와우축을 감싸는 가는 전극을 가진 CI532 기기를 이용한 인공와우 수술의 단기 경험: 기존의 기기와의 비교
Short-Term Experience in Cochlear Implantation with Slim Modiolar Electrode Array (CI532): Comparison to Previous Devices
2
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Background and Objectives There have been no reports in Korea regarding cochlear implantsurgeries using Cochlear Nucleus Profile Slim Modiolar electrode [CI532 (Cochlear Ltd.)],as it has been recently released in Korea. We aimed to investigate the short-term results of CI532and compare them with previous devices with perimodiolar or straight electrodes arrays fromthe same manufacturer. Subjects and Method From August 2018 to July 2019, 52 patients (26 adults; 26 children)who underwent cochlear implantation of CI532 were included. The intraoperative impedanceand evoked compound action potential (ECAP) threshold in each electrode were analyzed andcompared with the devices with a perimodiolar electrode array [Contour Advance® (CochlearLtd.)] and a lateral wall electrode array [CI422 and CI522 (Cochlear Ltd.)]. Postoperative changesof hearing thresholds at each frequency (250, 500, and 1000 Hz) and aided word recognitionscores (WRS) were also compared. Results CI532 showed significantly lower intraoperative impedance in the basal regionscompared to the lateral wall electrode array. The ECAP thresholds of CI532 in the apical electrodeswere significantly lower than that in the other two groups. After implantation, CI532showed a significant preservation of hearing thresholds at most frequencies and showed significantlyhigher preservation rates than the other electrodes. However, there was no differencebetween the three groups regarding the postoperative short-term aided WRS. Conclusion CI532 showed lower intraoperative impedances and ECAP thresholds, and bettershort-term hearing preservation outcomes compared to the other electrodes, suggesting thatCI532 electrode might be a better option with less traumatic insertion. However, there was nosignificant difference in the aided WRS, and further studies with a longer follow-up are necessaryto examine the difference of audiologic outcomes.
Abstract
Background and Objectives There have been no reports in Korea regarding cochlear implantsurgeries using Cochlear Nucleus Profile Slim Modiolar electrode [CI532 (Cochlear Ltd.)],as it has been recently released in Korea. We aimed to investigate the short-term results of CI532and compare them with previous devices with perimodiolar or straight electrodes arrays fromthe same manufacturer. Subjects and Method From August 2018 to July 2019, 52 patients (26 adults; 26 children)who underwent cochlear implantation of CI532 were included. The intraoperative impedanceand evoked compound action potential (ECAP) threshold in each electrode were analyzed andcompared with the devices with a perimodiolar electrode array [Contour Advance® (CochlearLtd.)] and a lateral wall electrode array [CI422 and CI522 (Cochlear Ltd.)]. Postoperative changesof hearing thresholds at each frequency (250, 500, and 1000 Hz) and aided word recognitionscores (WRS) were also compared. Results CI532 showed significantly lower intraoperative impedance in the basal regionscompared to the lateral wall electrode array. The ECAP thresholds of CI532 in the apical electrodeswere significantly lower than that in the other two groups. After implantation, CI532showed a significant preservation of hearing thresholds at most frequencies and showed significantlyhigher preservation rates than the other electrodes. However, there was no differencebetween the three groups regarding the postoperative short-term aided WRS. Conclusion CI532 showed lower intraoperative impedances and ECAP thresholds, and bettershort-term hearing preservation outcomes compared to the other electrodes, suggesting thatCI532 electrode might be a better option with less traumatic insertion. However, there was nosignificant difference in the aided WRS, and further studies with a longer follow-up are necessaryto examine the difference of audiologic outcomes.
참고문헌 (37)
- Hey M, "Objective, audiological and quality of life measures with the CI532Slim Modiolar electrode", Cochlear Implants Int, 20(2), pp.80~90 (2019)
- Lazard DS, "Pre-, per-and postoperative factors affecting performance of postlinguistically deaf adults using cochlear implants : A new conceptual model over time", PLoS One, 7(11), pp.e48739~ (2012)
- 김리석, "인공와우이식 아동의 장기간 청각 수행력의 발달", 대한이비인후-두경부외과학회지, 45(1), pp.18~21 (2002), 대한이비인후과학회
- Kim JR, "The relationship between electrically evoked compound action potential and speech perception in CI24RE implant users", Korean J Otorhinolaryngol-Head Neck Surg, 53(8), pp.470~474 (2010)
- Lehnhardt E, "Intracochlear placement of cochlear implant electrodes in soft surgery technique", HNO, 41(7), pp.356~359 (1993)
- 박세라, "저주파수의 잔존청력이 보존된 부분농 환자에서의 전기 청각 자극 인공와우 1예", 대한이비인후-두경부외과학회지, 55(11), pp.712~716 (2012), 대한이비인후과학회[DOI]
