The effect of the design of the abutment screw on the loosening force

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류치의학

초록

This study aimed to assess the stability of the screw joint by comparing the implant abutment screw with a Morse-tapered design in the head with the existing implant abutment screw with a butt joint design. After the implant was fastened with either a support screw with a Morse-tapered design or an abutment screw with a butt-joint design, the specimens were divided into six groups. The nonloaded specimens were assigned to two groups: Group 1 (Morse taper) and Group 2 (butt joint). The specimens subjected to 100,000 loading cycles were assigned to Groups 3 and 4. Those subjected to 500,000 loading cycles were assigned to Groups 5 and 6. Each abutment screw was tightened to 35 Ncm. For Groups 1 and 2, the removal torque was measured 1 hour after tightening. For the other four groups, prostheses were fabricated and connected, and a cyclic loading of 50 N was applied before measuring the removal torque. The mean removal torques for Groups 1–6 were 34.24±1.67, 27.30±0.74, 31.80±1.61, 27.20±0.93, 32.82±0.90, and 26.84±1.39 Ncm, respectively. Under all loading conditions, the abutment screws with the Morse-tapered design showed a higher mean removal torque than those with the butt-joint design. Implants assembled with abutment screws having a Morse-tapered connection design showed greater stability than those with a conventional butt-joint connection design.

Abstract

This study aimed to assess the stability of the screw joint by comparing the implant abutment screw with a Morse-tapered design in the head with the existing implant abutment screw with a butt joint design. After the implant was fastened with either a support screw with a Morse-tapered design or an abutment screw with a butt-joint design, the specimens were divided into six groups. The nonloaded specimens were assigned to two groups: Group 1 (Morse taper) and Group 2 (butt joint). The specimens subjected to 100,000 loading cycles were assigned to Groups 3 and 4. Those subjected to 500,000 loading cycles were assigned to Groups 5 and 6. Each abutment screw was tightened to 35 Ncm. For Groups 1 and 2, the removal torque was measured 1 hour after tightening. For the other four groups, prostheses were fabricated and connected, and a cyclic loading of 50 N was applied before measuring the removal torque. The mean removal torques for Groups 1–6 were 34.24±1.67, 27.30±0.74, 31.80±1.61, 27.20±0.93, 32.82±0.90, and 26.84±1.39 Ncm, respectively. Under all loading conditions, the abutment screws with the Morse-tapered design showed a higher mean removal torque than those with the butt-joint design. Implants assembled with abutment screws having a Morse-tapered connection design showed greater stability than those with a conventional butt-joint connection design.

관련 논문AI추천

Influence of the implant-abutment connection design and diameter on the screw joint stability신현모, 윤미정, 전영찬, 정창모, 허중보, Brian Myung Chang대한치과보철학회 · The Journal of Advanced of Prosthodontics2014 · 제6권 제2호Screw loosening and changes in removal torque relative to abutment screw length in a dental implant with external abutment connection after oblique cyclic loading이주희, 차현석대한치과보철학회 · The Journal of Advanced of Prosthodontics2018 · 제10권 제6호Influence of internal connection length on screw loosening in internal connection implants김선재, 김성태, 김지선, 김현철, 문홍석, 박영범, 이재훈, 최현민대한치과보철학회 · 대한치과보철학회지2017 · 제55권 제3호헤드 부분에 모스 테이퍼 구조를 갖는 지대주 나사가 임플란트 나사 안정성에 미치는 영향김정선, 김태훈, 김희중, 박현준대한턱관절교합학회 · 구강회복응용과학지2025 · 제41권 제1호The assessment of abutment screw stability between the external and internal hexagonal joint under cyclic loading김성훈, 양재호, 이재봉, 이태식, 한중석대한치과보철학회 · 대한치과보철학회지2008 · 제46권 제6호Effect of abutment screw length and cyclic loading on removal torque in external and internal hex implants배지명, 이진한, 조혜원, Hnd Hadi Mohammed대한치과보철학회 · The Journal of Advanced of Prosthodontics2016 · 제8권 제1호

이 간행물의 다른 논문 (10건)