Reliability and Validity of Measurement Using Smartphone ­Based Goniometer of Tibial External Rotation Angle in Standing Knee Flexion

2

조회

0

다운로드

2

인용

0

참고문헌

키워드

주제분류물리치료학

초록

The purpose of this study was to assess the intra-rater test-retest reliability of tibial external rotation angle measurement using a smartphone-based photographic goniometer, DrGoniometer (DrG) compared to a three-dimensional motion analysis system (Vicon). The current study showed an interchangeable method using DrG to measure the tibial external rotation angle in standing knee flexion at 90˚. Twelve healthy subjects participated in this study. A rest session was conducted 30 minutes later for within-day reliability and five days later for between-day intra-rater test-retest reliability. To assess the validity of the measurement using DrG, we used a three dimensional motion analysis system as a gold standard to measure the angle of tibial external rotation. Intra-class correlation coefficient (ICC) and the standard error of measurement (SEM) values were used to determine the within- and between- day intra-rater test-retest reliability of using DrG and a three dimensional motion analysis system. To assess validity, Pearson correlation coefficients were used for two measurement techniques. The measurement for tibial external rotation had high intra-rater test-retest reliability of within-day (ICC=.88) and between-day (ICC=.83) reliability using DrG and of within-day (ICC=.93) and between-day (ICC=.77) reliability using a three-dimentional motion analysis system. Tibial external rotation angle measurement using DrG was highly correlated with those of the three-dimensional motion analysis system (r=.86). These results represented that the tibial external rotation angle measurement using DrG showed acceptable reliability and validity compared with the use of three-dimensional motion analysis system.

Abstract

The purpose of this study was to assess the intra-rater test-retest reliability of tibial external rotation angle measurement using a smartphone-based photographic goniometer, DrGoniometer (DrG) compared to a three-dimensional motion analysis system (Vicon). The current study showed an interchangeable method using DrG to measure the tibial external rotation angle in standing knee flexion at 90˚. Twelve healthy subjects participated in this study. A rest session was conducted 30 minutes later for within-day reliability and five days later for between-day intra-rater test-retest reliability. To assess the validity of the measurement using DrG, we used a three dimensional motion analysis system as a gold standard to measure the angle of tibial external rotation. Intra-class correlation coefficient (ICC) and the standard error of measurement (SEM) values were used to determine the within- and between- day intra-rater test-retest reliability of using DrG and a three dimensional motion analysis system. To assess validity, Pearson correlation coefficients were used for two measurement techniques. The measurement for tibial external rotation had high intra-rater test-retest reliability of within-day (ICC=.88) and between-day (ICC=.83) reliability using DrG and of within-day (ICC=.93) and between-day (ICC=.77) reliability using a three-dimentional motion analysis system. Tibial external rotation angle measurement using DrG was highly correlated with those of the three-dimensional motion analysis system (r=.86). These results represented that the tibial external rotation angle measurement using DrG showed acceptable reliability and validity compared with the use of three-dimensional motion analysis system.

참고문헌 (34)

