질소의 끓는점에서의 산업용 온도계 비교 교정과 불확도 평가

Calibration by Comparison and Uncertainty Assessment of Industrial Thermometers at the Boiling Point of Nitrogen

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류공학일반

초록

We devised calibration procedure for industrial thermometers by a comparison method at the boiling point of nitrogen (~-196°C). The uncertainty of the calibration was 4 mK (k = 2). As experimentally demonstrated in this work, the effect of the atmospheric pressure on the boiling point of nitrogen can be easily detected by the thermometer. Therefore, when the boiling point of nitrogen is used for calibration of thermometer by comparison, either a reference thermometer must be used to provide the reference temperature or the effect of atmospheric pressure should be carefully considered. The use of a copper block with a large thermal mass soaked into the liquid nitrogen was proven to be more reliable, and the stability of the temperature immersed into the copper block was 1.4 mK. The temperatures at the thermometer wells, evaluated by the crossed-measurement method to compensate for the inaccuracy of the thermometers and the linear drift of the temperature of the copper block, were equivalent within 0.23 mK of standard uncertainty.

Abstract

We devised calibration procedure for industrial thermometers by a comparison method at the boiling point of nitrogen (~-196°C). The uncertainty of the calibration was 4 mK (k = 2). As experimentally demonstrated in this work, the effect of the atmospheric pressure on the boiling point of nitrogen can be easily detected by the thermometer. Therefore, when the boiling point of nitrogen is used for calibration of thermometer by comparison, either a reference thermometer must be used to provide the reference temperature or the effect of atmospheric pressure should be carefully considered. The use of a copper block with a large thermal mass soaked into the liquid nitrogen was proven to be more reliable, and the stability of the temperature immersed into the copper block was 1.4 mK. The temperatures at the thermometer wells, evaluated by the crossed-measurement method to compensate for the inaccuracy of the thermometers and the linear drift of the temperature of the copper block, were equivalent within 0.23 mK of standard uncertainty.

참고문헌 (13)

  1. H. Preston-Thomas, "The international temperature scale of 1990 (ITS-90)", Metrologia, 27(1), pp.3~10 (1990)
  2. 양인석, "초정밀 온도센서의 저온 영역 교정 장치 개발", 센서학회지, 17(3), pp.188~194 (2008), 한국센서학회
  3. 양인석, "산업용 저항 온도계의 저온 자동 비교 교정 시스템 개발", 센서학회지, 17(6), pp.429~436 (2008), 한국센서학회
  4. T. Nakano, "Comparison system for the calibration of capsule-type standard platinum resistance thermometers at NMIJ", AIST, 2008, pp.29(3), 881~889
  5. G. M. Chen, "Development of an automatic calibration device for high-accuracy low temperature thermometers", Science China Technological Sciences, 53(9), pp.2404~2407 (2010)
  6. M. Ballico, "A cryostat for automated mK-level thermometer calibrations from -202°C to 250°C", Int. J. Thermophys, 32, pp.2360~2369 (2011)
  7. R. E. Bedford, "Recommended values of temperature on the international temperature scale of 1990 for a selected set of secondary reference points", Metrologia, 33(2), pp.133~154 (1996)
  8. J. V. Nicholas, "Traceable Temperatures", pp.203~253 (2001), Johm Wiley & Sons
  9. C. W. Meyer, "ITS-90 nonuniqueness from PRT subrange inconsistencies over the range 24.56 K to 273.16 K", Metrologia, 43(5), pp.341~352 (2006)
  10. "Working group 3 report to CCT : May 2012"[URL]

관련 논문AI추천

이 간행물의 다른 논문 (11건)