본문 바로가기
KERD
논문KCI 등재

유방X선촬영 시 피폭선량 감소를 위한 유방촬영용 차폐복의 유용성에 관한 연구

A Study on the Usefulness of Breast Shielding Apron for Reducing Exposure Dose in Mammography

2

조회

0

다운로드

4

인용

0

참고문헌

키워드

주제분류방사선과학

초록

Mammography, conducted every two years, causes cancer due to regular exposure to radiation while reducing rate of death caused by breast cancer. The study evaluates the effect of breast shielding apron made to shield off scattered radiation that occurs to the breast when the opposite side breast is mammogramed. AGD was measured using ACR phantom, composed of 50% mammary glands and 50% fat, and radiation was measured before and after wearing the apron on the breast when the opposite side of the breast is mammogramed. When CC direction mammography was conducted to a breast, the AGD was 1.84 mGy. When CC direction and MLO direction mammography were done to a breast, the average dose detected from the opposite side breast from four directions(top to bottom and medial to lateral) was 140 μGy with maximum dose of 256 μGy at medial side. After putting on the apron, the dose, caused by scattered radiation, was not detected from any of the four directions. Using of breast shielding apron is expected to minimize the radiation exposure by blocking scattered radiation to the breast shielded, when mammography is done to the opposite side breast.

Abstract

Mammography, conducted every two years, causes cancer due to regular exposure to radiation while reducing rate of death caused by breast cancer. The study evaluates the effect of breast shielding apron made to shield off scattered radiation that occurs to the breast when the opposite side breast is mammogramed. AGD was measured using ACR phantom, composed of 50% mammary glands and 50% fat, and radiation was measured before and after wearing the apron on the breast when the opposite side of the breast is mammogramed. When CC direction mammography was conducted to a breast, the AGD was 1.84 mGy. When CC direction and MLO direction mammography were done to a breast, the average dose detected from the opposite side breast from four directions(top to bottom and medial to lateral) was 140 μGy with maximum dose of 256 μGy at medial side. After putting on the apron, the dose, caused by scattered radiation, was not detected from any of the four directions. Using of breast shielding apron is expected to minimize the radiation exposure by blocking scattered radiation to the breast shielded, when mammography is done to the opposite side breast.

참고문헌 (22)

  1. Jemal A, "Cancer statics, 2007", A Cancer Journal for Clinicias, 57(1), pp.43~66 (2007)
  2. "Breast Cancer Statistics" (2017)[URL]
  3. 김기석, "초기유방암에서 피하유방절제술 후 즉시유방재건술의 조기결과", Annals of Surgical Treatment and Research, 68(4), pp.264~270 (2005), 대한외과학회
  4. "Radiological Protection for Medical Exposure to Ionizing Radiatio n. Safety Standards Series No. RS-G-1.5" (2002)[URL]
  5. Yoon HS, "Nationwide Korean breast cancer data of 2002", Journal of Korean Breast Cancer Society, 7(2), pp.72~83 (2004)
  6. Miler AB, "Canadian national breast screening study:2.breast cancer detection and death rates among women aged 50 to 59 year", CMAJ, 147(10), pp.1477~1488 (1992)
  7. "IARC handbooks of cancer prevention: Breast cancer scre ening" (2002)[URL]
  8. Lauby-Secretan B, "Breast-canc er screening-viewpoint of the IARC working group", New England Journal of Medicine, 372(24), pp.2353~2358 (2015)
  9. 김정수, "건강검진센터의 의료방사선 피폭 품질관리 시스템 구축 운영 경험 보고", 방사선기술과학, 39(1), pp.13~17 (2016), 대한방사선과학회[DOI]
  10. Bouzarjomehri F, "The study of mean glandular dose in mammography in yazd and the factors affecting it", Iran J Radiol, 4, pp.29~35 (2006)
  11. Kim TH, "Dosimetric evaluation of average glandular absorption radiation dose in mammography", Journal of the Korean Radiological Society, 35(6), pp.999~1003 (1996)
  12. Calton RR, "Principles of radiographic image: An Art and a science" (2001), Delmar Thomson Learning
  13. International Commission on Radiological Protection, "The 2007 Recommendations of the International Commission on Radiological Protection", Ann ICRP, 37(2-4) (2007), Elsevier
  14. "Inspection standard Specification" (2002)[URL]
  15. Wu X, "Spectral dependence of glandular tissue dose in screen-film mammography", Radiology, 179(1), pp.143~148 (1991)
  16. Hurley SF, "The benefits and risks of mammographic screening for breast cancer", Epidemiol Rev, 14, pp.101~130 (1992)
  17. Yaffe MJ, "Risk of radiation-induced breast cancer from mammographic screening", Radiology, 258(1), pp.98~105 (2011)
  18. "Do se measurement in mammography; what are we mea suring?"[URL]
  19. 박혜숙, "평균 유선선량 측정방법의 표준화 및 유방 두께, 실질양상과의 상관관계 분석", 의학물리, 20(1), pp.21~29 (2009), 한국의학물리학회
  20. Jamal N, "A study of mean glandular dose during diagnostic mammography in Malaysia and some of the factors affecting it", The British Journal of Radiology, 76(904), pp.238~245 (2003)
  21. 정홍문, "디지털 유방촬영장비에 관한 공간선량율 평가", 한국방사선학회논문지, 9(1), pp.61~65 (2015), 한국방사선학회[DOI]
  22. 홍은애, "유방촬영 시 갑상샘의 피폭선량 경감을 위한 Al Face Block 제작 및 유용성에 관한 연구", 방사선기술과학, 36(1), pp.11~17 (2013), 대한방사선과학회