A Novel Magnetic Resonance Quality Assurance Phantom (KMRP-4): Multi-Site Comparison With the American College of Radiology Phantom

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

Purpose: To propose a novel standard magnetic resonance imaging (MRI) phantom, hereafter called the Korea Magnetic Resonance Phantom-4th edition (KMRP-4). Its related quality control (QC) assessment protocols and its comparison with the American College of Radiology (ACR) phantom and its QC assessment protocols. Materials and Methods: Internally, the KMRP-4 phantom is composed of cubic and triangular vessels, brain tissue structures, and a uniform region designed to facilitate a variety of QC protocols. Using magnetic resonance (MR) images of these structures, we quantitatively evaluated a total of 10 parameters, seven from those of existing ACR protocols (i.e., geometric accuracy, high-contrast spatial resolution, slice thickness accuracy, slice position accuracy, image intensity uniformity, percent signal ghosting, and low-contrast object detectability) and three additional parameters for evaluating vessel conspicuity, brain tissue contrast, and signal-to-noise ratio (SNR) introduced in the KMRP-4 protocols. Twentytwo MRI systems of 0.32–3.0 T static magnetic field strength were tested using both ACR and KMRP-4 phantoms. Mann–Whitney U-tests were performed on the seven evaluation items of the ACR method to compare KMRP-4 and ACR methods. Results: The results of Mann–Whitney U-test demonstrated that p-values were more than 0.05 for all seven items that could be assessed with both ACR and KMRP-4, indicating similar results between the two methods. Additionally, assessments of vessel conspicuity, brain tissue contrast, and SNR using the KMRP-4 method demonstrated utility of the KMRP-4 phantom. Conclusion: A novel standard phantom and related QC methods were developed to perform objective, observer-independent, and semi-automatic QC tests. Quantitative comparisons of MR images with KMPR-4 and ACR phantoms were performed. Results demonstrated the utility of the newly proposed KMRP-4 phantom and its related QC methods.

Abstract

Purpose: To propose a novel standard magnetic resonance imaging (MRI) phantom, hereafter called the Korea Magnetic Resonance Phantom-4th edition (KMRP-4). Its related quality control (QC) assessment protocols and its comparison with the American College of Radiology (ACR) phantom and its QC assessment protocols. Materials and Methods: Internally, the KMRP-4 phantom is composed of cubic and triangular vessels, brain tissue structures, and a uniform region designed to facilitate a variety of QC protocols. Using magnetic resonance (MR) images of these structures, we quantitatively evaluated a total of 10 parameters, seven from those of existing ACR protocols (i.e., geometric accuracy, high-contrast spatial resolution, slice thickness accuracy, slice position accuracy, image intensity uniformity, percent signal ghosting, and low-contrast object detectability) and three additional parameters for evaluating vessel conspicuity, brain tissue contrast, and signal-to-noise ratio (SNR) introduced in the KMRP-4 protocols. Twentytwo MRI systems of 0.32–3.0 T static magnetic field strength were tested using both ACR and KMRP-4 phantoms. Mann–Whitney U-tests were performed on the seven evaluation items of the ACR method to compare KMRP-4 and ACR methods. Results: The results of Mann–Whitney U-test demonstrated that p-values were more than 0.05 for all seven items that could be assessed with both ACR and KMRP-4, indicating similar results between the two methods. Additionally, assessments of vessel conspicuity, brain tissue contrast, and SNR using the KMRP-4 method demonstrated utility of the KMRP-4 phantom. Conclusion: A novel standard phantom and related QC methods were developed to perform objective, observer-independent, and semi-automatic QC tests. Quantitative comparisons of MR images with KMPR-4 and ACR phantoms were performed. Results demonstrated the utility of the newly proposed KMRP-4 phantom and its related QC methods.

참고문헌 (20)

  1. "Korean Institute for Accreditation of Medical Imaging (KIAMI)"[URL]
  2. 김유성, "Image Quality Improvement after Implementation of a CT Accreditation Program", Korean Journal of Radiology, 11(5), pp.553~559 (2010), 대한영상의학회
  3. American College of Radiology, "Phantom test guidance for the ACR MRI accreditation program", American College of Radiology, ~ (1998)
  4. Nordell B, "A rotating phantom for the study of flow effects in MR imaging", Magn Reson Imaging, 6, pp.695~705 (1988)
  5. Jovicich J, "Reliability in multi-site structural MRI studies : effects of gradient non-linearity correction on phantom and human data", Neuroimage, 30, pp.436~443 (2006)
  6. Mizowaki T, "Reproducibility of geometric distortion in magnetic resonance imaging based on phantom studies", Radiother Oncol, 57, pp.237~242 (2000)
  7. Krämer SC, "3.0 tesla magnetic resonance angiography of endovascular aortic stent grafts: phantom measurements in comparison with 1.5 tesla", Invest Radiol, 39, pp.413~417 (2004)
  8. Hattori K, "Development of MRI phantom equivalent to human tissues for 3.0-T MRI", Med Phys, 40, pp.032303~ (2013)
  9. Fellner C, "A high-resolution phantom for MRI", Magn Reson Imaging, 19, pp.899~904 (2001)
  10. Davids M, "Fully-automated quality assurance in multi-center studies using MRI phantom measurements", Magn Reson Imaging, 32, pp.771~780 (2014)
  11. Steckner MC, "Computing the modulation transfer function of a magnetic resonance imager", Med Phys, 21, pp.483~489 (1994)
  12. Hartung MP, "Magnetic resonance angiography : current status and future directions", J Cardiovasc Magn Reson, 13, pp.19~ (2011)
  13. ASTM International, "Standard test method for evaluation of MR image artifacts from passive implants F2119-07", ASTM International, ~ (2007)
  14. Draper NR, "Applied regression analysis (vol. 326)" (1998), John Wiley & Sons
  15. Whittall KP, "In vivo measurement of T2 distributions and water contents in normal human brain", Magn Reson Med, 37, pp.34~43 (1997)
  16. Deoni SC, "Visualization of thalamic nuclei on high resolution, multi-averaged T1 and T2 maps acquired at 1.5 T", Hum Brain Mapp, 25, pp.353~359 (2005)
  17. National Electrical Manufacturers Association, "Determination of signal-to-noise ratio (SNR) in diagnostic magnetic resonance imaging. NEMA Standards Publication MS 1-2008 (R2014, R2020)" (2020), NEMA
  18. National Electrical Manufacturers Association, "Characterization of phased array coils for diagnostic magnetic resonance images. NEMA Standards Publication MS 9-2008 (R2014, R2020)" (2020), NEMA
  19. Eldin AB, "Modern approaches to quality control" (2011), InTech
  20. Tofts P, "Quantitative MRI of the brain : measuring changes caused by disease", pp.55~81 (2005), John Wiley & Sons

관련 논문AI추천

이 간행물의 다른 논문 (7건)