디지털 방사선 영상의 편차지수를 이용한 의료영상 품질관리에 관한 연구

A Study on Quality Control for Medical Image by Using Deviation Index of Digital Radiology

0

조회

0

다운로드

5

인용

0

참고문헌

키워드

주제분류방사선과학

초록

In a digital radiation system using a Flat Panel Detector, we attempted to the quality control of digital radiography system using the Exposure Index and Deviation Index. Calibration was performed with the radiation quality suggested by the International Electrotechnical Commission, and through an experiment using a phantom, appropriate inspection radiation conditions applicable to medical institutions were selected. The study was conducted using the selected radiation conditions. Through those chest posterior anterior image, information such as examination conditions and exposure index was obtained. The deviation index was derived by analyzing the exposure index based on the target exposure index calculated by the phantom study. As for the analyzed exposure index, 97.1% was distributed within the range of ± 2.0 based on the deviation index. Quality control of medical images should be performed through management of inspection conditions through exposure index and deviation index and management of medical images.

Abstract

In a digital radiation system using a Flat Panel Detector, we attempted to the quality control of digital radiography system using the Exposure Index and Deviation Index. Calibration was performed with the radiation quality suggested by the International Electrotechnical Commission, and through an experiment using a phantom, appropriate inspection radiation conditions applicable to medical institutions were selected. The study was conducted using the selected radiation conditions. Through those chest posterior anterior image, information such as examination conditions and exposure index was obtained. The deviation index was derived by analyzing the exposure index based on the target exposure index calculated by the phantom study. As for the analyzed exposure index, 97.1% was distributed within the range of ± 2.0 based on the deviation index. Quality control of medical images should be performed through management of inspection conditions through exposure index and deviation index and management of medical images.

참고문헌 (20)

  1. Doi K, "Diagnostic image over the last 50 years : Research and development in medical imaging science and technology", Phys Med Biol, 51, pp.R5~27 (2006)
  2. "ICRP Publication 93. Managing patient dose in digital radiology", ~ (2003)
  3. Yoon YS, "Monte Carlo simulation-based feasibility study of a dose-area product meter built into a collimator for diagnostic X-ray", Radiat Prot Dosimetry, 162, pp.421~426 (2014)
  4. Shepard SJ, "An exposure indication for digital radiography : AAPM Task Group 116(Executive Summary)", Med Phys, 36, pp.2898~2914 (2009)
  5. "IEC 60601-1-3, Ed 2.0. Medical e lectrical equipment - Part 1-3: General requirements for basic safety and essential performance – Collateral Standard:Radiation protection in diagnostic X-ray equipment", International Electrotechnical Commission, ~ (2008)
  6. Shepard SJ, "An exposure indication for digital radiography : AAPM Task Group 116(Executive Summary)", Med Phys, 36, pp.2898~2914 (2009)
  7. An exposure indicator for digital radiography, , American Association of Physicists in Medicine, ~, pp.AAPM Report No.116 (2009)
  8. Moore QT, "Image gently : using exposure indicators to improve pediatric digital radiography", Radiol Technol, 84, pp.93~99 (2012)
  9. Yasumatsu S, "Effect of X-ray beam quality on determination of exposure index", Radiol Phys Technol, 9, pp.109~115 (2016)
  10. "IEC 62494-1, Ed 1.0. Medical electrical equipment–Exposure index of digital X-ray imaging systems. Part 1: Definition and requirements for general radiography", EInternational Electrotechnical Commission, ~ (2008)
  11. 정회원, "인체 흉부 모형 팬텀을 이용한 컴퓨터방사선영상에서 노출지수의 적용 사례 연구", 방사선기술과학, 41(6), pp.533~538 (2018), 대한방사선과학회[DOI]
  12. Nakamae M, "Influence of Interest Region and Interest Value Settings on Exposure Index Analysis Using Chest Clinical Images", Japanese Journal of Radiological Technology, 70(11), pp.1250~1257 (2014)
  13. Willis, CE, "Optimization and Quality Control of Computed Radiography", SPIE Medical Imaging VI:PACS Design and Evaluation, 2164, pp.178~185 (1994)
  14. Willis CE, "Objective Measures of Quality Assurance in a Computed Radiography-based Radiology Department", Physics of Medical Imaging, 2432, pp.588~599 (1995)
  15. Chotas HG, "Digital radiography with photostimulable storage phosphors : Control of detector latitude in chest imaging", Invest Radiol, 227, pp.823~828 (1992)
  16. Arreola M, "Management of pediatric radiation dose using Canon digital radiography", Pediatr Radiol, 34, pp.S221~6 (2004)
  17. 정회원, "다양한 디지털 방사선 시스템의 물리적 영상 특성 조사", 방사선기술과학, 40(3), pp.363~370 (2017), 대한방사선과학회[DOI]
  18. 박혜민, "IEC 규정 노출지수를 활용한 디지털 방사선 영상시스템에서의 영상 수용체간 특성평가 (Ⅰ)", 방사선기술과학, 42(4), pp.291~299 (2019), 대한방사선과학회
  19. Ongoing Quality Control in Digital Radiography, , American Association of Physicists in Medicine, ~, pp.AAPM Report No.151 (2015)
  20. 정회원, "CsI(TI) Indirect Flat Panel Detector의 선질에 따른 물리적 영상 평가", 방사선기술과학, 35(2), pp.109~117 (2012), 대한방사선과학회[DOI]

관련 논문AI추천

이 간행물의 다른 논문 (7건)