PET/CT 영상의 부분체적효과와 장기의 움직임으로 인해 감소된 SUV의 보정을 위한 회복계수의 비교

Comparison of Recovery Coefficients for Correction of Reduced SUV by Partial Volume Effect and Organ Movements in PET/CT Images

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

In this study, a recovery coefficient (RC) calculation was conducted that can correct the underestimation of the standardized uptake value (SUV) due to the partial volume effect (PVE) through phantom measurements and formulas. The experiment was conducted using a dynamic phantom capable of implement cranio-caudal movement at a respiratory rate of 15 times per minute along with the measured phantom experiment of the stopped state, and the RC of the moving state is calculated and compared. Ingenuity TF (Philips Healthcare, Netherland) was used as a positron emission tomography/ computed tomography (PET/CT) device. PET-CT Phantom (Biodex Medical System, USA) was used as a phantom for measurement. A phantom image in a stationary state was acquired, and a moving phantom image was acquired using the AZ-733V Respiratory Phantom (Anzai Medical Co, Japan) capable of breathing movement in the cranio-caudal direction under the same acquisition parameters. For RC calculation, the sphere maximum radioactivity concentration and the background mean radioactivity concentration of the acquired images were measured, and the initially determined sphere and background radioactivity concentrations were calculated. The calculated RC was 0.08 to 0.72. The size of sphere smaller, it was confirmed that the RC reduced. And the RC in the moving state reduced than in the stationary state. As a result of this study, the change of the RC was confirmed according to the size of spheres and the phantom moving. Using the RC derived by implement movement of breathing with the respiratory phantom, it is possible to considering correction of underestimated SUV by the partial volume effect of PET images and the patient movements.

Abstract

In this study, a recovery coefficient (RC) calculation was conducted that can correct the underestimation of the standardized uptake value (SUV) due to the partial volume effect (PVE) through phantom measurements and formulas. The experiment was conducted using a dynamic phantom capable of implement cranio-caudal movement at a respiratory rate of 15 times per minute along with the measured phantom experiment of the stopped state, and the RC of the moving state is calculated and compared. Ingenuity TF (Philips Healthcare, Netherland) was used as a positron emission tomography/ computed tomography (PET/CT) device. PET-CT Phantom (Biodex Medical System, USA) was used as a phantom for measurement. A phantom image in a stationary state was acquired, and a moving phantom image was acquired using the AZ-733V Respiratory Phantom (Anzai Medical Co, Japan) capable of breathing movement in the cranio-caudal direction under the same acquisition parameters. For RC calculation, the sphere maximum radioactivity concentration and the background mean radioactivity concentration of the acquired images were measured, and the initially determined sphere and background radioactivity concentrations were calculated. The calculated RC was 0.08 to 0.72. The size of sphere smaller, it was confirmed that the RC reduced. And the RC in the moving state reduced than in the stationary state. As a result of this study, the change of the RC was confirmed according to the size of spheres and the phantom moving. Using the RC derived by implement movement of breathing with the respiratory phantom, it is possible to considering correction of underestimated SUV by the partial volume effect of PET images and the patient movements.

참고문헌 (16)

  1. 박훈희, "PET", CT 검사에서 비스무스(bismuth) 차폐체의 적용에 따른 유용성 평가, 2014, pp.37(1), 49~56[DOI]
  2. 김희진, "간 전이 암 환자의 18F-FDG PET 기반 종양 영역 정의: 영상 인자와 자동 영상분할 기법 간의 관계분석", 의학물리, 24(2), pp.99~107 (2013), 한국의학물리학회
  3. 김정수, "핵의학 융합영상의 표준섭취계수 차이에 관한 연구", 방사선기술과학, 41(6), pp.553~560 (2018), 대한방사선과학회[DOI]
  4. Thie JA, "Understanding the Standardized Uptake Value, Its Methods, and Implications for Usage", The Journal of Nuclear Medicine, 45(9), pp.1431~1434 (2004)
  5. Kim CK, "Standardized Uptake Values of FDG : Body Surface Area Correction is Preferable to Body Weight Correction", The Journal of Nuclear Medicine, 5(1), pp.164~167 (1994)
  6. Hamill JJ, "Evaluation of CT-based lean-body SUV", Medical Physics, 40(9), pp.092054-1~092054-9 (2013)[DOI]
  7. F. Gallivanone, "A Partial Volume Effect Correction Tailored for 18F-FDG-PET Oncological Studies", BioMed Research International, 2013, pp.1~12 (2013), Hindawi Limited[DOI]
  8. Soret M, "Partial-Volume Effect in PET Tumor Imaging", The Journal of Nuclear Medicine, 48, pp.932~945 (2007)
  9. Erlandsson K, "MR Imaging-Guided Partial Volume Correction of PET Data in PET", MR Imaging, 2016, pp.11(2), 161~177
  10. Riou, O, "Integrating respiratory-gated PET-based target volume delineation in liver SBRT planning, a pilot study", Radiation Oncology, 9, pp.127~ (2014)
  11. Salavati A, "Application of Partial Volume Effect Correction and 4D PET in the Quantification of FDG Avid Lung Lesions", Molecular Imaging and Biology, 17, pp.140~148 (2015)
  12. 한기택, "LYSO 섬광체를 이용한 감마선 분광용 광섬유 방사선 센서의 개발", 센서학회지, 21(4), pp.287~292 (2012), 한국센서학회[DOI]
  13. NEMA NU 2-2001Performance Measurements of Positron Emission Tomographs, "Hines H", National Electrical Manufacturers Association, ~ (2001)
  14. Mesoloras G, "Measurement of Tumor Volume for Fluorodeoxyglucose-Positive Primary and Metastatic Liver Tumors : Validation of Optimal Threshold Value Technique", Journal of Interventional Oncology, 2(2), pp.95~104 (2009)
  15. Soto-Cordova MM, "An IoT based Urban Areas Air Quality Monitoring Prototype", International Journal of Advanced Computer Science and Applications, 11(9), pp.711~716 (2020)
  16. Srinivas SM, "A recovery coefficient method for partial volume correction of PET images", Annals of Nuclear Medicine, 23, pp.341~348 (2009)

관련 논문AI추천

이 간행물의 다른 논문 (7건)