초록

PURPOSE. The purpose of this study was to achieve more retention and stability and to delay or prevent screw loosening. MATERIALS AND METHODS. Twenty implants (Implantium 3.4 mm, Dentium, Seoul, Korea) were divided into 2 groups (n = 20). In the first group, an adhesive material was applied around the screw of the abutments (test group). In the second group, the screws are soaked in saliva (control group). All the screws were torqued under 30 N/cm, Then, the samples were gone through a cyclic fatigue loading process. After cyclic loading, we detorqued screws and calculated detorque value. RESULTS. In comparison with the control group, all the implant screws in the test group were smeared with the adhesive material, showing significant higher detorque value. CONCLUSION. There are significantly higher detorque values in the group with adhesive. It is recommended to make biocompatible adhesive to reduce screw loosening.

Abstract

PURPOSE. The purpose of this study was to achieve more retention and stability and to delay or prevent screw loosening. MATERIALS AND METHODS. Twenty implants (Implantium 3.4 mm, Dentium, Seoul, Korea) were divided into 2 groups (n = 20). In the first group, an adhesive material was applied around the screw of the abutments (test group). In the second group, the screws are soaked in saliva (control group). All the screws were torqued under 30 N/cm, Then, the samples were gone through a cyclic fatigue loading process. After cyclic loading, we detorqued screws and calculated detorque value. RESULTS. In comparison with the control group, all the implant screws in the test group were smeared with the adhesive material, showing significant higher detorque value. CONCLUSION. There are significantly higher detorque values in the group with adhesive. It is recommended to make biocompatible adhesive to reduce screw loosening.

참고문헌 (13)

  1. Diez JS, "Influence of diamondlike carbon-coated screws on the implant-abutment interface", Int J Oral Maxillofac Implants, 27, pp.1055~1060 (2012)
  2. Cho SC, "Screw loosening for standard and wide diameter implants in partially edentulous cases: 3- to 7-year longitudinal data", Implant Dent, 13, pp.245~250 (2004)
  3. Jung RE, "A systematic review of the 5-year survival and complication rates of implant-supported single crowns", Clin Oral Implants Res, 19, pp.119~130 (2008)
  4. Siamos G, "Relationship between implant preload and screw loosening on implant-supported prostheses", J Oral Implantol, 28, pp.67~73 (2002)
  5. 조웅래, "Effect of cyclic loading and retightening on reverse torque value in external and internal implants", The Journal of Advanced of Prosthodontics, 7(4), pp.288~293 (2015), 대한치과보철학회[DOI]
  6. Cem Sahin, "Correlation between microleakage and screw loosening at implantabutment connection", The Journal of Advanced of Prosthodontics, 6(1), pp.35~38 (2014), 대한치과보철학회[DOI]
  7. Tsumita M, "Effect of fatigue loading on the screw joint stability of zirconium abutment", J Prosthodont Res, 57, pp.219~223 (2013)
  8. Binon PP, "The effect of implant", abutment hexagonal misfit on screw joint stability, 1996, pp.9, 149~160
  9. Zeno HA, "The effect of tissue entrapment on screw loosening at the implant", abutment interface of external- and internal-connection implants: An in vitro study, 2016, pp.25, 216~223
  10. Ha CY, "The influence of abutment angulation on screw loosening of implants in the anterior maxilla", Int J Oral Maxillofac Implants, 26, pp.45~55 (2011)
  11. Shigley JE, "Standard handbook of machine design", pp.23~26 (1986), McGraw-Hill
  12. Khraisat A, "Effect of lateral cyclic loading on abutment screw loosening of an external hexagon implant system", J Prosthet Dent, 91, pp.326~334 (2004)
  13. Gross M, "Microleakage at the abutment-implant interface of osseointegrated implants: a comparative study", Int J Oral Maxillofac Implants, 14, pp.94~100 (1999)