노출지수를 이용한 요추 X선 촬영의 조사야 유효성 평가

Effect of Field Size on the Clinical Exposure Index for Lumbar Spine X-ray Examination

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

The field size of the lumbar spine X-ray examination, which belongs to the most frequent examination in general radiography, is 5 times wider than the width of the lumbar spine. Exposure index (EI) as per International Electrotechnical Commission has a proportional relationship with the dose incident on the image receptor for clinical protocols in addition to RQA5, which is a calibration beam quality. In this study, the effectiveness of the set field size was evaluated through the change of EI according to the size of field during lumbar spine X-ray examinations. Lumbar anterior- posterior and lateral examinations was performed using a whole-body phantom, and the national average exposure conditions of Korea investigated in 2017 were introduced for the X-ray exposure. As a result of comparing the EI displayed on the console of digital radiography system for the three field size in ① 18 × 36 cm2 ② 25 × 36 cm2 ③ 36 × 36 cm2, the EI values showed a tendency to increase as the field size increased. Since the patient dose, such as organ dose around the lumbar spine, increases as the field size becomes larger, thus, if the EI obtained from the field size at a level that does not interfere with diagnosis is set as a reference, the effectiveness of the field size can be evaluated through the EI displayed on the console when the lumbar spine X-ray examination is conducted.

Abstract

The field size of the lumbar spine X-ray examination, which belongs to the most frequent examination in general radiography, is 5 times wider than the width of the lumbar spine. Exposure index (EI) as per International Electrotechnical Commission has a proportional relationship with the dose incident on the image receptor for clinical protocols in addition to RQA5, which is a calibration beam quality. In this study, the effectiveness of the set field size was evaluated through the change of EI according to the size of field during lumbar spine X-ray examinations. Lumbar anterior- posterior and lateral examinations was performed using a whole-body phantom, and the national average exposure conditions of Korea investigated in 2017 were introduced for the X-ray exposure. As a result of comparing the EI displayed on the console of digital radiography system for the three field size in ① 18 × 36 cm2 ② 25 × 36 cm2 ③ 36 × 36 cm2, the EI values showed a tendency to increase as the field size increased. Since the patient dose, such as organ dose around the lumbar spine, increases as the field size becomes larger, thus, if the EI obtained from the field size at a level that does not interfere with diagnosis is set as a reference, the effectiveness of the field size can be evaluated through the EI displayed on the console when the lumbar spine X-ray examination is conducted.

참고문헌 (15)

  1. The Korean Society of Medical Imaging Technology, "Textbook of radiographic positioning and clinical diagnosis" (2019), Chung-Ku Publisher
  2. Assessment of radiation exposure of Korean popoulation by medical radiation, "Kim KP", Korea Centers for Disease Control and Prevention, ~, pp.Research Report No. 11-1790387-000030-01 (2020)
  3. Zetterberg LG, "Lumbar spine radiography – poor collimation practices after implementation of digital technology", The British Journal of Radiology, 84, pp.566~569 (2011)
  4. Development of the diagnostic reference level of general radiograpy: Twelve area including Brain, chest, pelvis, etc, "Do KH", Korea Centers for Disease Control and Prevention, ~, pp.Research Report No. 11-1352159-000916-01 (2017)
  5. Zhuo SH, "Geometrical dimensions of the lower lumbar vertebrae – analysis of data from digitised CT images", European Spine Journal, 9, pp.242~248 (2000)
  6. Kenichi K, "Efforts to protect radiation and reduce radiation exposure in FPD systems", Innervision, 27, pp.14~18 (2012)
  7. "IEC 62494-1. Ed 1.0. Medical electrical equipment – Exposure index of digital X-ray imaging system-Part 1: Definitions and requirements for general radiography", International Electrotechnical Commission, ~ (2008)
  8. 박혜민, "IEC 규정 노출지수를 활용한 디지털 방사선 영상시스템에서의 영상 수용체간 특성평가 (Ⅰ)", 방사선기술과학, 42(4), pp.291~299 (2019), 대한방사선과학회
  9. Park HM, "Use of clinical exposure index and deviation index based on national diagnostic reference level as dose-optimization tools for general radiography in Korea", Radiation Protection Dosimetry, 191(4), pp.439~451 (2020)
  10. 정회원, "인체 흉부 모형 팬텀을 이용한 컴퓨터방사선영상에서 노출지수의 적용 사례 연구", 방사선기술과학, 41(6), pp.533~538 (2018), 대한방사선과학회[DOI]
  11. Park HM, "Feasibility of displayed exposure index in IEC standard framework as a dose optimisation tool for digital radiography systems", Radiation Protection Dosimetry, 189(3), pp.384~394 (2020)
  12. 정회원, "디지털 방사선 영상의 편차지수를 이용한 의료영상 품질관리에 관한 연구", 방사선기술과학, 43(2), pp.115~121 (2020), 대한방사선과학회
  13. Zhang M, "Optimization of the radiological protection of patients undergoing digital radiography", Journal of Digital Imaging, 25, pp.196~200 (2012)
  14. 김재겸, "유리선량계를 이용한 요추검사의 장기선량 및 영상의 평가", 방사선기술과학, 39(1), pp.1~11 (2016), 대한방사선과학회[DOI]
  15. Yoon YS, "Proper management of the clinical exposure index based on body thickness using dose optimization tools in digital chest radiography: A phantom stduy", International Journal of Environmental Research and Public Health, 18(10), pp.5203~ (2021)

관련 논문AI추천

이 간행물의 다른 논문 (12건)