- 장정훈, "와우이식술의 예후인자에 대한 최신지견", 대한이비인후-두경부외과학회지, 57(11), pp.738~747 (2014), 대한이비인후과학회[DOI]
- Jeong J, "Intraindividual comparison of psychophysical parameters between perimodiolar and lateral-type electrode arrays in patients with bilateral cochlear implants", Otol Neurotol, 36(2), pp.228~234 (2015)
- Lathuillière M, "Cone-beam computed tomography in children with cochlear implants : The effect of electrode array position on ECAP", Int J Pediatr Otorhinolaryngol, 92, pp.27~31 (2017)
- Wackym PA, "Electrophysiologic effects of placing cochlear implant electrodes in a perimodiolar position in young children", Laryngoscope, 114(1), pp.71~76 (2004)
- Patrick JF, "The development of the Nucleus Freedom Cochlear implant system", Trends Amplif, 10(4), pp.175~200 (2006)
- Ahn JH, "Facial nerve stimulation after cochlear implantation according to types of Nucleus 24-channel electrode arrays", Acta Otolaryngol, 129(6), pp.588~591 (2009)
- Wanna GB, "Impact of electrode design and surgical approach on scalar location and cochlear implant outcomes", Laryngoscope, 124(Suppl 6), pp.S1~S7 (2014)
- Kim YR, "Characteristics and pathogenesis of facial nerve stimulation after cochlear implant surgeries : A single-centre retrospective analysis from 1151 patients", Clin Otolaryngol, 43(5), pp.1396~1400 (2018)
- Briggs RJ, "Development and evaluation of the modiolar research array--multi-centre collaborative study in human temporal bones", Cochlear Implants Int, 12(3), pp.129~139 (2011)
- Park LR, "Audiological outcomes and map characteristics in children with perimodiolar and slim straight array cochlear implants in opposite ears", Otol Neurotol, 38(9), pp.e320~e326 (2017)
- McJunkin JL, "Early outcomes with a slim, modiolar cochlear implant electrode array", Otol Neurotol, 39(1), pp.e28~e33 (2018)
- van Dijk B, "Clinical results of AutoNRT, a completely automatic ECAP recording system for cochlear implants", Ear Hear, 28(4), pp.558~570 (2007)
- Venail F, "Modeling of auditory neuron response thresholds with cochlear implants", Biomed Res Int, 2015, pp.394687~ (2015)
- Balkany TJ, "Conservation of residual acoustic hearing after cochlear implantation", Otol Neurotol, 27(8), pp.1083~1088 (2006)
- Skarzynski H, "Towards a consensus on a hearing preservation classification system", Acta Otolaryngol Suppl, (564), pp.3~13 (2013)
- Aschendorff A, "Clinical investigation of the Nucleus Slim Modiolar electrode", Audiol Neurootol, 22(3), pp.169~179 (2017)
- Friedmann DR, "Surgical experience and early outcomes with a slim perimodiolar electrode", Otol Neurotol, 40(3), pp.e304~e310 (2019)
- Gordon KA, "From nucleus 24 to 513 : Changing cochlear implant design affects auditory response thresholds", Otol Neurotol, 34(3), pp.436~442 (2013)
- Gordin A, "Evolution of cochlear implant arrays result in changes in behavioral and physiological responses in children", Otol Neurotol, 30(7), pp.908~915 (2009)
- Perenyi A, "Electrophysiological measurements with electrode types of different perimodiolar properties and the same cochlear implant electronics-a retrospective comparison study", J Otolaryngol Head Neck Surg, 48(1), pp.46~ (2019)
- Clark GM, "Cochlear implantation:Osteoneogenesis, electrode-tissue impedance, and residual hearing", Ann Otol Rhinol Laryngol Suppl, 166, pp.40~42 (1995)
- Shaul C, "Electrical impedance as a biomarker for inner ear pathology following lateral wall and peri-modiolar cochlear implantation", Otol Neurotol, 40(5), pp.e518~e526 (2019)
- Kawano A, "Intracochlear factors contributing to psychophysical percepts following cochlear implantation", Acta Otolaryngol, 118(3), pp.313~326 (1998)
- Nadol JB Jr, "Foreign body or hypersensitivity granuloma of the inner ear after cochlear implantation: One possible cause of a soft failure?", Otol Neurotol, 29(8), pp.1076~1084 (2008)
- Cuda D, "Assessment of cochlear trauma and telemetry measures after cochlear implantation : A comparative study between Nucleus®CI512 and CI532 electrode arrays", Audiol Res, 9(1), pp.223~ (2019)
- Skarzynski H, "Hearing Preservation Classification", J Hear Sci, 2(2), pp.95~96 (2012)
- Ramos-Macías A, "Hearing preservation with the Slim Modiolar Electrode Nucleus CI532®cochlear implant : a preliminary experience", Audiol Neurootol, 22(6), pp.317~325 (2017)
- Lenarz T, "Hearing conservation surgery using the Hybrid-L electrode. Results from the first clinical trial at the Medical University of Hannover", Audiol Neurootol, 14(Suppl 1), pp.22~31 (2009)
- Gantz BJ, "Multicenter clinical trial of the Nucleus Hybrid S8 cochlear implant : Final outcomes", Laryngoscope, 126(4), pp.962~973 (2016)
- Roland JT Jr, "United States multicenter clinical trial of the cochlear nucleus hybrid implant system", Laryngoscope, 126(1), pp.175~181 (2016)
- Cuda D, "Cochlear implantation with the Nucleus Slim Modiolar electrode(CI532) : A preliminary experience", Eur Arch Otorhinolaryngol, 274(12), pp.4141~4148 (2017)