본문 바로가기
KERD
논문KCI 등재

인터벤션 시술 시 환자의 선량감소를 위한 3D 프린팅 재료의 적용성 평가

Feasibility of the 3D Printing Materials for Radiation Dose Reduction in Interventional Radiology

2

조회

0

다운로드

3

인용

0

참고문헌

키워드

주제분류방사선과학

초록

Interventional radiology is performed under real-time fluoroscopy, and patients are exposed to a wide range of exposures for a long period of time depending on the examination and procedure. However, studies on radiation protection for patients during an intervention are insufficient. This study aims to evaluate the doses exposed during the intervention and the applicability of 3D printing materials. The organ dose for each intervention site was evaluated using a monte carlo simulatio. Also, the dose reduction effect of the critical organs was calculated when using a shielding device using 3D printing materials. As a result, the organ dose distribution for each intervention site showed a lower dose distribution for organs located far from the x-ray tube. It was analyzed that the influence of scattered rays was higher in the superficial organs of the back of the human body where x-rays were incident. The dose reduction effect on the critical o rgan u sing t he 3 D printing shield s howed the highest testis among t he g onads, a nd in the case o f other organs, the dose reduction effect gradually decreased in the order of the eye, thyroid, breast, and ovary. Accordingly, it is judged that the 3D printed shield will be sufficiently usable as a shielding device for the radiation protection of critical organs.

Abstract

Interventional radiology is performed under real-time fluoroscopy, and patients are exposed to a wide range of exposures for a long period of time depending on the examination and procedure. However, studies on radiation protection for patients during an intervention are insufficient. This study aims to evaluate the doses exposed during the intervention and the applicability of 3D printing materials. The organ dose for each intervention site was evaluated using a monte carlo simulatio. Also, the dose reduction effect of the critical organs was calculated when using a shielding device using 3D printing materials. As a result, the organ dose distribution for each intervention site showed a lower dose distribution for organs located far from the x-ray tube. It was analyzed that the influence of scattered rays was higher in the superficial organs of the back of the human body where x-rays were incident. The dose reduction effect on the critical o rgan u sing t he 3 D printing shield s howed the highest testis among t he g onads, a nd in the case o f other organs, the dose reduction effect gradually decreased in the order of the eye, thyroid, breast, and ovary. Accordingly, it is judged that the 3D printed shield will be sufficiently usable as a shielding device for the radiation protection of critical organs.

참고문헌 (17)

  1. Radiation protection guidelines to reduce patient dose in interventional procedures, "Ministry of Food and Drug Safety", ~, pp.Radiation Safety Management Series. No.36 (2014)
  2. 박혁, "중재적 방사선 시술 시 환자선량에 대한 참고 준위 비교", 한국방사선학회논문지, 6(1), pp.11~17 (2012), 한국방사선학회
  3. Valentin J, "Avoidance of radiation injuries from medical interventional procedures. ICRP Publication 85", Annals of the ICRP, 30(2), pp.7~67 (2000)
  4. Radiation protection guidelines following interventional radiation procedures, "Ministry of Food and Drug Safety", ~, pp.Radiation Safety Management Series. No.11 (2006)
  5. 김정훈, "혈관조영 및 중재적 시술 분야 내 종사자의 눈에 대한 선량평가", 한국방사선학회논문지, 12(2), pp.209~216 (2018), 한국방사선학회[DOI]
  6. 임영기, "몬테카를로 방법을 이용한 중재방사선시술자에 대한 선량평가", 방사선방어학회지, 39(4), pp.213~217 (2014), 테스트학회
  7. Ceh J, "Bismuth infusion of ABS enables additive manufacturing of complex radiological phantoms and shielding equipment", Sensors, 17(3), pp.459~ (2017)
  8. Interventional radiation treatment device performance evaluation technical information, "Ministry of Food and Drug Safety", ~, pp.Radiation Safety Management Series. No.11 (2011)
  9. Kim JH, "A study on the annual cumulative radiation exposure of patients undergoing transhepatic artery chemoembolization(TACE)", Journal of Korean Society of Cardio-Vascular Interventional Technology, 18(1), pp.69~76 (2015)
  10. Balter S, "Fluoroscopically guided interventional procedures : A review of radiation effects on patients skin and hair", Radiology, 254(2), pp.326~341 (2010)
  11. Schueler BA, "An investigation of operator exposure in interventional radiology", Radiographics, 26(5), pp.1533~1541 (2006)
  12. 강병삼, "인터벤션 시술 시 면적선량계를 이용한 환자 방사선 선량 평가", 방사선기술과학, 40(1), pp.27~34 (2017), 대한방사선과학회[DOI]
  13. 박형신, "중재적방사선검사에서 환자 피폭선량에 관한 연구", 방사선기술과학, 35(4), pp.299~308 (2012), 대한방사선과학회
  14. 정우경, "투시와 중재시술의 방사선 피폭과 저감화 방법", 대한의사협회지, 54(12), pp.1269~1276 (2011), 대한의사협회
  15. Rehani MM, "Radiological protection in fluoroscopically guided procedures performed outside the imaging department", Annals of the ICRP, 40(6), pp.1~102 (2010)
  16. Paek SM, "A study on the reduction of exposure dose by shielding the major organs of patients in interventional procedures (TACE)", Journal of Korean Society of Cardio-vascular Interventional Technology, 15(1), pp.103~109 (2012)

관련 논문AI추천