본문 바로가기
KERD
논문KCI 등재

Evaluation of the Automated Immunohematology Analyzer DAYMATE M

2

조회

0

다운로드

2

인용

0

참고문헌

키워드

초록

Background: An automated immunohematology analyzer, DAYMATE M (DAY Medical, Switzerland), has been recently developed. The potential of this analyzer to improve test results has been evaluated. Methods: A total of 300 blood samples from Seoul St. Mary’s hospital and Incheon St. Mary’s hospital were tested for ABO and RhD typing. In addition, 336 antibody screening test (AST) samples and 82 patients treated with hematopoietic stem cell transplantation (HSCT) were included. AST results by DAYMATE M were compared with those obtained by a manual method using DS-Screening II (Bio-Rad Laboratories, Switzerland) and red blood cells from Selectogen (Ortho-Clinical diagnostics Inc., USA). Results: Of the 300 patients enrolled, 87, 73, 79, and 61 had type A, B, O, and AB blood, respectively. The concordance rate was 99.9% for cell typing and 97.0% for serum typing. One discordant case was classified as type B instead of AB, and six discordant serum-typing cases were type A, but classified as type AB. Among the 336 AST samples, the concordance rate was 93.2%. From 136 positive cases, six were discordant. Within the 82 HSCT-treated patients, the concordance rate for ABO blood typing was 92.2%. Among the six discordant cases, DAYMATE M typed four cases as donor type where the standard method typed them as the recipient blood type. Conclusions: The DAYMATE M automated immunohematology analyzer performs reliably for ABO and RhD typing, as well as for ASTs and on samples from patients treated with HSCT.

Abstract

Background: An automated immunohematology analyzer, DAYMATE M (DAY Medical, Switzerland), has been recently developed. The potential of this analyzer to improve test results has been evaluated. Methods: A total of 300 blood samples from Seoul St. Mary’s hospital and Incheon St. Mary’s hospital were tested for ABO and RhD typing. In addition, 336 antibody screening test (AST) samples and 82 patients treated with hematopoietic stem cell transplantation (HSCT) were included. AST results by DAYMATE M were compared with those obtained by a manual method using DS-Screening II (Bio-Rad Laboratories, Switzerland) and red blood cells from Selectogen (Ortho-Clinical diagnostics Inc., USA). Results: Of the 300 patients enrolled, 87, 73, 79, and 61 had type A, B, O, and AB blood, respectively. The concordance rate was 99.9% for cell typing and 97.0% for serum typing. One discordant case was classified as type B instead of AB, and six discordant serum-typing cases were type A, but classified as type AB. Among the 336 AST samples, the concordance rate was 93.2%. From 136 positive cases, six were discordant. Within the 82 HSCT-treated patients, the concordance rate for ABO blood typing was 92.2%. Among the six discordant cases, DAYMATE M typed four cases as donor type where the standard method typed them as the recipient blood type. Conclusions: The DAYMATE M automated immunohematology analyzer performs reliably for ABO and RhD typing, as well as for ASTs and on samples from patients treated with HSCT.

참고문헌 (18)

