본문 바로가기
KERD
논문KCI 등재

1.5T 자기공명영상을 이용한 물리적 영상 특성에 대한 연구

Study on the Physical Imaging Characteristics by Using Magnetic Resonance Imaging 1.5T

0

조회

0

다운로드

4

인용

0

참고문헌

키워드

주제분류방사선과학

초록

This study was purpose to quantitative evaluation of noise power spectrum(NPS) and studied the quantitative evaluation and characteristics of modulation transfer function(MTF) by obtain the optimal edge image by using Coil in magnetic resonance imaging(MRI) equipment through Fujita theory using edge method. The MRI equipment was used (Tim AVANTO 1.5T, Siemense healthcare system, Germany) and the head matrix coil were 12channels(elements) receive coil. The NPS results of showed the best value of 0.004 based on the T2 Nyquist frequency of 1.0 mm-1, and the MTF results of showed that the T1 and T2 values were generally better than the T1 CE and T1 CE FC values. The characteristics of this study were to explain the characteristic method of image quality evaluation in general. To present the quantitative evaluation process and results in the evaluation of MRI image characteristics in radiology

Abstract

This study was purpose to quantitative evaluation of noise power spectrum(NPS) and studied the quantitative evaluation and characteristics of modulation transfer function(MTF) by obtain the optimal edge image by using Coil in magnetic resonance imaging(MRI) equipment through Fujita theory using edge method. The MRI equipment was used (Tim AVANTO 1.5T, Siemense healthcare system, Germany) and the head matrix coil were 12channels(elements) receive coil. The NPS results of showed the best value of 0.004 based on the T2 Nyquist frequency of 1.0 mm-1, and the MTF results of showed that the T1 and T2 values were generally better than the T1 CE and T1 CE FC values. The characteristics of this study were to explain the characteristic method of image quality evaluation in general. To present the quantitative evaluation process and results in the evaluation of MRI image characteristics in radiology

참고문헌 (17)

  1. Hahm HK, "The study on the subject development of MRI image quality evaluation Department of Radiology", master’s, pp.2008
  2. Miyati T, "Measurements of MTF and SNR(f) using a subtraction method im MRI", Physics in Medicine and Biology, 47, pp.2961~2972 (2002)
  3. Steckner MC, "Computing the modulation transfer function of a magnetic resonance imager", Med. Phys, 21, pp.483~489 (1994)
  4. 김기원, "컴퓨터단층검사에서 AAPM Phantom을 이용한 변조전달함수 평가", 방사선기술과학, 39(2), pp.193~198 (2016), 대한방사선과학회[DOI]
  5. 손순룡, "Evaluation of Image Quality for Various Electronic Portal Imaging Devices in Radiation Therapy", 방사선기술과학, 38(4), pp.451~461 (2015), 대한방사선과학회[DOI]
  6. 김기원, "유방영상에서 엣지를 이용한 물리적 영상 평가", 방사선기술과학, 39(3), pp.415~420 (2016), 대한방사선과학회[DOI]
  7. 손순룡, "Measurement of Image Quality According to the Time of Computed Radiography System", 방사선기술과학, 38(4), pp.365~374 (2015), 대한방사선과학회[DOI]
  8. 정회원, "CsI(TI) Indirect Flat Panel Detector의 선질에 따른 물리적 영상 평가", 방사선기술과학, 35(2), pp.109~117 (2012), 대한방사선과학회[DOI]
  9. 정회원, "다양한 디지털 방사선 시스템의 물리적 영상 특성 조사", 방사선기술과학, 40(3), pp.363~370 (2017), 대한방사선과학회[DOI]
  10. 민정환, "간접평판형 검출기에서 변조전달함수 측정 시 Edge 각도에 따른 비교 연구", 방사선기술과학, 42(4), pp.259~263 (2019), 대한방사선과학회
  11. 민정환, "간접평판형 검출기에서 국제전자기술위원회 기준을 통한 잡음전력스펙트럼 비교 연구", 방사선기술과학, 41(5), pp.457~462 (2018), 대한방사선과학회[DOI]
  12. Mohapatra SM, "Transfer function measurement and analysis for magnetic resonance imager", Med. Phys, 18(6), pp.1141~1144 (1991)
  13. EEC Concerted Research Project, "Identification and characterization of biological tissues by NMR. Concerted Research Project of the European Economic Community: IV. Protocols and test objects for the assessment of MRI equipment", Magn. Reson. Imaging, 6, pp.195~199 (1988)
  14. Fujita H, "A simple method for determining the modulation transfer function in digital radiography", IEEE Trans Med Imaging, 11(1), pp.34~39 (1992)
  15. Samei E, "A method for measuring the presampled MTF of digital radiographic systems using an edge test device", Medical Physics, 25, pp.102~ (1998)
  16. Greer PB, "Evaluation of an algorithm for the assessment of the MTF using an edge method", Medical Physics, 27, pp.2048~ (2000)
  17. 최관우, "고관절의 자기공명관절조영검사 시 가돌리늄 함유량이 높은 조영제의 유용성", 한국산학기술학회논문지, 14(11), pp.5682~5688 (2013), 한국산학기술학회[DOI]