다양한 소독방법이 PPDO-hybrid-PLGA nanofiber scaffold의 형태와 세포부착에 미치는 영향에 관한 연구
EFFECT OF DIFFERENT STERILIZATION METHODS ON THE SURFACE MORPHOLOGY OF PPDO-hybrid-PLGA NANOFIBER SCAFFOLD AND ATTACHMENTS OF PC12 CELL
2
조회
0
다운로드
1
인용
0
참고문헌
초록
Abstract
Objectives : the effect of different sterilization methods on the surface morphology of PPDO-hybrid-PLGA nanofiber scaffold and attachments of PC12 cell were investigated. Methods : Poly (p-dioxone)-hybrid-Poly (lactide-glycolide) (PPDO-hybrid-PLGA) nanofiber scaffold, fabricated in a tube form with 1.5 mm internal diameter, 0.2 mm thickness and 5 mm length, was prepared using electrospinning method. To study the surface morphology using SEM, The study group and control group in respective were; Control:Non-sterilized scaffold, Group I:scaffold sterilized with 70% Alcohol, Group II: scaffold sterilized with Ethylene Oxide at 65 ℃ , and Group Ⅲ: scaffold sterilized with Ethylene Oxide at 37 ℃. To investigate viability of the PC12 cell on the scaffold, The study group and control group in respective were; Control: sterilized with 70% Alcohol, Group I: sterilized with Ethylene Oxide at 65 ℃, and Group II: sterilized with Ethylene Oxide at 37 ℃. Results : 1. The surface morphology was slightly changed in Group I, II and GroupⅢ, compared with control. 2. The attachment of PC12 cells in Group I, II was not higher than in control Discussion : The attachment of PC12 cell is not influenced by different sterilization methods.
참고문헌 (29)
- 김상헌, "조직공학용 다공성 스캐폴드", 고분자과학과 기술, 16(4), pp.468~477 (2005)
- Park TG, "Degradation of poly(lactic-co-glycolic acid) microspheres: effect of copolymer composition", Biomaterials, 16(15), pp.1123~1130 (1995)
- Chia NK, "Controlled degradation of multilayered poly(lactide-co-glycolide) films using electron beam irradiation", J Biomed Mater Res A, 84(4), pp.980~987 (2008)
- Kang SW, "The effect of microsphere degradation rate on the efficacy of polymeric microspheres as bulking agents: an 18-month follow-up study", J Biomed Mater Res B Appl Biomater, 80(1), pp.253~259 (2007)
- Wu L, "Effects of porosity and pore size on in vitro degradation of three-dimensional porous poly(D,L-lactide-co-glycolide) scaffolds for tissue engineering", J Biomed Mater Res, 75(4), pp.767~777 (2005)
- Lu L, "In vitro degradation of thin poly(DL-lactic-co-glycolic acid) films", J Biomed Mater Re, 46(2), pp.236~244 (1999)
- Watzinger F, "Biodegradable polymer membrane used as septal splint", Int J Oral Maxillofac Surg, 37(5), pp.473~477 (2008)
- Suuronen R, "Biodegradable fracture-fixation devices in maxillofacial surgery", Int J Oral Maxillofac Surg, 22(1), pp.50~57 (1993)
- Kuzin MI, "Surgical absorbable suture materials", Khirurgiia (Mosk), (9), pp.152~157 (1990)
- Suuronen R, "Bioabsorbable plates and screws: Current state of the art in facial fracture repair", J Craniomaxillofac Trauma, 6(1), pp.19~30 (2000)
- Owen GR, "A PLGA membrane controlling cell behaviour for promoting tissue regeneration", Biomaterials, 26(35), pp.7447~7456 (2005)
- Murphy KG, "Guided tissue regeneration for the treatment of periodontal intrabony and furcation defects. A systematic review", Ann Periodontol, 8(1), pp.266~302 (2003)
- de Ruiter GC, "Methods for in vitro characterization of multichannel nerve tubes", J Biomed Mater Res, 84(3), pp.643~651 (2008)
- Lleixa Calvet J, "Sterilization effects on starPEG coated polymer surfaces: characterization and cell viability", J Mater Sci Mater Med, 19(4), pp.1631~1636 (2008)
- Montanari L CM, "Gamma irradiation effects on poly(DL-lactide-co-glycolide) microspheres", J Control Release, 56, pp.219~229 (1998)
- Sintzel MB S-AK, "Influence of irradiation sterilization on a semi-solid poly(ortho ester)", Int J Pharm, 175, pp.165~176 (1998)
- Grayson AC, "Size and temperature effects on poly(lactic-co-glycolic acid) degradation and microreservoir device performance", Biomaterials, 26(14), pp.2137~2145 (2005)
- Loo JS, "Degradation of poly(lactide-co-glycolide) (PLGA) and poly(L-lactide) (PLLA) by electron beam radiation", Biomaterials, 26(12), pp.1359~1367 (2005)
- Holy CE, "Optimizing the sterilization of PLGA scaffolds for use in tissue engineering", Biomaterials, 22(1), pp.25~34 (2001)
- Moorer WR, "Antiviral activity of alcohol for surface disinfection", Int J Dent Hyg, 1(3), pp.138~142 (2003)
- Gardener JF PM, "Introduction to sterilization, disinfection and infection control" (1991), NY:Churchill Livingston
- Calis S, "Influence of irradiation sterilization on poly(lactide-co-glycolide) microspheres containing anti-inflammatory drugs", Farmaco, 57(1), pp.55~62 (2002)
- Carrascosa C, "Effect of gammasterilization process on PLGA microspheres loaded with insulinlike growth factor-I (IGF-I)", J Biomater Appl, 18(2), pp.95~108 (2003)
- Friess W, "Sterilization of gentamicin containing collagen", PLGA microparticle composites, 2006, pp.63(2), 176~187
- Shearer H, "Effects of common sterilization methods on the structure and properties of poly(D,L lactic-co-glycolic acid) scaffolds", Tissue Eng, 12(10), pp.2717~2727 (2006)
- Yaman A, "Alternative methods of terminal sterilization for biologically active macromolecules", Curr Opin Drug Discov Devel, 4(6), pp.760~763 (2001)
- Mendes GC, "Ethylene oxide sterilization of medical devices: a review", Am J Infect Control, 35(9), pp.574~581 (2007)
- Pietrzak WS, "The influence of temperature on the degradation rate of LactoSorb copolymer", J Craniofac, 14(2), pp.176~183 (2003)
- Ernst RR, "Ethylene oxide gaseous sterilization. I. Concentration and temperature effects", Appl Microbiol, 10, pp.337~341 (1962)