Comparison of removal torques between laser-treated and SLAtreated implant surfaces in rabbit tibiae

2

조회

0

다운로드

1

인용

0

참고문헌

키워드

초록

PURPOSE. The purpose of this study was to compare removal torques and surface topography between laser treated and sandblasted, large-grit, acid-etched (SLA) treated implants. ATERIALS AND METHODS. Laser treated implants (experimental group) and SLA-treated implants (control group) 8 mm in length and 3.4 mm in diameter were inserted into both sides of the tibiae of 12 rabbits. Surface analysis was accomplished using a field emission scanning electron microscope (FE-SEM; Hitachi S-4800; Japan) under ×25, ×150 and ×1,000 magnification. Surface components were analyzed using energy dispersive spectroscopy (EDS). Rabbits were sacrificed after a 6-week healing period. The removal torque was measured using the MGT-12 digital torque meter (Mark-10 Co., Copiague, NY, USA). RESULTS. In the experimental group, the surface analysis showed uniform porous structures under ×25, ×150 and ×1,000 magnification. Pore sizes in the experimental group were 20-40 mm and consisted of numerous small pores, whereas pore sizes in the control group were 0.5-2.0 mm. EDS analysis showed no significant difference between the two groups. The mean removal torque in the lasertreated and the SLA-treated implant groups were 79.4 Ncm (SD = 20.4; range 34.6-104.3 Ncm) and 52.7 Ncm (SD = 17.2; range 18.7-73.8 Ncm), respectively. The removal torque in the laser-treated surface implant group was significantly higher than that in the control group (P=.004). CONCLUSION. In this study, removal torque values were significantly higher for laser-treated surface implants than for SLA-treated surface implants.

Abstract

PURPOSE. The purpose of this study was to compare removal torques and surface topography between laser treated and sandblasted, large-grit, acid-etched (SLA) treated implants. ATERIALS AND METHODS. Laser treated implants (experimental group) and SLA-treated implants (control group) 8 mm in length and 3.4 mm in diameter were inserted into both sides of the tibiae of 12 rabbits. Surface analysis was accomplished using a field emission scanning electron microscope (FE-SEM; Hitachi S-4800; Japan) under ×25, ×150 and ×1,000 magnification. Surface components were analyzed using energy dispersive spectroscopy (EDS). Rabbits were sacrificed after a 6-week healing period. The removal torque was measured using the MGT-12 digital torque meter (Mark-10 Co., Copiague, NY, USA). RESULTS. In the experimental group, the surface analysis showed uniform porous structures under ×25, ×150 and ×1,000 magnification. Pore sizes in the experimental group were 20-40 mm and consisted of numerous small pores, whereas pore sizes in the control group were 0.5-2.0 mm. EDS analysis showed no significant difference between the two groups. The mean removal torque in the lasertreated and the SLA-treated implant groups were 79.4 Ncm (SD = 20.4; range 34.6-104.3 Ncm) and 52.7 Ncm (SD = 17.2; range 18.7-73.8 Ncm), respectively. The removal torque in the laser-treated surface implant group was significantly higher than that in the control group (P=.004). CONCLUSION. In this study, removal torque values were significantly higher for laser-treated surface implants than for SLA-treated surface implants.

참고문헌 (23)

