본문 바로가기
KERD
논문KCI 등재

Accuracy of a direct drill-guiding system with minimal tolerance of surgical instruments used for implant surgery: a prospective clinical study

2

조회

0

다운로드

4

인용

0

참고문헌

키워드

초록

PURPOSE. A recently introduced direct drill-guiding implant surgery system features minimal tolerance of surgical instruments in the metal sleeve by using shank-modified drills and a sleeve-incorporated stereolithographic guide template. The purpose of this study was to evaluate the accuracy of this new guided surgery system in partially edentulous patients using geometric analyses. MATERIALS AND METHODS. For the study, 21 implants were placed in 11 consecutive patients using the direct drill-guiding implant surgery system. The stereolithographic surgical guide was fabricated using cone-beam computed tomography, digital scanning, computer-aided design and computer-assisted manufacturing, and additive manufacturing processes. After surgery, the positional and angular deviations between planned and placed implants were measured at the abutment level using implant-planning software. The Kruskal-Wallis test and Mann-Whitney U test were used to compare the deviations (α=.05). RESULTS. The mean horizontal deviations were 0.593 mm (SD 0.238) mesiodistally and 0.691 mm (SD 0.344) buccolingually. The mean vertical deviation was 0.925 mm (SD 0.376) occlusogingivally. The vertical deviation was significantly larger than the horizontal deviation (P=.018). The mean angular deviation was 2.024 degrees (SD 0.942) mesiodistally and 2.390 degrees (SD 1.142) buccolingually. CONCLUSION. The direct drill-guiding implant surgery system demonstrates high accuracy in placing implants. Use of the drill shank as the guiding component is an effective way for reducing tolerance.

Abstract

PURPOSE. A recently introduced direct drill-guiding implant surgery system features minimal tolerance of surgical instruments in the metal sleeve by using shank-modified drills and a sleeve-incorporated stereolithographic guide template. The purpose of this study was to evaluate the accuracy of this new guided surgery system in partially edentulous patients using geometric analyses. MATERIALS AND METHODS. For the study, 21 implants were placed in 11 consecutive patients using the direct drill-guiding implant surgery system. The stereolithographic surgical guide was fabricated using cone-beam computed tomography, digital scanning, computer-aided design and computer-assisted manufacturing, and additive manufacturing processes. After surgery, the positional and angular deviations between planned and placed implants were measured at the abutment level using implant-planning software. The Kruskal-Wallis test and Mann-Whitney U test were used to compare the deviations (α=.05). RESULTS. The mean horizontal deviations were 0.593 mm (SD 0.238) mesiodistally and 0.691 mm (SD 0.344) buccolingually. The mean vertical deviation was 0.925 mm (SD 0.376) occlusogingivally. The vertical deviation was significantly larger than the horizontal deviation (P=.018). The mean angular deviation was 2.024 degrees (SD 0.942) mesiodistally and 2.390 degrees (SD 1.142) buccolingually. CONCLUSION. The direct drill-guiding implant surgery system demonstrates high accuracy in placing implants. Use of the drill shank as the guiding component is an effective way for reducing tolerance.

참고문헌 (31)

