Primary implant stability in a bone model simulating clinical situations for the posterior maxilla: an in vitro study

1

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류치의학

초록

Purpose: The aim of this study was to determine the influence of anatomical conditions on primary stability in the models simulating posterior maxilla. Methods: Polyurethane blocks were designed to simulate monocortical (M) and bicortical (B) conditions. Each condition had four subgroups measuring 3 mm (M3, B3), 5 mm (M5, B5), 8 mm (M8, B8), and 12 mm (M12, B12) in residual bone height (RBH). After implant placement, the implant stability quotient (ISQ), Periotest value (PTV), insertion torque (IT), and reverse torque (RT) were measured. Two-factor ANOVA (two cortical conditions×four RBHs) and additional analyses for simple main effects were performed. Results: A significant interaction between cortical condition and RBH was demonstrated for all methods measuring stability with two-factor ANOVA. In the analyses for simple main effects, ISQ and PTV were statistically higher in the bicortical groups than the corresponding monocortical groups, respectively. In the monocortical group, ISQ and PTV showed a statistically significant rise with increasing RBH. Measurements of IT and RT showed a similar tendency, measuring highest in the M3 group, followed by the M8, the M5, and the M12 groups. In the bicortical group, all variables showed a similar tendency, with different degrees of rise and decline. The B8 group showed the highest values, followed by the B12, the B5, and the B3 groups. The highest coefficient was demonstrated between ISQ and PTV. Conclusions: Primary stability was enhanced by the presence of bicortex and increased RBH, which may be better demonstrated by ISQ and PTV than by IT and RT.

Abstract

Purpose: The aim of this study was to determine the influence of anatomical conditions on primary stability in the models simulating posterior maxilla. Methods: Polyurethane blocks were designed to simulate monocortical (M) and bicortical (B) conditions. Each condition had four subgroups measuring 3 mm (M3, B3), 5 mm (M5, B5), 8 mm (M8, B8), and 12 mm (M12, B12) in residual bone height (RBH). After implant placement, the implant stability quotient (ISQ), Periotest value (PTV), insertion torque (IT), and reverse torque (RT) were measured. Two-factor ANOVA (two cortical conditions×four RBHs) and additional analyses for simple main effects were performed. Results: A significant interaction between cortical condition and RBH was demonstrated for all methods measuring stability with two-factor ANOVA. In the analyses for simple main effects, ISQ and PTV were statistically higher in the bicortical groups than the corresponding monocortical groups, respectively. In the monocortical group, ISQ and PTV showed a statistically significant rise with increasing RBH. Measurements of IT and RT showed a similar tendency, measuring highest in the M3 group, followed by the M8, the M5, and the M12 groups. In the bicortical group, all variables showed a similar tendency, with different degrees of rise and decline. The B8 group showed the highest values, followed by the B12, the B5, and the B3 groups. The highest coefficient was demonstrated between ISQ and PTV. Conclusions: Primary stability was enhanced by the presence of bicortex and increased RBH, which may be better demonstrated by ISQ and PTV than by IT and RT.

참고문헌 (31)

