자기공명영상에서 ACR 팬텀을 이용한 해상력 특성 평가

Evaluation of the Resolution Characteristics by Using American College of Radiology Phantom for Magnetic Resonance Imaging

2

조회

0

다운로드

3

인용

0

참고문헌

키워드

주제분류방사선과학

초록

This study was purpose to quantitative assessment of the resolution characteristics by using American college of radiology(ACR) phantom for magnetic resonance imaging (MRI). The MRI equipment was used (Achiva 3.0T MRI, Philips system, Netherlands) and the head/neck matrix shim SENSE head coil were 32 channels(elements) receive MR coil. And the MRI equipment was used (Discovery MR 750, 3.0T MRI, GE medical system, America) and the head/neck matrix shim MC 3003G-32R 32-CH head coil were receive MR coil. As for the modulation transfer function(MTF) comparison result by using ACR magnetic resonance imaging phantom, the MTF value of the ACR standard T2 image in GE equipment is 0.199 when the frequency is 1.0 mm-1 and the MTF value of the hospital T2 image in Philips equipment is 0.528. It was used efficiently by using a general sequence more than the standard sequence method using the ACR phantom. In addition it is significant that the quantitative quality assurance evaluation method for resolution characteristics was applied mutatis mutandis, and the result values of the physical image characteristics of the 3.0T MRI device were presented.

Abstract

This study was purpose to quantitative assessment of the resolution characteristics by using American college of radiology(ACR) phantom for magnetic resonance imaging (MRI). The MRI equipment was used (Achiva 3.0T MRI, Philips system, Netherlands) and the head/neck matrix shim SENSE head coil were 32 channels(elements) receive MR coil. And the MRI equipment was used (Discovery MR 750, 3.0T MRI, GE medical system, America) and the head/neck matrix shim MC 3003G-32R 32-CH head coil were receive MR coil. As for the modulation transfer function(MTF) comparison result by using ACR magnetic resonance imaging phantom, the MTF value of the ACR standard T2 image in GE equipment is 0.199 when the frequency is 1.0 mm-1 and the MTF value of the hospital T2 image in Philips equipment is 0.528. It was used efficiently by using a general sequence more than the standard sequence method using the ACR phantom. In addition it is significant that the quantitative quality assurance evaluation method for resolution characteristics was applied mutatis mutandis, and the result values of the physical image characteristics of the 3.0T MRI device were presented.

참고문헌 (22)

