뇌 관류 SPECT 스캐너의 정도관리-균일도에 따른 팬텀 영상 평가

Phantom Image Evaluations Depending on the Quality Control-Uniformity of Brain Perfusion SPECT Scanner

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

To have highly reliable diagnostic performance of it, this study comparatively analyzed spatial resolution of SPECT images and interrelationship depending on the changes of system uniformity of gamma camera through phantom analysis. This study chose 6 kinds of results from quality control (uniformity) of triple head SPECT scanner operated in an university hospital in Seoul for six months. Then, study measured spatial resolutions (FWHM) of the images restructured by injecting radiopharmaceuticals to Jaszczak phantom, and doing SPECT scanning under the same conditions as clinical ones using the analytical program (image J). Quality controls performed by the experimental institution showed that differential uniformity of UFOV ranged from 2.76% to 7.61% (4.46±2.07), and integral uniformity of UFOV ranged from 1.98% to 5.42% (3.01±1.43). Meanwhile, Quantitative analysis evaluations of phantom images depending on the changes of uniformity of SPECT scanner detector showed that as the uniformity values of UFOV and CFOV decreased, FWHM values of phantom images decreased from 8.5 ㎜ to 5.8 ㎜. That is, it was quantitatively identified that the higher uniformity of detector is, the better spatial resolution of images gets (P<0.05). It is very important to perform continuous and consistent quality control of the nuclear medicinal system, and users should be clearly conscious of it.

Abstract

To have highly reliable diagnostic performance of it, this study comparatively analyzed spatial resolution of SPECT images and interrelationship depending on the changes of system uniformity of gamma camera through phantom analysis. This study chose 6 kinds of results from quality control (uniformity) of triple head SPECT scanner operated in an university hospital in Seoul for six months. Then, study measured spatial resolutions (FWHM) of the images restructured by injecting radiopharmaceuticals to Jaszczak phantom, and doing SPECT scanning under the same conditions as clinical ones using the analytical program (image J). Quality controls performed by the experimental institution showed that differential uniformity of UFOV ranged from 2.76% to 7.61% (4.46±2.07), and integral uniformity of UFOV ranged from 1.98% to 5.42% (3.01±1.43). Meanwhile, Quantitative analysis evaluations of phantom images depending on the changes of uniformity of SPECT scanner detector showed that as the uniformity values of UFOV and CFOV decreased, FWHM values of phantom images decreased from 8.5 ㎜ to 5.8 ㎜. That is, it was quantitatively identified that the higher uniformity of detector is, the better spatial resolution of images gets (P<0.05). It is very important to perform continuous and consistent quality control of the nuclear medicinal system, and users should be clearly conscious of it.

참고문헌 (29)