  1. Albrektsson T., "Osseointegrated titanium implants. Requirements for ensuring a long-lasting, direct bone-to-implant anchorage in man", Acta Orthop Scand, 52, pp.155~170 (1981)
  2. Roberts WE., "Osseointegration in oral rehabilitation", pp.163~164 (1993), Quintessence
  3. Carlsson L., "Removal torques for polished and rough titanium implants", Int J Oral Maxillofac Implants, 3, pp.21~24 (1988)
  4. Gotfredsen K., "Histomorphometric and removal torque analysis for TiO2-blasted titanium implants. An experimental study on dogs", Clin Oral Implants Res, 3, pp.77~84 (1992)
  5. Cordioli G., "Removal torque and histomorphometric investigation of 4 different titanium surfaces : an experimental study in the rabbit tibia", Int J Oral Maxillofac Implants, 15, pp.668~674 (2000)
  6. Klokkevold PR., "Osseointegration enhanced by chemical etching of the titanium surface. A torque removal study in the rabbit", Clin Oral Implants Res, 8, pp.442~447 (1997)
  7. Wennerberg A., "A histomorphometric and removal torque study of screw-shaped titanium implants with three different surface topographies", Clin Oral Implants Res, 6, pp.24~30 (1995)
  8. Ferguson SJ., "Biomechanical comparison of different surface modifications for dental implants", Int J Oral Maxillofac Implants, 23, pp.1037~1046 (2008)
  9. Marin C., "The effect of alterations on resorbable blasting media processed implant surfaces on early bone healing : a study in rabbits", Implant Dent, 20, pp.167~177 (2011)
  10. Kang SH., "Comparison of removal torques for lasertreated titanium implants with anodized implants", J Craniofac Surg, 22, pp.1491~1495 (2011)
  11. Hofmann AA., "Progression of human bone ingrowth into porous-coated implants. Rate of bone ingrowth in humans", Acta Orthop Scand, 68, pp.161~166 (1997)
  12. Hall J., "Stimulation of directed bone growth at oxidized titanium implants by macroscopic grooves: an in vivo study", Clin Implant Dent Relat Res, 7, pp.S76~S82 (2005)
  13. Mangano C., "Bone response to modified titanium surface implants in nonhuman primates(Papio ursinus)and humans : histological evaluation", J Oral Implantol, 34, pp.17~24 (2008)
  14. Schwartz Z., "Effect of titanium surface roughness on chondrocyte proliferation, matrix production, and differentiation depends on the state of cell maturation", J Biomed Mater Res, 30, pp.145~155 (1996)
  15. Hallgren C., "An in vivo study of bone response to implants topographically modified by laser micromachining", Biomaterials, 24, pp.701~710 (2003)
  16. Gaggl A., "Scanning electron microscopical analysis of laser-treated titanium implant surfaces--a comparative study", Biomaterials, 21, pp.1067~1073 (2000)
  17. Cho SA., "A removal torque of the laser-treated titanium implants in rabbit tibia", Biomaterials, 24, pp.4859~4863 (2003)
  18. Kang SH., "Comparison of removal torques for lasertreated titanium implants with anodized implants", J Craniofac Surg, 22, pp.1491~1495 (2011)
  19. Palmquist A., "Biomechanical, histological and ultrastructural analyses of laser micro-and nano-structured titanium implant after 6 months in rabbit", J Biomed Mater Res B Appl Biomater, 97, pp.289~298 (2011)
  20. Wie H., "Hot isostatic pressing-processed hydroxyapatite-coated titanium implants : light microscopic and scanning electron microscopy investigations", Int J Oral Maxillofac Implants, 13, pp.837~844 (1998)
  21. Han CH., "Quantitative and qualitative investigations of surface enlarged titanium and titanium alloy implants", Clin Oral Implants Res, 9, pp.1~10 (1998)
  22. Sennerby L., "Early tissue response to titanium implants inserted in rabbit cortical bone. Part I. Light microscopic obserrvations", J Mater Sci Mater Med, 4, pp.240~250 (1993)
  23. Rong M., "The early osseointegration of the laser-treated and acid-etched dental implants surface : an experimental study in rabbits", J Mater Sci Mater Med, 20, pp.1721~1728 (2009)

관련 논문AI추천

Comparison of removal torques between laser-etched and modified sandblasted acid-etched Ti implant surfaces in rabbit tibias박경순, 조성암, Abdel Ghani Ibrahim Al Awamleh대한치과보철학회 · The Journal of Advanced of Prosthodontics2018 · 제10권 제1호Comparison of the removal torque and a histomorphometric evaluation of the RBM treated implants with the RBM followed by laser treated implants: an experimental study in rabbits김은경, Eun Young Park, Hae Ok Sohn영남대학교 의과대학 · Journal of Yeungnam Medical Science2019 · 제36권 제1호The effects of saline soaking on the removal torque of titanium implants in rabbit tibia after 10 days박정현, 조성암대한치과보철학회 · 대한치과보철학회지2019 · 제57권 제4호Comparison of removal torque of dual-acid etched and single-acid etched implants in rabbit tibias김종진, 조성암대한치과보철학회 · 대한치과보철학회지2019 · 제57권 제4호Biomechanical evaluation of dental implants with different surfaces: Removal torque and resonance frequency analysis in rabbits고정우, 김성훈, 양재호, 이재봉, 한중석대한치과보철학회 · The Journal of Advanced of Prosthodontics2009 · 제1권 제2호Effects of dental implant surface treated with sandblasting large grit acid-etching and femtosecond laser on implant stability, marginal bone volume, and histological results in a rabbit model김지현, 김현덕, 손영탁, 이규복, 이재목, 이종훈, 전만식, 조호성, 진명욱, 황준호, 손큰바다, Sm Abu Saleah대한치과보철학회 · The Journal of Advanced of Prosthodontics2025 · 제17권 제2호

이 간행물의 다른 논문 (10건)