The change of rotational freedom following different insertion torques in three implant systems with implant driver

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

STATEMENT OF PROBLEM. Implant drivers are getting popular in clinical dentistry. Unlike to implant systems with external hex connection, implant drivers directly engage the implant/abutment interface. The deformation of the implant/abutment interface can be introduced while placing an implant with its implant driver in clinical situations. PURPOSE. This study evaluated the change of rotational freedom between an implant and its abutment after application of different insertion torques. MATERIAL AND METHODS. Three kinds of internal connection implants were utilized for the current study (4.5 × 12 mm Xive, 4.3 × 11.5 mm Inplant Magicgrip, 4.3 × 12 mm Implantium MF). An EstheticBase, a 2-piece top, a Dual abutment was used for its corresponding implant system. The rotational freedom between an implant and its abutment were measured before and after applying 45, 100 Ncm insertion torque. Repeated measures ANOVA was used for statistical analysis. RESULTS. Under 45 Ncm insertion torque, the rotational freedom between an implant and its abutment was significantly increased in Xive (P = .003). However, no significant change was noted in Inplant Magicgrip and Implantium MF. Under 100 Ncm torque, both in Xive (P = .0005) and Implatium MF (P = .03) resulted in significantly increased rotational freedom between the implant and its abutment. DISCUSSION. The design of the implant/implant driver interface effectively prevented the deformation of implant/abutment interface. Little change was noted in the rotational freedom between an implant and its abutment, even though the insertion torque was far beyond clinical application. CONCLUSIONS. The implant/abutment joint of internally connecting implants were quite stable under insertion torque in clinical situation.

Abstract

STATEMENT OF PROBLEM. Implant drivers are getting popular in clinical dentistry. Unlike to implant systems with external hex connection, implant drivers directly engage the implant/abutment interface. The deformation of the implant/abutment interface can be introduced while placing an implant with its implant driver in clinical situations. PURPOSE. This study evaluated the change of rotational freedom between an implant and its abutment after application of different insertion torques. MATERIAL AND METHODS. Three kinds of internal connection implants were utilized for the current study (4.5 × 12 mm Xive, 4.3 × 11.5 mm Inplant Magicgrip, 4.3 × 12 mm Implantium MF). An EstheticBase, a 2-piece top, a Dual abutment was used for its corresponding implant system. The rotational freedom between an implant and its abutment were measured before and after applying 45, 100 Ncm insertion torque. Repeated measures ANOVA was used for statistical analysis. RESULTS. Under 45 Ncm insertion torque, the rotational freedom between an implant and its abutment was significantly increased in Xive (P = .003). However, no significant change was noted in Inplant Magicgrip and Implantium MF. Under 100 Ncm torque, both in Xive (P = .0005) and Implatium MF (P = .03) resulted in significantly increased rotational freedom between the implant and its abutment. DISCUSSION. The design of the implant/implant driver interface effectively prevented the deformation of implant/abutment interface. Little change was noted in the rotational freedom between an implant and its abutment, even though the insertion torque was far beyond clinical application. CONCLUSIONS. The implant/abutment joint of internally connecting implants were quite stable under insertion torque in clinical situation.

참고문헌 (13)

  1. English CE, "Externally hexed implants, abutments, and transfer devices: a comprehensive overview", Implant Dent, 1, pp.273~282 (1992)
  2. Widmark G, "Mk III: a third generation of the self-tapping Brånemark System implant, including the new Stargrip internal grip design. A 1-year prospective four-center study", Clin Implant Dent Relat Res, 5, pp.273~279 (2003)
  3. Bambini F, "Comparative analysis of deformation of two implant", abutment connection systems during implant insertion. An in vitro study, 2005, pp.54, 129~138
  4. Binon PP, "Evaluation of machining accuracy and consistency of selected implants, standard abutments, and laboratory analogs", Int J Prosthodont, 8, pp.162~178 (1995)
  5. Binon PP, "The effect of implant", abutment hexagonal misfit on screw joint stability, 1996, pp.9, 149~160
  6. Binon PP, "The effect of eliminating implant", abutment rotational misfit on screw joint stability, 1996, pp.9, 511~519
  7. Perriard J, "Fatigue resistance of ITI implant-abutment connectors -- a comparison of the standard cone with a novel internally keyed design", Clin Oral Implants Res, 13, pp.542~549 (2002)
  8. Cehreli MC, "Dynamic fatigue resistance of implant-abutment junction in an internally notched morse-taper oral implant: influence of abutment design", Clin Oral Implants Res, 15, pp.459~465 (2004)
  9. Khraisat A, "Fatigue resistance of two implant", abutment joint designs, 2002, pp.88, 604~610
  10. Merz BR, "Mechanics of the implant-abutment connection: an 8-degree taper compared to a butt joint connection", Int J Oral Maxillofac Implants, 15, pp.519~526 (2000)
  11. Norton MR, "An in vitro evaluation of the strength of an internal conical interface compared to a butt joint interface in implant design", Clin Oral Implants Res, 8, pp.290~298 (1997)
  12. de Barros Carrilho GP, "Comparison of external and internal hex implants’rotational freedom: a pilot study", Int J Prosthodont, 18, pp.165~166 (2005)
  13. Squier RS, "Removal torques of conical, tapered implant abutments: the effects of anodization and reduction of surface area", Int J Oral Maxillofac Implants, 17, pp.24~27 (2002)

관련 논문AI추천

Rotational tolerances of a titanium abutment in the as-received condition and after screw tightening in a conical implant connectionGiuseppe Troiano, Khrystyna Zhurakivska, Luigi Laino, Rosario Prisco대한치과보철학회 · The Journal of Advanced of Prosthodontics2021 · 제13권 제6호Influence of the implant-abutment connection design and diameter on the screw joint stability신현모, 윤미정, 전영찬, 정창모, 허중보, Brian Myung Chang대한치과보철학회 · The Journal of Advanced of Prosthodontics2014 · 제6권 제2호Effect of abutment screw length and cyclic loading on removal torque in external and internal hex implants배지명, 이진한, 조혜원, Hnd Hadi Mohammed대한치과보철학회 · The Journal of Advanced of Prosthodontics2016 · 제8권 제1호The assessment of abutment screw stability between the external and internal hexagonal joint under cyclic loading김성훈, 양재호, 이재봉, 이태식, 한중석대한치과보철학회 · 대한치과보철학회지2008 · 제46권 제6호Comparison of fit accuracy and torque maintenance of zirconia and titanium abutments for internal tri-channel and externalhex implant connectionsElaheh Beyabanaki, Hakimeh Siadat, Marzieh Alikhasi, Niloufar Mousavi대한치과보철학회 · The Journal of Advanced of Prosthodontics2017 · 제9권 제4호Influence of internal connection length on screw loosening in internal connection implants김선재, 김성태, 김지선, 김현철, 문홍석, 박영범, 이재훈, 최현민대한치과보철학회 · 대한치과보철학회지2017 · 제55권 제3호

이 간행물의 다른 논문 (9건)