본문 바로가기
KERD
논문KCI 등재

적외선 분광학에 의한 간질액 글루코즈 농도 측정의 정확도 향상

Accuracy improvement in the interstitial glucose measurement based on infrared spectroscopy

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류공학일반

초록

Glucose concentrations in the interstitial fluid were measured based on optical spectroscopy. Prediction of glucosepreprocesing as well as by selecting an optimal wavelength region. For this purpose, artificial interstitial fluid sampleswere prepared where their glucose levels varied between 0 and 10 g/dl. Infrared spectral regions where glucose absorptionlies were investigated. A region of 1000-1500 cm1 produced the best accuracy among the regions of 1000-1500 cm1,4000-4545 cm1 and 5500-6500 cm1. Further accuracy improvement in 1000-1500 cm1 was achieved by selectingspecific wavelength bands based on a loading vector analysis method. For the samples whose glucose concentrations1. However, theloading vector optimized band of 1002-1095 cm1 reduced the prediction error up to 47 % (SEP=0.0125 g/dl andR=0.9970).

Abstract

Glucose concentrations in the interstitial fluid were measured based on optical spectroscopy. Prediction of glucosepreprocesing as well as by selecting an optimal wavelength region. For this purpose, artificial interstitial fluid sampleswere prepared where their glucose levels varied between 0 and 10 g/dl. Infrared spectral regions where glucose absorptionlies were investigated. A region of 1000-1500 cm1 produced the best accuracy among the regions of 1000-1500 cm1,4000-4545 cm1 and 5500-6500 cm1. Further accuracy improvement in 1000-1500 cm1 was achieved by selectingspecific wavelength bands based on a loading vector analysis method. For the samples whose glucose concentrations1. However, theloading vector optimized band of 1002-1095 cm1 reduced the prediction error up to 47 % (SEP=0.0125 g/dl andR=0.9970).

참고문헌 (16)

  1. R. G. Tiessen, "Glucose gradient differences in subcutaneous tissue of healthy volunteers assessed with ultraslow microdialysis and a nanolitre glucose sensor", Life Sciences, 70(21), pp.2457~2466 (2002)
  2. S. Gebhart, "Glucose sensing in transdermal body fluid collected under continuous vacuum pressure via micropores in the stratum corneum", Diabetes Technology & Therapeutics, 5(2), pp.159~166 (2003)
  3. B. Buckingham, "Response to nocturnal alarms using a real-time glucose sensor", Diabetes Technology & Therapeutics, 7(3), pp.440~447 (2005)
  4. B. D. Cameron, "Measurement of the glucose transport time delay between the blood and aqueous humor of the eye for the eventual development of a noninvasive glucose sensor", Diabetes Technology & Therapeutics, 3(2), pp.201~207 (2001)
  5. Y-J Kim, "Determination of glucose in whole blood samples by mid-infrared spectroscopy", Applied Optics, 42(4), pp.745~749 (2003)
  6. 성호경, "생리학", pp.268~275 (1991), 의학문화사, 서울 마포구
  7. N. Fogh-Andersen, "Composition of Interstitial Fluid", Clinical Chemistry, 41(10), pp.1522~1525 (1995)
  8. 김연주, "실험계획법에 의한 혈중 글루코즈 측정 시 타 성분의 영향 분석", 의공학회 지, 22(6), pp.497~502 (2001)
  9. K. S. Y. Yeung, "Near-infrared spectroscopy for bioprocess monitoring and control", Biotechnology and Bioengineering, 63(6), pp.684~693 (1999)
  10. H. Chung, "Simultaneous measurements of glucose, glutamine, ammonia, lactate, and glutamate in aqueous solutions by near-infrared spectroscopy", Applied spectroscopy, 50(2), pp.270~276 (1996)
  11. A. C. Guybn, "Textbook of medical physiology, Sunders", pp.383~389 (1986), Philadelphia USA
  12. Y.C. Shen, "Determination of glucose concentration in whole blood using fouriertransform infrared spectroscopy", J. Biological Physics, 29(2-3), pp.129~133 (2003)
  13. J. P. Bantle, "Glucose measurement in patients with diabetes mellitus with dermal interstitial fluid", The J. laboratory and clinical medicine, 130(4), pp.436~441 (1997)
  14. Y-J. Kim, "Multicomponent assay for human serum using mid-infrared transmission spectroscopy based on component-optimized spectral region selected by first loading vector analysis", Applied Spectroscopy, 56(5), pp.625~632 (2002)
  15. Y-J Kim, "Prediction of glucose in whole blood by near-infrared spectroscopy: Influence of wavelength region, preprocessing, and hemoglobin concentration", J. Biomedical Optics, 11(4), pp.041128-1~041128-7 (2006)
  16. 전계진, "총 헤모글 로빈농도를 비침습적으로 측정하기 위한 400 - 2500 nm 대역의 흡수스펙트럼 분석", 센서학회지, 10(6), pp.273~278 (2001)

관련 논문AI추천