Sinking and fit of abutment of locking taper implant system

1

조회

0

다운로드

3

인용

0

참고문헌

키워드

초록

STATEMENT OF PROBLEM. Unlike screw-retention type, fixture-abutment retention in Locking taper connection depends on frictional force so it has possibility of abutment to sink. PURPOSE. In this study, Bicon. Implant System, one of the conical internal connection implant system, was used with applying loading force to the abutments connected to the fixture. Then the amount of sinking was measured. MATERIAL AND METHODS. 10 Bicon. implant fixtures were used. First, the abutment was connected to the fixture with finger force. Then it was tapped with a mallet for 3 times and loads of 20 kg corresponding to masticatory force using loading application instrument were applied successively. The abutment state, slightly connected to the fixture without pressure was considered as a reference length, and every new abutment length was measured after each load’s step was added. The amount of abutment sinking (mm) was gained by subtracting the length of abutmentfixture under each loading condition from reference length. RESULTS. It was evident, that the amount of abutment sinking in Bicon. Implant System increased as loads were added. When loads of 20 kg were applied more than 5 - 7 times, sinking stopped at 0.45 ± 0.09 mm. CONCLUSION. Even though locking taper connection type implant shows good adaption to occlusal force, it has potential for abutment sinking as loads are given. When locking taper connection type implant is used, satisfactory loads are recommended for precise abutment location.

Abstract

STATEMENT OF PROBLEM. Unlike screw-retention type, fixture-abutment retention in Locking taper connection depends on frictional force so it has possibility of abutment to sink. PURPOSE. In this study, Bicon. Implant System, one of the conical internal connection implant system, was used with applying loading force to the abutments connected to the fixture. Then the amount of sinking was measured. MATERIAL AND METHODS. 10 Bicon. implant fixtures were used. First, the abutment was connected to the fixture with finger force. Then it was tapped with a mallet for 3 times and loads of 20 kg corresponding to masticatory force using loading application instrument were applied successively. The abutment state, slightly connected to the fixture without pressure was considered as a reference length, and every new abutment length was measured after each load’s step was added. The amount of abutment sinking (mm) was gained by subtracting the length of abutmentfixture under each loading condition from reference length. RESULTS. It was evident, that the amount of abutment sinking in Bicon. Implant System increased as loads were added. When loads of 20 kg were applied more than 5 - 7 times, sinking stopped at 0.45 ± 0.09 mm. CONCLUSION. Even though locking taper connection type implant shows good adaption to occlusal force, it has potential for abutment sinking as loads are given. When locking taper connection type implant is used, satisfactory loads are recommended for precise abutment location.

참고문헌 (16)

  1. Binon PP, "Implants and components: entering the new millennium", Int J Oral Maxillofac Implants, 15, pp.76~94 (2000)
  2. Goodacre CJ, "Clinical complications of osseointegrated implants", J Prosthet Dent, 81, pp.537~552 (1999)
  3. Schwarz MS, "Mechanical complications of dental implants", Clin Oral Implants Res, 11, pp.156~158 (2000)
  4. Jemt T, "Osseointegrated implants for single tooth replacement: a 1-year report from a multicenter prospective study", Int J Oral Maxillofac Implants, 6, pp.29~36 (1991)
  5. Taylor TD, "Prosthodontic problems and limitations associated with osseointegration", J Prosthet Dent, 79, pp.74~78 (1998)
  6. Albrektsson T, "A multicenter report on osseointegrated oral implants", J Prosthet Dent, 60, pp.75~84 (1988)
  7. Keating K, "Connecting abutments to Dental implants “An Engineers perspective”", Irish Dentist, pp.43~46 (2001)
  8. Chapman RJ, "The locking taper attachment for implant abutments: use and reliability", Implant Dent, 5, pp.257~261 (1996)
  9. Gibbs CH, "Occlusal forces during chewing and swallowing as measured by sound transmission", J Prosthet Dent, 46, pp.443~449 (1981)
  10. Robert G. Craig, "Restorative Dental Materials, 6th ed", pp.60~62 (1980), C.V. Mosby Co
  11. Anderson DJ, "Measurement of stress in mastication. I", J Dent Res, 35, pp.664~670 (1956)
  12. Anderson DJ, "Measurement of stress in mastication. II", J Dent Res, 35, pp.671~673 (1956)
  13. Richter EJ, "In vivo vertical forces on implants", Int J Oral Maxillofac Implants, 10, pp.99~108 (1995)
  14. Van Eijden TM, "Three dimensional analysis of human biteforce magnitude and moment", Arch Oral Biol, 36, pp.535~539 (1991)
  15. 이한라, "하중조건에 따른 원추형 내측연결 임플랜트 시스템에서 지대주 침하 및 적합에 관한 연구", 구강회복응용과학지, 24(1), pp.77~90 (2008), 대한턱관절교합학회
  16. Dailey B, "Axial displacement of abutments into implants and implant replicas, with the tapered cone-screw internal connection, as a function of tightening atorque", Int J Oral Maxillofac Implants, 24, pp.251~256 (2009)

관련 논문AI추천

Effect of cyclic loading on axial displacement of abutment into implant with internal tapered connection: a pilot study곽재영, 김성균, 설현우, 한종현, 허성주대한치과보철학회 · 대한치과보철학회지2013 · 제51권 제4호The assessment of abutment screw stability between the external and internal hexagonal joint under cyclic loading김성훈, 양재호, 이재봉, 이태식, 한중석대한치과보철학회 · 대한치과보철학회지2008 · 제46권 제6호Mechanical analysis of conventional and small diameter conical implant abutmentsAdriana Cla′udia Lapria Faria, Izabela Cristina Maurl′cio Moris, Maria da Gloria Chiarello de Mattos, Renata Cristina Silveira Rodrigues, Ricardo Faria Ribeiro대한치과보철학회 · The Journal of Advanced of Prosthodontics2012 · 제4권 제3호Influence of the implant-abutment connection design and diameter on the screw joint stability신현모, 윤미정, 전영찬, 정창모, 허중보, Brian Myung Chang대한치과보철학회 · The Journal of Advanced of Prosthodontics2014 · 제6권 제2호Influence of abutment materials on the implant-abutment joint stability in internal conical connection type implant systems양동석, 윤미정, 정창모, 조재영, 허재찬, 허중보대한치과보철학회 · The Journal of Advanced of Prosthodontics2014 · 제6권 제6호Influence of the implant abutment types and the dynamic loading on initial screw loosening김은숙, 신수연대한치과보철학회 · The Journal of Advanced of Prosthodontics2013 · 제5권 제1호

이 간행물의 다른 논문 (7건)