SPECT/CT의 획득시간 증감에 따른 방사능농도 추정치의 변화
Variation on Estimated Values of Radioactivity Concentration according to the Change of the Acquisition Time of SPECT/CT
0
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
With the recent development of precision medicine(Theranostics), interest and utilization of the quantitative function of SPECT/CT are increasing. This study aims to investigate the effect on the radioactivity concentration estimate by the increase or decrease in the total time of SPECT/CT imaging conditions. A standard image was obtained by the conditions of a total acquisition time of 600 sec(10 sec/f × 120 frames) by diluting 99mTc 91.76 MBq in a cylindrical phantom filled with sterile water, and a comparative image was obtained by increasing the total acquisition time by -90%, -75%, -50%, -25%, +50%, +100%. The CNR, radioactive concentration estimate(cps/ml), and the variation rate(%) of the recovery coefficient(RC) were analyzed by measuring the overall coefficient of interest in each image. The results[CNR, Radiation Concentration, RC] by the change in the number of projections for each increase or decrease rate(-90%, -75%, -50%, -25%, +50%, +100%) of total acquisition time are as follows. [-89.5%, +3.90%, 1.04] at –90%, [-77.9%, +2.71%, 1.03] at –75%, [-55.6%, +1.85%, 1.02] at –50%, [-33.6%, +1.37%, 1.01] at –25%, [-33.7%, +0.71%, 1.01] at +50%, [+93.2%, +0.32%, 1.00] at +100%. and also The results[CNR, Radiation Concentration, RC] by the acquisition time change for each increase or decrease rate(-90%, -75%, -50%, -25%, +50%, +100%) of total acquisition time are as follows. [-89.3%, -3.55%, 0.96] at –90%, [-73.4%, -0.17%, 1.00] at –75%, [-49.6%, -0.34%, 1.00] at –50%, [-24.9%, 0.03%, 1.00] at –25%, [+49.3%, -0.04%, 1.00] at +50%, [+99.0%, +0.11%, 1.00] at +100%. Image quality(CNR) showed a pattern of change in proportion to the increase or decrease in the total acquisition time of SPECT/CT, but the result at quantitative evaluation showed a change of less than 5% in all experimental conditions, maintaining quantitative accuracy(RC less than 0.05) without much influence.
Abstract
With the recent development of precision medicine(Theranostics), interest and utilization of the quantitative function of SPECT/CT are increasing. This study aims to investigate the effect on the radioactivity concentration estimate by the increase or decrease in the total time of SPECT/CT imaging conditions. A standard image was obtained by the conditions of a total acquisition time of 600 sec(10 sec/f × 120 frames) by diluting 99mTc 91.76 MBq in a cylindrical phantom filled with sterile water, and a comparative image was obtained by increasing the total acquisition time by -90%, -75%, -50%, -25%, +50%, +100%. The CNR, radioactive concentration estimate(cps/ml), and the variation rate(%) of the recovery coefficient(RC) were analyzed by measuring the overall coefficient of interest in each image. The results[CNR, Radiation Concentration, RC] by the change in the number of projections for each increase or decrease rate(-90%, -75%, -50%, -25%, +50%, +100%) of total acquisition time are as follows. [-89.5%, +3.90%, 1.04] at –90%, [-77.9%, +2.71%, 1.03] at –75%, [-55.6%, +1.85%, 1.02] at –50%, [-33.6%, +1.37%, 1.01] at –25%, [-33.7%, +0.71%, 1.01] at +50%, [+93.2%, +0.32%, 1.00] at +100%. and also The results[CNR, Radiation Concentration, RC] by the acquisition time change for each increase or decrease rate(-90%, -75%, -50%, -25%, +50%, +100%) of total acquisition time are as follows. [-89.3%, -3.55%, 0.96] at –90%, [-73.4%, -0.17%, 1.00] at –75%, [-49.6%, -0.34%, 1.00] at –50%, [-24.9%, 0.03%, 1.00] at –25%, [+49.3%, -0.04%, 1.00] at +50%, [+99.0%, +0.11%, 1.00] at +100%. Image quality(CNR) showed a pattern of change in proportion to the increase or decrease in the total acquisition time of SPECT/CT, but the result at quantitative evaluation showed a change of less than 5% in all experimental conditions, maintaining quantitative accuracy(RC less than 0.05) without much influence.
참고문헌 (15)
- Radiological Protection in Therapy with Radiopharmaceuticals, "ICRP", 11~, pp.ICRP Publication 140 (2018)
- Arda K, "Theranostics : The Role of Quantitative Nuclear", Semin Radiat Oncol, 31(1), pp.28~36 (2021)
- Bin H, "Effects of shortened acquisition time on accuracy and precision of quantitative estimates of organ activity", Am Assoc Phys Med, 37(4), pp.1807~1815 (2010)
- 박훈희, "Usefulness of CT based SPECT Fusion Image in the lung Disease:Preliminary Study", 방사선기술과학, 35(1), pp.59~64 (2012), 대한방사선과학회
- Performance Measurements of Gamma Cameras, "NEMA", 44~47, pp.NEMA Standards Publication NU 1-2012 (2013)
- Quantitative Nuclear Medicine Imaging:Concepts, Requirements and Methods, "IAEA", 36~, pp.IAEA Human Health Reports No.9 (2014)
- Nakamura Y, "Usefulness of the collimator detector response(CDR)recovery and the scatter correction by the effective scatter source estimation(ESSE)method in myocardial SPECT study", Nihon Hoshasen Gijutsu Gakkai Zasshi, 66(6), pp.609~617 (2010)
- Johannes Z, "Quantitative accuracy of clinical 99mTc SPECT", CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction, 2010, pp.51, 921~928
- Kim JG, "A Study on the Tendency of Dose value According to Dose calibrator Measurement Depth and Volume", J Nucl Med Technol, 24(1), pp.20~26 (2020)
- Alfonso RM, "The Generalized Contrast-to-Noise Ratio : A Formal Definition for Lesion Detectability", IEEE Trans Ultrason Ferroelectr Freq Control, 67(4), pp.745~759 (2020)
- Anil A, "Improving Spatial Resolution Using Incoherent Subtraction of Receive Beams Having Different Apodizations", IEEE Trans Ultrason Ferroelectr Freq Control, 66(1), pp.5~17 (2019)
- 이재영, "Breast PET CT 영상 재구성 변화에 따른 대조도 대 잡음비와 신호 대 잡음비의 비교평가", 방사선기술과학, 40(1), pp.79~85 (2017), 대한방사선과학회[DOI]
- 김기원, "흉추 측면검사 영상의 CNR과 SNR 측정의 비교 연구", 방사선기술과학, 36(4), pp.273~280 (2013), 대한방사선과학회
- Marin I, "Establishment of a clinical SPECT", CT protocol for imaging of 161Tb, 2020, pp.7(45), 6~7
- "Help Me Understand Genetics ‘Precision Medicine’"[URL]