Establishment of Age- and Gender-Specific Reference Ranges for 36 Routine and 57 Cell Population Data Items in a New Automated Blood Cell Analyzer, Sysmex XN-2000

2

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류병리학

초록

We established age- and gender-specific reference ranges for the 36 routine complete blood cell (CBC) and 57 cell population data (CPD) items in the Sysmex XN-2000 (Sysmex, Japan). In total, 280 peripheral blood samples were obtained from an equal number of healthy adults. Values for 36 routine items and 57 CPD items were obtained for each sample, and the results were categorized into six subgroups (N>39 in each subgroup) according to patient age (20-40, 41-60, and >60 yr) and gender (male and female), and compared with respect to age and gender differences. The majority of data items (22 of 36 routine CBC items and 44 of 57 CPD items) exhibited significant differences (P≤0.05) in their results with respect to age or gender, and several red cell-, lymphocyte-, and platelet-related data tended to decrease in women or older adults. These results provide a basis for establishing age- and gender-specific reference ranges for routine and CPD items in Sysmex XN-2000. Furthermore, these reference ranges could be used to determine clinical significance for new items of Sysmex XN-2000 in further studies.

Abstract

We established age- and gender-specific reference ranges for the 36 routine complete blood cell (CBC) and 57 cell population data (CPD) items in the Sysmex XN-2000 (Sysmex, Japan). In total, 280 peripheral blood samples were obtained from an equal number of healthy adults. Values for 36 routine items and 57 CPD items were obtained for each sample, and the results were categorized into six subgroups (N>39 in each subgroup) according to patient age (20-40, 41-60, and >60 yr) and gender (male and female), and compared with respect to age and gender differences. The majority of data items (22 of 36 routine CBC items and 44 of 57 CPD items) exhibited significant differences (P≤0.05) in their results with respect to age or gender, and several red cell-, lymphocyte-, and platelet-related data tended to decrease in women or older adults. These results provide a basis for establishing age- and gender-specific reference ranges for routine and CPD items in Sysmex XN-2000. Furthermore, these reference ranges could be used to determine clinical significance for new items of Sysmex XN-2000 in further studies.

참고문헌 (6)

  1. Qiao R, "Complete blood count reference intervals and age- and sex-related trends of North China Han population", Clin Chem Lab Med, 52, pp.1025~1032 (2014)
  2. Hoffmann JJ, "Effect of age and gender on reference intervals of red blood cell distribution width (RDW) and mean red cell volume (MCV)", Clin Chem Lab Med, 53, pp.2015~2019 (2015)
  3. Adeli K, "Complex biological profile of hematologic markers across pediatric, adult, and geriatric ages: establishment of robust pediatric and adult reference intervals on the basis of the Canadian Health Measures Survey", Clin Chem, 61, pp.1075~1086 (2015)
  4. Sehgal KK, "Reference range evaluation of complete blood count parameters with emphasis on newer research parameters on the complete blood count analyzer Sysmex XE-2100", Indian J Pathol Microbiol, 56, pp.120~124 (2013)
  5. Ko YJ, "Reference interval for immature platelet fraction on Sysmex XN hematology analyzer: a comparison study with Sysmex XE-2100", Clin Chem Lab Med, 53, pp.1091~1097 (2015)
  6. Clinical and Laboratory Standards Institute, "Defining, establishing, and verifying reference intervals in the clinical laboratory; approved guideline, Third edition. CLSI document EP28-A3c", Clinical and Laboratory Standards Institute, ~ (2008)

관련 논문AI추천

Complete Blood Count Reference Intervals and Patterns of Changes Across Pediatric, Adult, and Geriatric Ages in Korea조선, 김수영, 나은희, 조한익대한진단검사의학회 · Annals of Laboratory Medicine2018 · 제38권 제6호자동혈구분석기 Sysmex XN-2000의 성능 및 Low WBC Mode를 이용한 백혈구 감별계산 정확성 평가김혜란, 송새암, 신정환, 오승환, 이자영, 이정녀, 전경란대한진단검사의학회 · Laboratory Medicine Online2015 · 제5권 제4호Performance Evaluation of the Mindray BC-6200 Hematology Analyzer; Comparison with Sysmex XE-2100 and Manual Microscopy김한아, 박미경, 이창훈, 이태환, 허미나대한진단검사의학회 · Laboratory Medicine Online2022 · 제12권 제4호Effective and Practical Complete Blood Count Delta Check Method and Criteria for the Quality Control of Automated Hematology AnalyzersCho Young-Uk, Jang Seongsoo, Kim Min-Sun, Kim Seung-Il, Namgoong Seung, Park Chan-Jeoung대한진단검사의학회 · Annals of Laboratory Medicine2023 · 제43권 제5호An Automated Draft Report Generator for Peripheral Blood Smear Examinations Based on Complete Blood Count Parameters고선영, 권정아, 김상욱, 김영곤, 남명현, 문연숙, 박성준, 윤수영, 이현아, 임채승, 장은아대한진단검사의학회 · Annals of Laboratory Medicine2018 · 제38권 제6호Maintenance of Platelet Counts with Low Level QC Materials and the Change in P-LCR according to Hemolysis with XN-9000심문정, Hyun-A Lee대한임상검사과학회 · 대한임상검사과학회지2018 · 제50권 제4호

이 간행물의 다른 논문 (18건)