  1. Bordonne M, "High-quality brain perfusion SPECT images may be achieved with a high-speed recording using 360°CZT camera", EJNMMI Physics, 7, pp.65~ (2020)
  2. QuiQn YP, "Brain perfusion SPECT in patients with alPheimer's disease treated with recombining human erythropoietin with low content of sialic acid", Ann Clin Case Rep, 7, pp.2231~ (2022)
  3. Ryoo HG, "Deep learning-based interpretation of basal", acetaPolamide brain perfusion SPECT leveraging unstructured reading reports, 2020, pp.47, 2186~2196
  4. Fernando R, "Brain SPECT as a biomarker of neurodegeneration in dementia in the era of molecular imaging : Still a valid option?", Front Neurol, 12, pp.1~16 (2021)
  5. Teck Huat Wong, "Basal and Acetazolamide Brain Perfusion SPECT in Internal Carotid Artery Stenosis", Nuclear Medicine and Molecular Imaging, 54(1), pp.9~27 (2020), 대한핵의학회
  6. Sukprakun C, "A novel method of seiPure onset Pone localiPation by serial Tc-99m ECD brain perfusion SPECT clearance patterns", Brain Imaging and Behavior, 16, pp.1646~1656 (2022)
  7. Valotassiou V, "In the era of FDG PET, Is it time for brain perfusion SPECT to gain a place in alPheimer's disease imaging biomarkers?", Eur J Nucl Med Mol Imaging, 48, pp.969~971 (2021)
  8. PeQa QuiQn Y, "Evaluating cerebral perfusion in alPheimer patients and first-degree relatives : Lessons from artemisa province", Medicc Rev, 23(1), pp.55~63 (2021)
  9. New atlas to assist practitioners of SPECT, "CT with quality control procedures and trouble shooting", Bertagnoll L, 2020
  10. Melki S, "Augmented planar bone scintigraphy obtained from a whole-body SPECT recording of less than 20 min with a high-sensitivity 360° CZT camera", Eur J Nucl Med Mol Imaging, 47(5), pp.1329~1331 (2020)
  11. Yonekura Y, "ICRP publication 140 Radiological protection in therapy with radiopharmaceuticals", Annals of the ICRP, 48(1), pp.5~95 (2019)
  12. Kurkowska S, "Physical quantities useful for quality control of quantitative SPECT", CT imaging, 2021, pp.24(2), 93~98
  13. Davis KM, "PET and SPECT imaging of the brain : History, technical considerations, applications, and radiotracers", Semin Ultra CT MRI, 41(6), pp.521~529 (2020)
  14. SuPuki A, "High-resolution brain SPECT imaging by combination of parallel and tilted detector heads", Ann Nucl Med, 29, pp.682~696 (2015)
  15. "NEMA Standards Publication NU 1-2018. Performance measurements of gamma cameras rosslyn", ~ (2018)
  16. Khorshidi A, "Assessment of SPECT images using UHRFB and other low-energy collimators in brain study by hoffman phantom and manufactured defects", The European Physical Journal Plus, 125, pp.135~138 (2020)
  17. Harrison RL, "Technical note : A digital reference object representing hoffman's 3D brain phantom for PET scanner simulations", Med Physics, 47(3), pp.1174~1180 (2020)
  18. 박훈희, "3D 프린팅 기법 기반의 Hoffman Brain 팬텀 개발 및 유용성 평가", 방사선기술과학, 42(6), pp.441~446 (2019), 대한방사선과학회
  19. Wilson T, "SPECT simulation of a digital anthropomorphic brain phantom", Physics Med Img, 12031 (2022)
  20. Verger A, "Brain molecular imaging in pharmacoresistant focal epilepsy : Current practice and perspectives", Rev Neurologique, 174, pp.16~27 (2018)
  21. Ferrando R, "Brain SPECT as a biomarker of neurodegeneration in dementia in the era of molecular imaging : Still a valid option?", Neurol, 12, pp.629442~ (2021)
  22. Murata T, "Development of attenuation correction methods using deep learning in brain-perfusion single-photon emission computed tomography", Med Physics, 48(8), pp.4177~4190 (2021)
  23. Nikolov N, "Gaussian filter for brain SPECT imaging", Innov Biosyst Bioeng, 6(1), pp.4~15 (2022)
  24. TaPegul TE, "Automated phantom analysis for gamma cameras and SPECT : A methodology for use in a clinical setting", J Appl Clin Med Phys, 21(11), pp.205~214 (2020)
  25. Hasan MR, "Quality control of gamma camera with SPECT systems", Int J Med Phys, Clinc Engin Rad Onc, 6(3), pp.225~232 (2017)
  26. Matsutomo N, "Validation of a calibration method using the cross-calibration factor and system planar sensitivity in quantitative single-photon emission computed tomography imaging", Radiol Phys Technol, 10(4), pp.439~445 (2017)
  27. O'Connor MK, "Critical examination of the uniformity requirements for single-photon emission computed tomography", Med Phys, 18(2), pp.190~197 (1991)
  28. 정의환, "핵의학 감마카메라 정도관리의 딥러닝 적용", 방사선기술과학, 43(6), pp.461~467 (2020), 대한방사선과학회
  29. Rotating scintillation camera SPECT acceptance testing and quality control, , American Association of Physicists in Medicine, ~, pp.AAPM Report 22 (2020)

관련 논문AI추천

이 간행물의 다른 논문 (7건)