본문 바로가기
KERD
논문KCI 등재

Does matching relation exist between the length and the tilting angle of terminal implants in the all-on-four protocol? stress distributions by 3D finite element analysis

2

조회

0

다운로드

1

인용

0

참고문헌

키워드

초록

PURPOSE. To explore whether there is matching relation between the length and the tilting angle of terminal implants in the All-on-Four protocol by studying the effects of different implant configurations on stress distributions of implant, bone, and framework. MATERIALS AND METHODS. Four implants were employed to support a full-arch fixed prosthesis and five three-dimensional finite element models were established with CT images, based on the length (S and L) and distal tilt angle (0°, 30° and 45°) of terminal implants for an edentulous mandible, which named: Tilt0-S, Tilt30-S, Tilt30-L, Tilt45-S and Tilt45-L. An oblique 240 N was loaded at second molar. The von Mises Stresses were analyzed. The implants were consecutively named #1 to #4 from the loading point. RESULTS. 1) Tilt0-S had the greatest stress on the implants, with the other groups exhibiting variable reductions; the four implants of Tilt45-L demonstrated the greatest reduction in stress. 2) Tilt0-S had the greatest stress at bone around #1 implant neck, and Tilt45-L exhibited the least stress, which was a 36.3% reduction compared to Tilt0-S. 3) The greatest stress in the framework was found on the cantilevers distal to #1 implant. Tilt45-S exhibited the least stress. CONCLUSION. Matching different length and tilting angle of the terminal implants led to variable stress reductions on implants, bone and the superstructure. By optimizing implant configuration, the reduction of stress on implants and surrounding bone could be maximized. Under the present condition, Tilt45-L was the preferred configuration. Further clinical testings are required.

Abstract

PURPOSE. To explore whether there is matching relation between the length and the tilting angle of terminal implants in the All-on-Four protocol by studying the effects of different implant configurations on stress distributions of implant, bone, and framework. MATERIALS AND METHODS. Four implants were employed to support a full-arch fixed prosthesis and five three-dimensional finite element models were established with CT images, based on the length (S and L) and distal tilt angle (0°, 30° and 45°) of terminal implants for an edentulous mandible, which named: Tilt0-S, Tilt30-S, Tilt30-L, Tilt45-S and Tilt45-L. An oblique 240 N was loaded at second molar. The von Mises Stresses were analyzed. The implants were consecutively named #1 to #4 from the loading point. RESULTS. 1) Tilt0-S had the greatest stress on the implants, with the other groups exhibiting variable reductions; the four implants of Tilt45-L demonstrated the greatest reduction in stress. 2) Tilt0-S had the greatest stress at bone around #1 implant neck, and Tilt45-L exhibited the least stress, which was a 36.3% reduction compared to Tilt0-S. 3) The greatest stress in the framework was found on the cantilevers distal to #1 implant. Tilt45-S exhibited the least stress. CONCLUSION. Matching different length and tilting angle of the terminal implants led to variable stress reductions on implants, bone and the superstructure. By optimizing implant configuration, the reduction of stress on implants and surrounding bone could be maximized. Under the present condition, Tilt45-L was the preferred configuration. Further clinical testings are required.

참고문헌 (28)