  1. Roback JD, "An automatable format for accurate immunohematology testing by ow cytometry", Transfusion, 43, pp.918~927 (2003)
  2. Bajpai M, "Automation in Immunohematology", Asian J Transfus Sci, 6, pp.140~144 (2012)
  3. 신소연, "수혈 전 검사 장비인 AutoVue Innova와 Techno TwinStation의 평가", Annals of Laboratory Medicine, 28(3), pp.214~220 (2008), 대한진단검사의학회
  4. Sturgeon P, "Automation of routine bloodtyping procedures", Vox Sang, 8, pp.438~451 (1963)
  5. Lapierre Y, "The gel test : a new way to detect red cell antigen-antibody reactions", Transfusion, 30, pp.109~113 (1990)
  6. Sinor LT, "Advances in solid-phase red cell adherence methods and transfusion serology", Transfus Med Rev, 6, pp.26~31 (1992)
  7. Voak D, "The status of new methods for the detection of red cell agglutination", Transfusion, 39, pp.1037~1040 (1999)
  8. Duguid JK, "New technology in hospital blood banking", J Clin Pathol, 46, pp.585~588 (1993)
  9. Casina TS, "In search of the Holy Grail : comparison of antibody screening methods", Immunohematology, 22, pp.196~202 (2006)
  10. Malomgré W, "Recent and future trends in blood group typing", Anal Bioanal Chem, 393, pp.1443~1451 (2009)
  11. Tanabe K, "Interinstitutional variation in the measurement of anti-A", B antibodies: the Japanese ABO-Incompatible Transplantation Committee survey, 2007, pp.84, S13~S16
  12. Validation Task Force of the International Society of Blood Transfusion Working Party on Information Technology, "ISBT Guidelines for validation of automated systems in blood establishments", Vox Sang, 98, pp.1~19 (2010)
  13. LA33-A, "Validation of automated systems for immunohematological testing before implementation; approved guidelines. I", pp.2010, Clinical and Laboratory Standards Institute
  14. Fung MK, "Technical manual", pp.371~378 (2014), American Association of Blood Banks
  15. 장미애, "ABO 및 RhD 혈액형 검사와 항체 선별검사를 위한 혈액은행 자동화장비 Galileo NEO와 QWALYS-3의 평가", 대한수혈학회지, 25(3), pp.235~242 (2014), 대한수혈학회
  16. 박영춘, "ABO 및 RhD 혈액형 검사와 비예기항체 선별검사를 위한 자동화장비 IH-1000의 평가", 대한수혈학회지, 23(2), pp.127~135 (2012), 대한수혈학회
  17. 조수연, "Comparative Analysis of Clinical Samples Showing Weak Serum Reaction on AutoVue System Causing ABO Blood Typing Discrepancies", Annals of Laboratory Medicine, 37(2), pp.117~123 (2017), 대한진단검사의학회
  18. Morelati F, "Evaluation of a new automated instrument for pretransfusion testing", Transfusion, 38, pp.959~965 (1998)

관련 논문AI추천

ABO-RhD 혈액형 검사와 비예기항체 선별검사를 위한 혈액은행 자동화장비 IH-500과 VISION Max의 평가구선회, 권계철, 김선영, 김지명, 박유미, 임영애, 임진숙대한진단검사의학회 · Laboratory Medicine Online2017 · 제7권 제4호Economic and Workflow Analysis of a Blood Bank Automated System김형회, 신경화, 이은엽, 장철훈대한진단검사의학회 · Annals of Laboratory Medicine2013 · 제33권 제4호Comparative Analysis of Clinical Samples Showing Weak Serum Reaction on AutoVue System Causing ABO Blood Typing Discrepancies김형회, 김혜림, 신경화, 이주미, 이현지, 장철훈, 조수연대한진단검사의학회 · Annals of Laboratory Medicine2017 · 제37권 제2호교차시험 검사를 위한 혈액은행 자동화 장비 ORTHO VISION의 평가구모세, 구선회, 권계철, 김선영, 김지명, 임진숙대한진단검사의학회 · Laboratory Medicine Online2019 · 제9권 제4호Development of Cryopreserved Red Blood Cell Panels as Biological Reference Standards for Performance Evaluation of ABO and D Blood Grouping Reagents고대현, 김신영, 김현옥, 임지향, 조윤경, 최승준대한진단검사의학회 · Laboratory Medicine Online2022 · 제12권 제1호수혈 전 검사 장비인 AutoVue Innova와 Techno TwinStation의 평가고지선, 구선회, 권계철, 박종우, 성지연, 송정훈, 신소연대한진단검사의학회 · Annals of Laboratory Medicine2008 · 제28권 제3호