PET/CT에서 Phantom을 이용한 Fluorine-18, Gallium-68 방사성 핵종의 PET 영상 평가

Evaluation of PET Image for Fluorine-18 and Gallium-68 using Phantom in PET/CT

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

The purpose of this study is to compare PET imaging performance with Fluorine-18 (18F) and Gallium-68 (68Ga) for influence of physical properties of PET tracer. Measurement were performed on a Siemens Biograph mCT64 PET/CT scanner using NEMA IEC body phantom and Flangeless Esser PET phantom containing filled with 18F and 68Ga. Emission scan duration(ESD) was set to 1, 2, 3, 4 and 5min/bed for 68Ga and 1min/bed for 18F. The PET image were evaluated in terms of contrast, spatial resolution. Under same condition, The percentage of contrast recovery measured in the phantom ranged from 16.88% to 72.56% for 68Ga and from 27.51% to 74.43% for 18F and The FWHM value to evaluate spatial resolution was 10.96 ㎜ for 68Ga and 9.19 ㎜ for 18F. For this study, 18F produces better image contrast and spatial resolution than 68Ga due to higher positron yield and lower positron energy (18F: 96.86%, 633.5 keV, 68Ga: 88.9%, 1899 keV), The physical properties of PET tracer effect on the PET image. 68Ga image applying ESD of 3, 4, 5min/bed were showed similar to 18F image with ESD of 1min/bed. This study suggests that increasing ESD for acquiring 68Ga PET image seem to be similar to 18F image.

Abstract

The purpose of this study is to compare PET imaging performance with Fluorine-18 (18F) and Gallium-68 (68Ga) for influence of physical properties of PET tracer. Measurement were performed on a Siemens Biograph mCT64 PET/CT scanner using NEMA IEC body phantom and Flangeless Esser PET phantom containing filled with 18F and 68Ga. Emission scan duration(ESD) was set to 1, 2, 3, 4 and 5min/bed for 68Ga and 1min/bed for 18F. The PET image were evaluated in terms of contrast, spatial resolution. Under same condition, The percentage of contrast recovery measured in the phantom ranged from 16.88% to 72.56% for 68Ga and from 27.51% to 74.43% for 18F and The FWHM value to evaluate spatial resolution was 10.96 ㎜ for 68Ga and 9.19 ㎜ for 18F. For this study, 18F produces better image contrast and spatial resolution than 68Ga due to higher positron yield and lower positron energy (18F: 96.86%, 633.5 keV, 68Ga: 88.9%, 1899 keV), The physical properties of PET tracer effect on the PET image. 68Ga image applying ESD of 3, 4, 5min/bed were showed similar to 18F image with ESD of 1min/bed. This study suggests that increasing ESD for acquiring 68Ga PET image seem to be similar to 18F image.

참고문헌 (22)

  1. 김인영, "FDG를 이용한 Brain PET에서 Glucose Level이 영상에 미치는 영향", 방사선기술과학, 40(2), pp.275~280 (2017), 대한방사선과학회[DOI]
  2. 정재민, "양전자 방출핵종 68Ga을 이용한 NOTA와 DOTA의 표지 및 시험관내 특성 연구", Nuclear Medicine and Molecular Imaging, 43(4), pp.330~336 (2009), 대한핵의학회
  3. Fitzpatrick C, "<TEX>$^{68}Ga$<", TEX>-PSMAPET, Anticancer Research, pp.2017, 37(6)[DOI]
  4. Harshad R K, "Molecular imaging using PET", CT applying <TEX>$^{68}Ga$<, Journal of Postgraduate Medicine, Education and Research, pp.2013, 47(1)[DOI]
  5. Ronald B, "FDG PET and PET", CT: EANM procedure guidelines for tumour PET imaging: version 1.0, 2010, pp.37(1), 181~200
  6. Conti M, "Physics of pure and non-pure positron emitters for PET: a review and discussion", European Journal of Nuclear Medicine and Molecular Imaging Physics, 3(1), pp.1~17 (2016)
  7. National Electrical Manufacturers Association, "Performance Measurements of Positron Emission Tomographs. NEMA Standards Publication NU 2-2012", ~ (2012)
  8. Oliveira JF, "Influence of the time of flight and acquisition time on image quality in positron emission tomography", computed tomography scanners, Polytechnic Institute of Lisbon, 2016
  9. 이재영, "Breast PET CT 영상 재구성 변화에 따른 대조도 대 잡음비와 신호 대 잡음비의 비교평가", 방사선기술과학, 40(1), pp.79~85 (2017), 대한방사선과학회[DOI]
  10. ", Scientific Reports volume 8"[URL]
  11. Ko HS, "A study on comparative analysis of SUVs before and after correction with use of recovery coefficient (RC) in partial volume effect (PVE)", The Imaging Science Journal, 62(1), pp.11~15 (2014)[DOI]
  12. Isabel M, "Gallium-68: A New Trend in PET Radiopharmacy", Current Radiopharmaceuticals, 5(2), pp.142~149 (2012)[DOI]
  13. Canadas, "Performance evaluation for <TEX>$^{68}Ga$<", TEX> and <TEX>$^{18}F$<, IEEE Nuclear Science Symposium conference record. Nuclear Science Symposium, pp.2010[DOI]
  14. SanchezCrespo A, "Comparison of Gallium-68 and Fluorine-18 imaging characteristics in positron emission tomography", Applied Radiation and Isotopes, 76, pp.55~62 (2013)[DOI]
  15. Levin CS, "Calculation of positron range and its effect on the fundamental limit of positron emission tomography system spatial resolution", Physics in Medicine &. Biology, 45(2), pp.781~799 (1999)
  16. Susanne Z, "NEMA image quality phantom measurements and attenuation correction in integrated PET", MR hybrid imaging, 2015, pp.2(18), 1~17
  17. Habib J, "Quantitative Analysis in Nuclear Medicine Imaging", pp.531~ (2006), Springer Science & Business Media
  18. Soret M, "Partial-Volume Effect in PET Tumor Imaging", Journal of Nuclear Medicine, 48(6), pp.932~945 (2007)[DOI]
  19. Soile PS, "Head-to-Head Comparison of <TEX>$^{68}Ga$<", TEX>-Citrate and <TEX>$^{18}F$<, CT for Detection of Infectious Foci in Patients with Staphylococcus aureus Bacteraemia, pp.European Journal of Nuclear Medicine and Molecular Imaging, 2009[DOI]
  20. Kayani I, "Functional Imaging of Neuroendocrine Tumors With Combined PET", CT Using <TEX>$^{68}Ga$<, TEX>-FDG, pp.The Cancer Journal, 2008[DOI]
  21. ZhaoX, "Comparison between <TEX>$^{68}Ga$<", TEX>-labelled PSMA and <TEX>$^{18}F$<, CT in the diagnostic value of clear cell renal cell carcinoma, pp.European Urology Supplements, 2018[DOI]
  22. Bertolli O, "PET iterative reconstruction incorporating an efficient positron range correction method", Physica Medica: European Journal of Medical Physics, 32(2), pp.323~330 (2016)

관련 논문AI추천

이 간행물의 다른 논문 (9건)