A comparative study on the initial stability of different implants placed above the bone level using resonance frequency analysis

2

조회

0

다운로드

0

인용

0

참고문헌

초록

PURPOSE. This study evaluated the initial stability of different implants placed above the bone level in different types of bone. MATERIALS AND METHODS. As described by Lekholm and Zarb, cortical layers of bovine bone specimens were trimmed to a thickness of 2 mm, 1 mm or totally removed to reproduce bone types II, III, and IV respectively. Three Implant system (Branemark System(R). Mk III TiUniteTM, Straumann Standard Implant SLA(R)., and Astra Tech MicrothreadTM-OsseoSpeedTM) were tested. Control group implants were placed in level with the bone, while test group implants were placed 1, 2, 3, and 4 mm above the bone level. Initial stability was evaluated by resonance frequency analysis. Data was statistically analyzed by one-way analysis of variance in confidence level of 95%. The effective implant length and the Implant Stability Quotient (ISQ) were compared using simple linear regression analysis. RESULTS. In the control group, there was a significant difference in the ISQ values of the 3 implants in bone types III and IV (P<.05). The ISQ values of each implant decreased with increased effective implant length in all types of bone. In type II bone, the decrease in ISQ value per 1-mm increase in effective implant length of the Branemark and Astra implants was less than that of the Straumann implant. In bone types III and IV, this value in the Astra implant was less than that in the other 2 implants. CONCLUSION. The initial stability was much affected by the implant design in bone types III, IV and the implant design such as the short pitch interval was beneficial to the initial stability of implants placed above the bone level.

Abstract

PURPOSE. This study evaluated the initial stability of different implants placed above the bone level in different types of bone. MATERIALS AND METHODS. As described by Lekholm and Zarb, cortical layers of bovine bone specimens were trimmed to a thickness of 2 mm, 1 mm or totally removed to reproduce bone types II, III, and IV respectively. Three Implant system (Branemark System(R). Mk III TiUniteTM, Straumann Standard Implant SLA(R)., and Astra Tech MicrothreadTM-OsseoSpeedTM) were tested. Control group implants were placed in level with the bone, while test group implants were placed 1, 2, 3, and 4 mm above the bone level. Initial stability was evaluated by resonance frequency analysis. Data was statistically analyzed by one-way analysis of variance in confidence level of 95%. The effective implant length and the Implant Stability Quotient (ISQ) were compared using simple linear regression analysis. RESULTS. In the control group, there was a significant difference in the ISQ values of the 3 implants in bone types III and IV (P<.05). The ISQ values of each implant decreased with increased effective implant length in all types of bone. In type II bone, the decrease in ISQ value per 1-mm increase in effective implant length of the Branemark and Astra implants was less than that of the Straumann implant. In bone types III and IV, this value in the Astra implant was less than that in the other 2 implants. CONCLUSION. The initial stability was much affected by the implant design in bone types III, IV and the implant design such as the short pitch interval was beneficial to the initial stability of implants placed above the bone level.

참고문헌 (30)