  1. Alam M, "Measurement of rotational laxity of the knee: In vitro comparison of accuracy between the tibia, overlying skin, and foot", Am J Sports Med, 39(12), pp.2575~2581 (2011)
  2. Alanay A, "Reliability and validity of adapted turkish version of scoliosis research society-22 (SRS-22) questionnaire", Spine (Phila Pa 1976), 30(21), pp.2464~2468 (2005)
  3. Al-Hadithy N, "Smartphones in orthopaedics", Int Orthop, 36(8), pp.1543~1547 (2012)
  4. Atkinson G, "Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine", Sports Med, 26(4), pp.217~238 (1998)
  5. Brochard S, "Accuracy and reliability of three methods of recording scapular motion using reflective skin markers", Proc Inst Mech Eng H, 225(1), pp.100~105 (2011)
  6. Bruce WD, "Surgical correction of miserable malalignment syndrome", J Pediatr Orthop, 24(4), pp.392~396 (2004)
  7. Costigan PA, "Semiautomatic three-dimensional knee motion assessment system", Med Biol Eng Comput, 30(3), pp.343~350 (1992)
  8. Crossley KM, "Analysis of outcome measures for persons with patellofemoral pain: Which are reliable and valid?", Arch Phys Med Rehabil, 85(5), pp.815~822 (2004)
  9. Dunlevy C, "Procedural considerations for photographic-based joint angle measurements", Physiother Res Int, 10(4), pp.190~200 (2005)
  10. Eckhoff DG, "Knee version associated with anterior knee pain", Clin Orthop Relat Res, (339), pp.152~155 (1997)
  11. Ege T, "Use of the iPhone for radiographic evaluation of hallux valgus", Skeletal Radiol, 42(2), pp.269~273 (2013)
  12. Ferber R, "A comparison of within- and between-day reliability of discrete 3D lower extremity variables in runners", J Orthop Res, 20, pp.1139~1145 (2002)
  13. Ferriero G, "Reliability of a new application for smartphones (DrGoniometer) for elbow angle measurement", PM R, 3(12), pp.1153~1154 (2011)
  14. Ferriero G, "Reliability of a smartphone-based goniometer for knee joint goniometry", Int J Rehabil Res, 36(2), pp.146~151 (2013)
  15. Ford KR, "Reliability of landing 3D motion analysis: Implications for longitudinal analyses", Med Sci Sports Exerc, 39(11), pp.2021~2028 (2007)
  16. Fouilleron N, "Proximal tibial derotation osteotomy for torsional tibial deformities generating patello-femoral disorders", Orthop Traumatol Surg Res, 96(7), pp.785~792 (2010)
  17. Jacquot F, "Measuring the cobb angle with the iPhone in kyphoses: A reliability study", Int Orthop, 36(8), pp.1655~1660 (2012)
  18. Kadaba MP, "Repeatability of kinematic, kinetic, and electromyographic data in normal adult gait", J Orthop Res, 7(6), pp.849~860 (1989)
  19. Lee RY, "A real-time gyroscopic system for three-dimensional measurement of lumbar spine motion", Med Eng Phys, 25(10), pp.817~824 (2003)
  20. Lindqvist B, "A three-dimensional photographic method for documentation and measurement of dental conditions", J Orofac Orthop, 59(2), pp.90~99 (1998)
  21. Mikashima Y, "Clinical results of isolated reconstruction of the medial patellofemoral ligament for recurrent dislocation and subluxation of the patella", Acta Orthop Belg, 72(1), pp.65~71 (2006)
  22. Osternig LR, "Patterns of tibial rotary torque in knees of healthy subjects", Med Sci Sports Exerc, 12(3), pp.195~199 (1980)
  23. Qiao J, "Reliability analysis of a smartphone-aided measurement method for the cobb angle of scoliosis", J Spinal Disord Tech, 25(4), pp.E88~E92 (2012)
  24. Queen RM, "Repeatability of lower extremity kinetics and kinematics for standardized and self-selected running speeds", Gait Posture, 23(3), pp.282~287 (2006)
  25. Sahrmann SA, "Movement System Impairment Syndromes of the Extremities, Cervical and Thoracic Spines", pp.424~425 (2011), Mosby
  26. Tashman S, "Abnormal rotational knee motion during running after anterior cruciate ligament reconstruction", Am J Sports Med, 32(4), pp.975~983 (2004)
  27. Taussig G, "Tibial torsion", Rev Chir Orthop Reparatrice Appar Mot, 68(1), pp.20~23 (1982)
  28. Turner MS, "The association between tibial torsion and knee joint pathology", Clin Orthop Relat Res, (302), pp.47~51 (1994)
  29. Turner MS, "The effect of tibial torsion of the pathology of the knee", J Bone Joint Surg Br, 63-B(3), pp.396~398 (1981)
  30. von Porat A, "Knee kinematics and kinetics during gait, step and hop in males with a 16 year old ACL injury compared with matched controls", Knee Surg Sports Traumatol Arthrosc, 14(6), pp.546~554 (2006)
  31. von Porat A, "Reliability and validity of videotaped functional performance tests in ACL-injured subjects", Physiother Res Int, 13(2), pp.119~130 (2008)
  32. Vrtovec T, "A review of methods for quantitative evaluation of spinal curvature", Eur Spine J, 18(5), pp.593~607 (2009)
  33. Webster KE, "Three dimensional motion analysis of within and between day repeatability of tibial rotation during pivoting", Knee, 17(5), pp.329~333 (2010)
  34. Wilkerson GB, "Biomechanical and neuromuscular effects of ankle taping and bracing", J Athl Train, 37(4), pp.436~445 (2002)

관련 논문AI추천

이 간행물의 다른 논문 (7건)