A disposable electrochemical immunosensor system has been developed for the detection of herbicide in aqueous samples. Disposable screen printed carbon electrodes(SPCE) were used as basic electrodes and an enzyme, horseradish peroxidase (HRP), and anti-herbicide antibodies was immobilised on to the working electrode of SPCE by using avidin-biotin coupling reactions. An herbicide-glucose oxidase conjugates have been used for the competitive immunoreaction with sample herbicides. The enzymatic reaction between the conjugated glucose oxidase and glucose added generates hydrogen peroxide, which was reduced by the peroxidase immobilised. The latter process caused an electrical current change, due to direct re-reduction of peroxidase by a direct electron transfer mechanism, which was measured to determine the herbicides in the sample. The optimal operational condition was found to be: 20 μgl-1 deglycosylated avidin loading to the working electrode and working potential +50 mV vs. Ag/AgCl. The total assay time was 15 min after sample addition. The detection limits for herbicides, atrazine and simazine, were found to be 3 ppb and 10 ppb, respectively.

Abstract

A disposable electrochemical immunosensor system has been developed for the detection of herbicide in aqueous samples. Disposable screen printed carbon electrodes(SPCE) were used as basic electrodes and an enzyme, horseradish peroxidase (HRP), and anti-herbicide antibodies was immobilised on to the working electrode of SPCE by using avidin-biotin coupling reactions. An herbicide-glucose oxidase conjugates have been used for the competitive immunoreaction with sample herbicides. The enzymatic reaction between the conjugated glucose oxidase and glucose added generates hydrogen peroxide, which was reduced by the peroxidase immobilised. The latter process caused an electrical current change, due to direct re-reduction of peroxidase by a direct electron transfer mechanism, which was measured to determine the herbicides in the sample. The optimal operational condition was found to be: 20 μgl-1 deglycosylated avidin loading to the working electrode and working potential +50 mV vs. Ag/AgCl. The total assay time was 15 min after sample addition. The detection limits for herbicides, atrazine and simazine, were found to be 3 ppb and 10 ppb, respectively.

참고문헌 (15)

  1. S. Schuman, "A clinical historical overview of pesticide health issues", Occup. Med, 12, pp.203~207 (1997)
  2. B.Johnson, "A review of health-based comparative risk assessments in the United States", Rev. Environ. Health, 15, pp.273~287 (2000)
  3. P. Hansen, "New biosensors for environmental analysis", EXS, 81, pp.109~120 (1997)
  4. I. Karube, "Biosensors for environmental monitoring", Ann. NY Acad. Sci, 864, pp.23~36 (1998)
  5. A. Warsinke, "Electrochemical immunoassays", Fresenius J. Anal.Chem, 366, pp.622~634 (2000)
  6. R. Scheller1, "Research and development in biosensors", Curr. Opin.Biotechnol, 12, pp.35~40 (2001)
  7. 박덕수, "파라세타몰 검출을 위한 전기화학적 다중효소 바이오센서", 센서학회지, 17(1), pp.29~34 (2008), 한국센서학회
  8. S. Kroger, "Imprinted polymer-based sensor system for herbicidesusing differential-pulse voltammetry on screen-printedelectrodes", Anal. Chem, 71, pp.3698~3702 (1999)
  9. T. Neufeld, "A micro flow injection electrochemical biosensor for organophosphorus pesticides", Biosens. Bioelectr, 15, pp.323~329 (2000)
  10. S. Chang, "Disposable tyrosinase-peroxidase bi-enzyme sensor for amperometric detection of phenols", Biosens. Bioelectr, 17, pp.1015~1023 (2002)
  11. 장승철, "Disposable in-field electrochemical potable sensor system for free available chlorine (FAC) detection", 센서학회지, 16(6), pp.449~456 (2007), 한국센서학회
  12. B. Dzantiev, "Determination of the herbicide chlorsulfuron by amperometric sensor based on separation-free bienzyme immunoassay", Sens. Actuat. B, 98, pp.254~261 (2004)
  13. Z. Marshall, "Modification of trinder's method for the assay of salicylates in biological fluids", Clinic. Chimica Acta, 54, pp.135~136 (1974)
  14. AOAC, "Association of Agricultural Chemists official method 974.27, Official methods of analysis of AOAC international" (2010), AOAC
  15. Council directive relating to the quality of water intended for human consumption, "80", 778, Official J. of the E.C.,, pp.1980, ~[DOI]