초록
For the accurate calibrations of thermometers at low temperatures, triple points of oxygen and argon have been realizedusing sealed-type triple point cells. The triple points were realized with both adiabatic and continuous heating methods.The resistance of thre capsule-type standard platinum resistance thermometers were measured at the triple points ofoxygen, argon, mercury and water for the calibrations of these thermometers in the range from 220oC to 0.01oC. Theextended uncertainty, with the coverage factor of 2, of the calibration in this range is smaller than 1 mK.
Abstract
For the accurate calibrations of thermometers at low temperatures, triple points of oxygen and argon have been realizedusing sealed-type triple point cells. The triple points were realized with both adiabatic and continuous heating methods.The resistance of thre capsule-type standard platinum resistance thermometers were measured at the triple points ofoxygen, argon, mercury and water for the calibrations of these thermometers in the range from 220oC to 0.01oC. Theextended uncertainty, with the coverage factor of 2, of the calibration in this range is smaller than 1 mK.
참고문헌 (6)
- H. Preston-Tomas, "The International Temperature Scale of 1990 (ITS-90)", Metrologia, 27(1), pp.3~10 (1990)
- 강기훈, "캡슐형 백금저항온도계 국 제비교를 위한 네온, 산소, 아르곤, 수은 및 물의 삼 중점 실현", 센서학회지, 9(3), pp.153~162 (2000)
- F. Pavese, "An international intercomparison of fixed points by means of sealed cells in the range 13.81 K - 90,686 K", Metrologia, 20(4), pp.127~144 (1984)
- 강기훈, "전기로 및 봉입형 응고점 셀 내의 온도구배가 미치는 표준백금저항온도계 온도측정의 불확도 요소 평가", 센서학회지, 13(6), pp.411~416 (2004), 한국센서학회
- 강기훈, "핵심국제비교용 무의 삼중점 셀 제작과 비교 실험 및 불확도 평가", 센서학회지, 13(2), pp.114~120 (2004), 한국센서학회
- B.W. Mangum, "On the international temperature scale of 1990 (ITS-90)", Metrologia, 34(5), pp.427~429 (1997)