  1. 민정환, "3.0T 자기공명영상을 이용한 해상력 특성에 대한 연구", 방사선기술과학, 43(4), pp.251~257 (2020), 대한방사선과학회
  2. 민정환, "3.0T 자기공명영상을 이용한 잡음전력스펙트럼 특성 평가", 방사선기술과학, 44(1), pp.31~37 (2021), 대한방사선과학회
  3. 이정휘, "자기공명영상 장치의 정도관리를 위한 ACR 팬텀의 유용성 및 적정성", 대한영상의학회지, 54(1), pp.47~55 (2006), 대한영상의학회
  4. Miyati T, "Measurements of MTF and SNR(f)using a subtraction method im MRI", Physics in Medicine and Biology, 47, pp.961~972 (2002)
  5. Steckner MC, "Computing the modulation transfer function of a magnetic resonance imager", Med Phys, 21, pp.483~489 (1994)
  6. Hahm HK, "The study on the subject development of MRI image quality evaluation", Department of Radiology, pp.2008
  7. 민정환, "간접평판형 검출기에서 국제전자기술위원회 기준을 통한 잡음전력스펙트럼 비교 연구", 방사선기술과학, 41(5), pp.457~462 (2018), 대한방사선과학회[DOI]
  8. 민정환, "간접평판형 검출기에서 변조전달함수 측정 시 Edge 각도에 따른 비교 연구", 방사선기술과학, 42(4), pp.259~263 (2019), 대한방사선과학회
  9. 정회원, "다양한 디지털 방사선 시스템의 물리적 영상 특성 조사", 방사선기술과학, 40(3), pp.363~370 (2017), 대한방사선과학회[DOI]
  10. 정회원, "CsI(TI) Indirect Flat Panel Detector의 선질에 따른 물리적 영상 평가", 방사선기술과학, 35(2), pp.109~117 (2012), 대한방사선과학회[DOI]
  11. 김기원, "유방영상에서 엣지를 이용한 물리적 영상 평가", 방사선기술과학, 39(3), pp.415~420 (2016), 대한방사선과학회[DOI]
  12. 손순룡, "Measurement of Image Quality According to the Time of Computed Radiography System", 방사선기술과학, 38(4), pp.365~374 (2015), 대한방사선과학회[DOI]
  13. 김기원, "컴퓨터단층검사에서 AAPM Phantom을 이용한 변조전달함수 평가", 방사선기술과학, 39(2), pp.193~198 (2016), 대한방사선과학회[DOI]
  14. Mohapatra SM, "Transfer function measurement and analysis for magnetic resonance imager", Med Phys, 18(6), pp.1141~1144 (1991)
  15. Fujita H, "A simple method for determining the modulation transfer function in digital radiography", IEEE Trans Med Imaging, 11(1), pp.34~39 (1992)
  16. Samei E, "A method for measuring the presampled MTF of digital radiographic systems using an edge test device", Medical Physics, 25, pp.102~113 (1998)
  17. Greer PB, "Evaluation of an algorithm for the assessment of the MTF using an edge method", Medical Physics, 27, pp.2048~2059 (2000)
  18. "Medical electrical equipment Characteristics of digital X-ray imaging devices Part 1: Determination of the detective quantum efficiency", International Electrotechnical Commission 62220-1, ~ (2003)
  19. Och JG, "Acceptance testing of magnetic resonance imaging systems: Report of AAPM Nuclear Magnetic Resonance Task Group No. 6", Med Phys, 19, pp.217~229 (1992)
  20. American College of Radiology, "Phantom test guidance for the ACR MRI accreditation program" (2000), American College of Radiology
  21. American College of Radiology, "MRI quality control manual" (2004), American College of Radiology
  22. 최영재, "유속신호증강효과의 자기공명혈관조영술을 이용한 뇌혈관검사에서 Half Scan Factor 적용한 영상 평가", 한국자기학회지, 26(3), pp.92~98 (2016), 한국자기학회

관련 논문AI추천

자기공명영상에서 ACR 팬텀을 이용한 잡음전력스펙트럼 평가민정환, 정회원대한방사선과학회 · Journal of Radiological Science and Technology2024 · 제47권 제1호3.0T 자기공명영상에서 ACR 팬텀을 이용한 품질평가민정환, 이수민, 정회원대한방사선과학회 · Journal of Radiological Science and Technology2025 · 제48권 제1호3.0T 자기공명영상을 이용한 해상력 특성에 대한 연구김기원, 김현수, 민정환, 손진현, 이시내, 정회원, 한송이, 한지현대한방사선과학회 · Journal of Radiological Science and Technology2020 · 제43권 제4호자기공명영상 화질 평가용 팬텀 개발에 관한 연구강현수, 김광기, 김동성, 김인수, 나동규, 변재호, 송인찬, 안창범, 이정휘, 장기원, 홍석주대한자기공명의과학회 · Investigative Magnetic Resonance Imaging2006 · 제10권 제2호A Novel Magnetic Resonance Quality Assurance Phantom (KMRP-4): Multi-Site Comparison With the American College of Radiology PhantomChang-Hyun Oh, Chulhyun Lee, Jang Gyu Cha, Ji Young Hwang, Jong-Min Kim, Seung Eun Jung, Suchit Kumar대한자기공명의과학회 · Investigative Magnetic Resonance Imaging2023 · 제27권 제1호1.5T 자기공명영상을 이용한 물리적 영상 특성에 대한 연구김기원, 김현수, 민정환, 박장호, 이시내, 정회원, 한지현대한방사선과학회 · Journal of Radiological Science and Technology2019 · 제42권 제5호

이 간행물의 다른 논문 (7건)