  1. Strassburger C, "Influence of implant and conventional prostheses on satisfaction and quality of life: A literature review. Part 2: Qualitative analysis and evaluation of the studies", Int J Prosthodont, 19, pp.339~348 (2006)
  2. Adell R, "A 15-year study of osseointegrated implants in the treatment of the edentulous jaw", Int J Oral Surg, 10, pp.387~416 (1981)
  3. Brånemark PI, "Ten-year survival rates of fixed prostheses on four or six implants ad modum Brånemark in full edentulism", Clin Oral Implants Res, 6, pp.227~231 (1995)
  4. Francetti L, "Immediate rehabilitation of the mandible with fixed full prosthesis supported by axial and tilted implants:interim results of a single cohort prospective study", Clin Implant Dent Relat Res, 10, pp.255~263 (2008)
  5. Capelli M, "Immediate rehabilitation of the completely edentulous jaw with fixed prostheses supported by either upright or tilted implants: a multicenter clinical study", Int J Oral Maxillofac Implants, 22, pp.639~644 (2007)
  6. Maló P, "“All-on-Four” immediate-function concept with Brånemark System implants for completely edentulous mandibles: a retrospective clinical study", Clin Implant Dent Relat Res, 5, pp.2~9 (2003)
  7. Shackleton JL, "Survival of fixed implant-supported prostheses related to cantilever lengths", J Prosthet Dent, 71, pp.23~26 (1994)
  8. Correa S, "Evaluation of the structural behavior of three and four implant-supported fixed prosthetic restorations by finite element analysis", J Prosthodont Res, 56, pp.110~119 (2012)
  9. Bellini CM, "Comparison of tilted versus nontilted implant-supported prosthetic designs for the restoration of the edentuous mandible: a biomechanical study", Int J Oral Maxillofac Implants, 24, pp.511~517 (2009)
  10. Skalak R, "Biomechanical consideration in osseointegrated prostheses", J Prosthet Dent, 49, pp.843~848 (1983)
  11. Fazi G, "Three-dimensional finite element analysis of different implant configurations for a mandibular fixed prosthesis", Int J Oral Maxillofac Implants, 26, pp.752~759 (2011)
  12. Naini RB, "Tilted or parallel implant placement in the completely edentulous mandible? A three-dimensional finite element analysis", Int J Oral Maxillofac Implants, 26, pp.776~781 (2011)
  13. Takayama H, "Osseointegration and Occlusal Rehabilitation", pp.265~280 (1990), Quintessence
  14. Stegaroiu R, "Influence of restoration type on stress distribution in bone around implants:a three-dimensional finite element analysis", Int J Oral Maxillofac Implants, 13, pp.82~90 (1998)
  15. Sertgöz A, "Finite element analysis of the effect of cantilever and implant length on stress distribution in an implant-supported fixed prosthesis", J Prosthet Dent, 76, pp.165~169 (1996)
  16. Rodriguez AM, "Cantilever and implant biomechanics: a review of the literature, Part 2", J Prosthodont, 3, pp.114~118 (1994)
  17. Kim KS, "Biomechanical comparison of axial and tilted implants for mandibular full-arch fixed prostheses", Int J Oral Maxillofac Implants, 26, pp.976~984 (2011)
  18. Gita Malathi K, "Mechanical complications with implants and implant prostheses", Indian J Dent Adv, 3, pp.555~558 (2011)
  19. Montero J, "Clinical evaluation of the incidence of prosthetic complications in implant crowns constructed with UCLA castable abutments. A cohort follow-up study", J Dent, 40, pp.1081~1089 (2012)
  20. Brägger U, "Biological and technical complications and failures with fixed partial dentures (FPD) on implants and teeth after four to five years of function", Clin Oral Implants Res, 12, pp.26~34 (2001)
  21. Naert I, "A study of 589 consecutive implants supporting complete fixed prostheses. Part II: Prosthetic aspects", J Prosthet Dent, 68, pp.949~956 (1992)
  22. Pjetursson BE, "A systematic review of the survival and complication rates of fixed partial dentures (FPDs) after an observation period of at least 5 years", Clin Oral Implants Res, 15, pp.625~642 (2004)
  23. Kreissl ME, "Technical complications of implant-supported fixed partial dentures in partially edentulous cases after an average observation period of 5 years", Clin Oral Implants Res, 18, pp.720~726 (2007)
  24. Zampelis A, "Tilting of splinted implants for improved prosthodontic support: a two-dimensional finite element analysis", J Prosthet Dent, 97, pp.S35~S43 (2007)
  25. Sertgöz A, "Finite element analysis study of the effect of superstructure material on stress distribution in an implant-supported fixed prosthesis", Int J Prosthodont, 10, pp.19~27 (1997)
  26. Rubo JH, "Finite-element analysis of stress on dental implant prosthesis", Clin Implant Dent Relat Res, 12, pp.105~113 (2010)
  27. Bevilacqua M, "Three-dimensional finite element analysis of load transmission using different implant inclinations and cantilever lengths", Int J Prosthodont, 21, pp.539~542 (2008)
  28. Korioth TW, "Influence of mandibular superstructure shape on implant stresses during simulated posterior biting", J Prosthet Dent, 82, pp.67~72 (1999)

관련 논문AI추천

Three-dimensional finite element analysis of buccally cantilevered implantsupported prostheses in a severely resorbed mandible권호범, 김명주, 임영준, Ghaith Alom대한치과보철학회 · The Journal of Advanced of Prosthodontics2021 · 제13권 제1호A two-short-implant-supported molar restoration in atrophic posterior maxilla: A finite element analysis박찬진, 송호용, 조리라, 허윤혁대한치과보철학회 · The Journal of Advanced of Prosthodontics2016 · 제8권 제4호Three dimensional finite element analysis of the stress distribution around the mandibular posterior implant during non-working movement according to the amount of cantilever김명래, 김선종, 김현주, 박은진, 박지만대한치과보철학회 · The Journal of Advanced of Prosthodontics2014 · 제6권 제5호Comparative evaluation of peri-implant stress distribution in implant protected occlusion and cuspally loaded occlusion on a 3 unit implant supported fixed partial denture: A 3D finite element analysis studyParamba Hitendrabhai Acharya, Sareen Subhash Duseja, Vilas Valjibhai Patel, Vishal Rajendrabhai Chauhan대한치과보철학회 · The Journal of Advanced of Prosthodontics2021 · 제13권 제2호Finite element analysis of stress distribution around a short implant according to a horizontally reduced bone in the mandible김병옥, 김희중, 이원표, Kim Kyung-Min, Lim Kyeong-Ok, Yang Min-Cheol치의학연구원 · Oral Biology Research2021 · 제45권 제1호Biomechanical investigation of maxillary implant-supported full-arch prostheses produced with different framework materials: a finite elements studyEmre Mumcu, Mirac Berke Topcu Ersöz대한치과보철학회 · The Journal of Advanced of Prosthodontics2022 · 제14권 제6호

이 간행물의 다른 논문 (10건)