  1. Zarb GA, "Osseointegration: A requiem for the periodontal ligament", Int J Periodontic Restorative Dent, 11, pp.88~91 (1991)
  2. Friberg B, "Early failures in 4,641 consecutively placed Bra􀆆nemark dental implants: a study from stage 1 surgery to the connection of completed prostheses", Int J Oral Maxillofac Implants, 6, pp.142~146 (1991)
  3. Jaffin RA, "The excessive loss of Branemark fixtures in type IV bone: a 5-year analysis", J Periodontol, 62, pp.2~4 (1991)
  4. O’Sullivan D, "Measurements comparing the initial stability of five designs of dental implants: a human cadaver study", Clin Implant Dent Relat Res, 2, pp.85~92 (2000)
  5. West JD, "Identification of stability changes for immediately placed dental implants", Int J Oral Maxillofac Implants, 22, pp.623~630 (2007)
  6. Albrektsson T, "Oral implant surfaces: Part 2-review focusing on clinical knowledge of different surfaces", Int J Prosthodont, 17, pp.544~564 (2004)
  7. Ostman PO, "Resonance frequency analysis measurements of implants at placement surgery", Int J Prosthodont, 19, pp.77~83 (2006)
  8. Balshi SF, "A resonance frequency analysis assessment of maxillary and mandibular immediately loaded implants", Int J Oral Maxillofac Implants, 20, pp.584~594 (2005)
  9. Meredith N, "Quantitative determination of the stability of the implant-tissue interface using resonance frequency analysis", Clin Oral Implants Res, 7, pp.261~267 (1996)
  10. "OsstellTM mentor Resonance Frequency Analyser. User`s manual", Integration Diagnostics Ltd, ~
  11. Meredith N, "Resonance frequency measurements of implant stability in vivo. A crosssectional and longitudinal study of resonance frequency measurements on implants in the edentulous and partially dentate maxilla", Clin Oral Implants Res, 8, pp.226~233 (1997)
  12. Adell R, "Surgical procedures in: Tissue integrated prostheses. Osseointegration in clinical dentistry", pp.199~210 (1985), Quintessence
  13. Misch CE, "Contemporary Implant Dentistry", pp.327~354 (1993), CV Mosby
  14. Johansson P, "Assessment of bone quality from cutting resistance duringimplant surgery", Int J Oral Maxillofac Implants, 9, pp.279~288 (1994)
  15. Friberg B, "Evaluation of bone density using cutting resistance measurements and microradiography: an in vitro study in pig ribs", Clin Oral Implants Res, 6, pp.164~171 (1995)
  16. Friberg B, "Identification of bone quality in conjunction with insertion of titanium implants. A pilot study in jaw autopsy specimens", Clin Oral Implants Res, 6, pp.213~219 (1995)
  17. Carr AB, "The relationship of Periotest values, biomaterial, and torque to failure in adult baboons", Int J Prosthodont, 8, pp.15~20 (1995)
  18. Manz M, "An evaluation of the Periotest system. Part I: Examiner reliability and repeatability of readings. Dental Implant Clinical Group (Planning Committee)", Implant Dent, 1, pp.142~146 (1992)
  19. Olive′J, "Periotest method as a measure of osseointegrated oral implant stability", Int J Oral Maxillofac Implants, 5, pp.390~400 (1990)
  20. van Steenberghe D, "Reproducibility and detection thresthold of periimplant diagnostics", Adv Dent Res, 7, pp.191~195 (1993)
  21. van Steenberghe D, "Damping characteristics of bone-to-implant interfaces. A clinical study with the Periotest device", Clin Oral Implants Res, 6, pp.31~39 (1995)
  22. Teerlinck J, "Periotest: an objective clinical diagnosis of bone apposition toward implants", Int J Oral Maxillofac Implants, 6, pp.55~61 (1991)
  23. Schulte W, "Periotest - a new measurement process for periodontal function", Zahna¨rtzl Mitt, 73, pp.1229~1240 (1983)
  24. Derhami K, "Assessment of the periotest device in baseline mobility measurements of craniofacial implants", Int J Oral Maxillofac Implants, 10, pp.221~229 (1995)
  25. Niimi A, "A comparative study of removal torque of endosseous implants in the fibula, iliac crest and scapula of cadavers: preliminary report", Clin Oral Implants Res, 8, pp.286~289 (1997)
  26. Sennerby L, "Resonance frequency analysis: measuring implant stability and osseointegration", Compend Contin Educ Dent, 19, pp.493~498 (1998)
  27. Meredith N, "The application of resonance frequency measurements to study the stability of titanium implants during healing in the rabbit tibia", Clin Oral Implants Res, 8, pp.234~243 (1997)
  28. Meredith N, "Assessment of implant stability as a prognostic determinant", Int J Prosthodont, 11, pp.491~501 (1998)
  29. Analisi della Frequenza di Risonanza (RFA), "Sennerby L", 19~32 (2002)
  30. Zix J, "Stability measurements of 1-stage implants in the maxilla by means of resonance frequency analysis: a pilot study", Int J Oral Maxillofac Implants, 20, pp.747~752 (2005)

관련 논문AI추천

이 간행물의 다른 논문 (8건)