Comparison of the Accuracy of Noninvasive Hemoglobin Sensor (NBM-200) and Portable Hemoglobinometer (HemoCue) with an Automated Hematology Analyzer (LH500) in Blood Donor Screening

1

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류병리학

초록

Background: The Hb levels of prospective blood donors are usually determined using a finger prick test. A new noninvasive Hb device has the advantage of not causing any sampling pain. The purpose of this study was to evaluate the accuracy of the noninvasive Hb sensor and to compare its measurements with those of a currently used portable hemoglobinometer. Methods: Hb was measured using a noninvasive Hb sensor (NBM-200; OrSense, Israel),a portable hemoglobinometer (HemoCue; HemoCue AB, Sweden), and an automated hematology analyzer (LH500; Beckman Coulter, USA). The correlations between Hb measurements taken by the NBM-200 and HemoCue with those by an automated hematology analyzer were assessed using intraclass correlation coefficients (ICCs). Hb measurements were compared among 3 different Hb level groups. Results: The mean Hb values of 506 blood donors were 14.1 g/dL by the NBM-200, 14.0g/dL by the LH500, and 14.3 g/dL by the HemoCue. The correlation between the LH500and the NBM-200 was substantial (ICC=0.69), while that between the LH500 and the HemoCue agreed almost perfectly (ICC=0.86). Conclusions: The possibility to judge to be eligible for donors who are ineligible to donate was substantial when using NBM-200. Even though the NBM-200 has the apparent advantage of noninvasiveness, its use in pre-screening should be given meticulous attention. Since pre-donation testing is crucial to protecting donors’ health, complete evaluation of the instrument should be performed prior to use.

Abstract

Background: The Hb levels of prospective blood donors are usually determined using a finger prick test. A new noninvasive Hb device has the advantage of not causing any sampling pain. The purpose of this study was to evaluate the accuracy of the noninvasive Hb sensor and to compare its measurements with those of a currently used portable hemoglobinometer. Methods: Hb was measured using a noninvasive Hb sensor (NBM-200; OrSense, Israel),a portable hemoglobinometer (HemoCue; HemoCue AB, Sweden), and an automated hematology analyzer (LH500; Beckman Coulter, USA). The correlations between Hb measurements taken by the NBM-200 and HemoCue with those by an automated hematology analyzer were assessed using intraclass correlation coefficients (ICCs). Hb measurements were compared among 3 different Hb level groups. Results: The mean Hb values of 506 blood donors were 14.1 g/dL by the NBM-200, 14.0g/dL by the LH500, and 14.3 g/dL by the HemoCue. The correlation between the LH500and the NBM-200 was substantial (ICC=0.69), while that between the LH500 and the HemoCue agreed almost perfectly (ICC=0.86). Conclusions: The possibility to judge to be eligible for donors who are ineligible to donate was substantial when using NBM-200. Even though the NBM-200 has the apparent advantage of noninvasiveness, its use in pre-screening should be given meticulous attention. Since pre-donation testing is crucial to protecting donors’ health, complete evaluation of the instrument should be performed prior to use.

참고문헌 (19)

  1. Cable RG, "Hemoglobin determination in blood donors", Transfus Med Rev, 9, pp.131~144 (1995)
  2. Boulton FE, "Improved strategy for screening prospective blood donors for anaemia", Transfus Med, 4, pp.221~225 (1994)
  3. James V, "Statistical analysis of inappropriate results from current Hb screening methods for blood donors", Transfusion, 43, pp.400~404 (2003)
  4. Ross DG, "Evaluation of hemoglobin screening methods in prospective blood donors", Vox Sang, 50, pp.78~80 (1986)
  5. Noiri E, "Pulse total-hemoglobinometer provides accurate noninvasive monitoring", Crit Care Med, 33, pp.2831~2835 (2005)
  6. Neville RG, "Evaluation of portable haemoglobinometer in general practice", Br Med J (Clin Res Ed), 294, pp.1263~1265 (1987)
  7. Lotfi R, "A noninvasive strategy for screening prospective blood donors for anemia", Transfusion, 45, pp.1585~1592 (2005)
  8. Gayat E, "Non-invasive measurement of hemoglobin: assessment of two different point-of-care technologies", PLoS One, 7, pp.e30065~ (2012)
  9. Landis JR, "The measurement of observer agreement for categorical data", Biometrics, 33, pp.159~174 (1977)
  10. Dhingra-Kumar N, "CuSO4 gravimetric method for Hb screening of prospective donors--should it be discarded?", Transfus Med, 7, pp.245~247 (1997)
  11. Sawant RB, "Evaluation of hemoglobin of blood donors deferred by the copper sulphate method for hemoglobin estimation", Transfus Apher Sci, 36, pp.143~148 (2007)
  12. Bhaskaram P, "Validation of hemoglobin estimation using Hemocue", Indian J Pediatr, 70, pp.25~28 (2003)
  13. Neufeld L, "Hemoglobin measured by Hemocue and a reference method in venous and capillary blood: a validation study", Salud Publica Mex, 44, pp.219~227 (2002)
  14. Prakash S, "Utility of HemoCue in estimation of hemoglobin against standard blood cell counter method", J Assoc Physicians India, 47, pp.995~997 (1999)
  15. Gordeuk VR, "Prevention of iron deficiency with carbonyl iron in female blood donors", Transfusion, 30, pp.239~245 (1990)
  16. Cogswell ME, "Innaccuracy of capillary hemoglobin testing for the detection of anemia and iron deficiency during pregnancy", Am J Epidemiol, 147(S5), pp.206~ (1998)
  17. Bahadur S, "Estimation of hemoglobin in blood donors: A comparative study using hemocue and cell counter", Transfus Apher Sci, 43, pp.155~157 (2010)
  18. Conway AM, "Measurement of haemoglobin using single drops of skin puncture blood: is precision acceptable?", J Clin Pathol, 51, pp.248~250 (1998)
  19. Mills AF, "Screening for anaemia: evaluation of a haemoglobinometer", Arch Dis Child, 64, pp.1468~1471 (1989)

관련 논문AI추천

이 간행물의 다른 논문 (13건)