A novel intravenous (IV) infusion monitoring sensor is presented to measure the drop rate in the drip chamber of an IV infusion set. It is based on a capacitive proximity sensor and detects the variation of the longitudinal electric field induced by the drop falling intothe drip chamber. Unlike the conventional capacitor sensor with two semi-cylindrical conductor plates, the proximity sensor for IV mon- itoring is composed of a pair of conductor rings which are mounted on the outer surface of the drip chamber with a specific gap between them. The characteristics of the proximity sensor for IV monitoring were investigated through three dimensional electrostatic sim-ulations. It showed quite superior performances in comparison with the conventional capacitor sensor. Especially, the proposed prox-imity sensor exhibits consistent sensitivity regardless of its mounting position on the drip chamber, operates normally though the dripchamber is tilted and shows robustness to the changes of the drop size and the drip factor of the IV infusion set. Thus, the proximity sensor for IV monitoring is more suitable for use in actual environment of IV therapy compared with the conventional capacitor sensor.

Abstract

A novel intravenous (IV) infusion monitoring sensor is presented to measure the drop rate in the drip chamber of an IV infusion set. It is based on a capacitive proximity sensor and detects the variation of the longitudinal electric field induced by the drop falling intothe drip chamber. Unlike the conventional capacitor sensor with two semi-cylindrical conductor plates, the proximity sensor for IV mon- itoring is composed of a pair of conductor rings which are mounted on the outer surface of the drip chamber with a specific gap between them. The characteristics of the proximity sensor for IV monitoring were investigated through three dimensional electrostatic sim-ulations. It showed quite superior performances in comparison with the conventional capacitor sensor. Especially, the proposed prox-imity sensor exhibits consistent sensitivity regardless of its mounting position on the drip chamber, operates normally though the dripchamber is tilted and shows robustness to the changes of the drop size and the drip factor of the IV infusion set. Thus, the proximity sensor for IV monitoring is more suitable for use in actual environment of IV therapy compared with the conventional capacitor sensor.

참고문헌 (9)

  1. F. C. Flack, "Variations of drop size in disposable administration sets used for intravenous infusion", J. Clin. Path, 28, pp.510~512 (1975)
  2. 박효순, "수액 소진 알람시스템을 위한 동일평면형 광센서의 연구 및 제작", 센서학회지, 24(2), pp.113~118 (2015), 한국센서학회
  3. V. V. Kamble, "Monitoring of Intravenous Drip Rate", Proceedings of International Conference on Biomedical Engineering (2001)
  4. P. Bustamante, "A new Wireless Sensor for Intravenous Dripping Detection", Int. J. on Adv. In Networks and Services, 3(1&2) (2010)
  5. D. R. Kirkby, "Capacitance sensing drop counter", J. Biomed. Eng, 11(2), pp.166~169 (1989)
  6. P. Bustamante, "A new wireless sensor for intravenous dripping detection", Proceedings of International Conference on Sensor Technologies and Applications (2007)
  7. D. I. Cour, "Drop Size in Disposable Sets for Intravenous Infusion", J. Acta Anaesthesiologica Scandinavica, 9(3), pp.145~154 (1965)
  8. M. Z. Aslam, "A High Resolution Capacitive Sensing System for the Measurement of Water Content in Crude Oil", Sensors, 14, pp.11351~11361 (2014)