  1. Meredith N, "Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis", Clin Oral Implants Res, 7, pp.261~267 (1996)
  2. Roccuzzo M, "Implant loading protocols for partially edentulous maxillary posterior sites", Int J Oral Maxillofac Implants, 24(Suppl), pp.147~157 (2009)
  3. Sennerby L, "Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implications", Periodontol 2000, 47, pp.51~66 (2008)
  4. Atsumi M, "Methods used to assess implant stability: current status", Int J Oral Maxillofac Implants, 22, pp.743~754 (2007)
  5. Meredith N, "Assessment of implant stability as a prognostic determinant", Int J Prosthodont, 11, pp.491~501 (1998)
  6. Shin SY, "The effects of defect type and depth, and measurement direction on the implant stability quotient value", J Oral Implantol, 41, pp.652~656 (2015)
  7. Brånemark PI, "An experimental and clinical study of osseointegrated implants penetrating the nasal cavity and maxillary sinus", J Oral Maxillofac Surg, 42, pp.497~505 (1984)
  8. Jensen J, "Varying treatment strategies for reconstruction of maxillary atrophy with implants: results in 98 patients", J Oral Maxillofac Surg, 52, pp.210~216 (1994)
  9. Ivanoff CJ, "Influence of mono- and bicortical anchorage on the integration of titanium implants. A study in the rabbit tibia", Int J Oral Maxillofac Surg, 25, pp.229~235 (1996)
  10. Van Oosterwyck H, "The influence of bone mechanical properties and implant fixation upon bone loading around oral implants", Clin Oral Implants Res, 9, pp.407~418 (1998)
  11. Bahat O, "Treatment planning and placement of implants in the posterior maxillae: report of 732 consecutive Nobelpharma implants", Int J Oral Maxillofac Implants, 8, pp.151~161 (1993)
  12. Akça K, "Biomechanical aspects of initial intraosseous stability and implant design: a quantitative micro-morphometric analysis", Clin Oral Implants Res, 17, pp.465~472 (2006)
  13. Schulte W, "Periotest to monitor osseointegration and to check the occlusion in oral implantology", J Oral Implantol, 19, pp.23~32 (1993)
  14. Aparicio C, "Validity and clinical significance of biomechanical testing of implant", bone interface, 2006, pp.17(Suppl 2), 2~7
  15. Devlin H, "A comparison of maxillary and mandibular bone mineral densities", J Prosthet Dent, 79, pp.323~327 (1998)
  16. Raghavendra S, "Early wound healing around endosseous implants: a review of the literature", Int J Oral Maxillofac Implants, 20, pp.425~431 (2005)
  17. Cakur B, "Relationship among Schneiderian membrane, Underwood's septa, and the maxillary sinus inferior border", Clin Implant Dent Relat Res, 15, pp.83~87 (2013)
  18. Miyamoto I, "Influence of cortical bone thickness and implant length on implant stability at the time of surgery--clinical, prospective, biomechanical, and imaging study", Bone, 37, pp.776~780 (2005)
  19. Akkocaoglu M, "Implant design and intraosseous stability of immediately placed implants: a human cadaver study", Clin Oral Implants Res, 16, pp.202~209 (2005)
  20. Nkenke E, "Implant stability and histomorphometry: a correlation study in human cadavers using stepped cylinder implants", Clin Oral Implants Res, 14, pp.601~609 (2003)
  21. Heidari B, "The influence of cortical width of sheep bone on the primary stability of dental implants: an in vitro study", J Long Term Eff Med Implants, 23, pp.87~91 (2013)
  22. 신승윤, "Bone cement grafting increases implant primary stability in circumferential cortical bone defects", Journal of Periodontal & Implant Science, 45(1), pp.30~35 (2015), 대한치주과학회
  23. Ahn SJ, "Differences in implant stability associated with various methods of preparation of the implant bed: an in vitro study", J Prosthet Dent, 107, pp.366~372 (2012)
  24. Yan X, "Comparing the influence of crestal cortical bone and sinus floor cortical bone in posterior maxilla bi-cortical dental implantation: a three-dimensional finite element analysis", Acta Odontol Scand, 73, pp.312~320 (2015)
  25. Hong J, "Quantitative biomechanical analysis of the influence of the cortical bone and implant length on primary stability", Clin Oral Implants Res, 23, pp.1193~1197 (2012)
  26. Friberg B, "Early failures in 4,641 consecutively placed Brånemark dental implants: a study from stage 1 surgery to the connection of completed prostheses", Int J Oral Maxillofac Implants, 6, pp.142~146 (1991)
  27. Jung UW, "Osseointegration of dental implants installed without mechanical engagement: a histometric analysis in dogs", Clin Oral Implants Res, 23, pp.1297~1301 (2012)
  28. Rea M, "Healing of implants installed in over- or under-prepared sites--an experimental study in dogs", Clin Oral Implants Res, 26, pp.442~446 (2015)
  29. Ivanoff CJ, "Influence of bicortical or monocortical anchorage on maxillary implant stability: a 15-year retrospective study of Brånemark System implants", Int J Oral Maxillofac Implants, 15, pp.103~110 (2000)
  30. Krafft T, "Osstell resonance frequency measurement values as a prognostic factor in implant dentistry", J Oral Implantol, 41, pp.e133~e137 (2015)
  31. Molly L, "Bone density and primary stability in implant therapy", Clin Oral Implants Res, 17(Suppl 2), pp.124~135 (2006)

관련 논문AI추천

Evaluation of the correlation between insertion torque and primary stability of dental implants using a block bone test구영, 구기태, 김명덕, 김병국, 김성태, 김태일, 님궁희, 류인철, 설양조, 이용무, 이은희, Dorjpalam Bayarchimeg대한치주과학회 · Journal of Periodontal & Implant Science2013 · 제43권 제1호Primary stability of implants with peri-implant bone defects of various widths: an in vitro investigation허익, 신승윤, 신승일, 임현진, 임현창, 정종혁, 홍지연대한치주과학회 · Journal of Periodontal & Implant Science2019 · 제49권 제1호The effect of various thread designs on the initial stability of taper implants권호범, 김명주, 박주희, 임영준대한치과보철학회 · The Journal of Advanced of Prosthodontics2009 · 제1권 제1호Effect of bone quality and implant surgical technique on implant stability quotient (ISQ) value곽재영, 김성균, 윤홍기, 이수영, 허성주대한치과보철학회 · The Journal of Advanced of Prosthodontics2011 · 제3권 제1호Relationship between cortical bone thickness and implant stability at the time of surgery and secondary stability after osseointegration measured using resonance frequency analysisIrena Sailer, Kenko Tanaka, Kensuke Yamauchi, Nobuhiro Yoda, Ryosuke Iwama, Shinnosuke Nogami, Tetsu Takahashi대한치주과학회 · Journal of Periodontal & Implant Science2018 · 제48권 제6호Maxillary Sinus의 inferior wall의 피질골 두께가 임플란트의 초기안정성에 미치는 영향에 대한 in vitro 연구강봉서, 권택가, 김석규, 양성은, 이철원대한통합치과학회 · 대한통합치과학회지2013 · 제2권 제2호

이 간행물의 다른 논문 (6건)