175 (1996)

관련 논문AI추천

이 간행물의 다른 논문 (9건)