  1. Garber DA, "Restoration-driven implant placement with restoration-generated site development", Compend Contin Educ Dent, 16, pp.796~ (1995)
  2. Becker CM, "Surgical guide for dental implant placement", J Prosthet Dent, 83, pp.248~251 (2000)
  3. D’Souza KM, "Types of implant surgical guides in dentistry: a review", J Oral Implantol, 38, pp.643~652 (2012)
  4. Marchack CB, "CAD", CAM-guided implant surgery and fabrication of an immediately loaded prosthesis for a partially edentulous patient, 2007, pp.97, 389~394
  5. Park C, "Accuracy of implant placement using precision surgical guides with varying occlusogingival heights: an in vitro study", J Prosthet Dent, 101, pp.372~381 (2009)
  6. 이승록, "Anthropometric analysis of maxillary anterior buccal bone of Korean adults using cone-beam CT", The Journal of Advanced of Prosthodontics, 2(3), pp.92~96 (2010), 대한치과보철학회
  7. Koop R, "Tolerance within the sleeve inserts of different surgical guides for guided implant surgery", Clin Oral Implants Res, 24, pp.630~634 (2013)
  8. Aalam AA, "Functional restoration of implants on the day of surgical placement in the fully edentulous mandible: a case series", Clin Implant Dent Relat Res, 7, pp.10~16 (2005)
  9. Barone A, "Immediate restoration of single implants placed immediately after tooth extraction", J Periodontol, 77, pp.1914~1920 (2006)
  10. Holst S, "Precision for computer-guided implant placement: using 3D planning software and fixed intraoral reference points", J Oral Maxillofac Surg, 65, pp.393~399 (2007)
  11. 육신영, "A survey of the satisfaction of patients who have undergone implant surgery with and without employing a computer-guided implant surgical template", The Journal of Advanced of Prosthodontics, 6(5), pp.395~405 (2014), 대한치과보철학회[DOI]
  12. Stumpel LJ 3rd, "Cast-based guided implant placement: a novel technique", J Prosthet Dent, 100, pp.61~69 (2008)
  13. Noharet R, "Accuracy of implant placement in the posterior maxilla as related to 2 types of surgical guides: a pilot study in the human cadaver", J Prosthet Dent, 112, pp.526~532 (2014)
  14. Cassetta M, "The Influence of the Tolerance between Mechanical Components on the Accuracy of Implants Inserted with a Stereolithographic Surgical Guide: A Retrospective Clinical Study", Clin Implant Dent Relat Res, 17, pp.580~588 (2015)
  15. Van Assche N, "Tolerance within a surgical guide", Clin Oral Implants Res, 21, pp.455~458 (2010)
  16. Widmann G, "Accuracy in computer-aided implant surgery-a review", Int J Oral Maxillofac Implants, 21, pp.305~313 (2006)
  17. D’haese J, "A prospective study on the accuracy of mucosally supported stereolithographic surgical guides in fully edentulous maxillae", Clin Implant Dent Relat Res, 14, pp.293~303 (2012)
  18. Block MS, "Computed tomography-guided surgery:complications associated with scanning, processing, surgery, and prosthetics", J Oral Maxillofac Surg, 67, pp.13~22 (2009)
  19. Vercruyssen M, "Guided surgery: accuracy and efficacy", Periodontol 2000, 66, pp.228~246 (2014)
  20. Cassetta M, "The intrinsic error of a stereolithographic surgical template in implant guided surgery", Int J Oral Maxillofac Surg, 42, pp.264~275 (2013)
  21. Verhamme LM, "A clinically relevant accuracy study of computerplanned implant placement in the edentulous maxilla using mucosa-supported surgical templates", Clin Implant Dent Relat Res, 17, pp.343~352 (2015)
  22. Van de Wiele G, "The accuracy of guided surgery via mucosa-supported stereolithographic surgical templates in the hands of surgeons with little experience", Clin Oral Implants Res, 26, pp.1489~1494 (2015)
  23. Van Assche N, "Accuracy of computer-aided implant placement", Clin Oral Implants Res, 23, pp.112~123 (2012)
  24. Verhamme LM, "A clinically relevant validation method for implant placement after virtual planning", Clin Oral Implants Res, 24, pp.1265~1272 (2013)
  25. 이재호, "An assessment of template-guided implant surgery in terms of accuracy and related factors", The Journal of Advanced of Prosthodontics, 5(4), pp.440~447 (2013), 대한치과보철학회[DOI]
  26. Koyanagi K, "Development and clinical application of a surgical guide for optimal implant placement", J Prosthet Dent, 88, pp.548~552 (2002)
  27. De Kok IJ, "Radiographic stents: integrating treatment planning and implant placement", Dent Clin North Am, 58, pp.181~192 (2014)
  28. Greenstein G, "Using drill stops on twist drills to promote safety and efficiency when creating osteotomies for dental implants", J Am Dent Assoc, 145, pp.371~375 (2014)
  29. Schwarz L, "Autostop drilling device for implant site preparation: in vitro test of eccentric sensor position", Int J Oral Maxillofac Implants, 30, pp.1041~1046 (2015)
  30. Juodzbalys G, "Inferior alveolar nerve injury associated with implant surgery", Clin Oral Implants Res, 24, pp.183~190 (2013)
  31. Nelogi S, "A surgical guide for optimal placement and immediate restoration of implant", J Oral Implantol, 40, pp.623~626 (2014)

관련 논문AI추천

Accuracy of computer-aided template-guided oral implant placement: a prospective clinical studyCarlo Maiorana, Mario Beretta, Pier Paolo Poli대한치주과학회 · Journal of Periodontal & Implant Science2014 · 제44권 제4호An assessment of template-guided implant surgery in terms of accuracy and related factors김명주, 김명진, 김성민, 박지만, 이재호, 이종호대한치과보철학회 · The Journal of Advanced of Prosthodontics2013 · 제5권 제4호Influence of template design on the accuracy of static computer-assisted implant surgery배아란, 홍성진, Hyeonseong Son, Jinmook Chung, Junghyun Park대한치과보철학회 · The Journal of Advanced of Prosthodontics2025 · 제17권 제1호Reliability of a chairside CAD-CAM surgical guide for dental implant surgery on the anterior maxilla: An in vitro study이석원, 이성복, 이유진, Kyung Lhi Kang, Phyo Ei Ei Htay, Sung Ok Hong대한치과보철학회 · The Journal of Advanced of Prosthodontics2023 · 제15권 제5호Effects of a modified surgical protocol on the positional accuracy of dental implants placed using fully guided implant surgery in the partially edentulous posterior ridge with distal extension: a dentiform model study송영우, 유승하, 정의원대한치과보철학회 · The Journal of Advanced of Prosthodontics2024 · 제16권 제1호Clinical Precautions for Implant Placement using Computer-guided Implant Surgical Guide: A Systematic Review배재휘, 이수영대한구강악안면임플란트학회 · 대한구강악안면임플란트학회지2021 · 제25권 제4호

이 간행물의 다른 논문 (12건)