고정밀 백금저항온도계를 이용한 교정기관의 온도교정능력 수행평가
Performance Assessment on Temperature Calibration Capability of the Calibration Laboratories Using High-Precision Platinum Resistance Thermometers
2
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Calibration capabilities for thermometer calibration by comparison method were assessed using high-precision industrial platinumresistance thermometers (IPRT). It was found in the performance assessment that out of 31 laboratories who participated, 28 laboratories resulted magnitude of En number less than 1 at every calibration points they submitted results in the range from 50 to 500 . The results of about 75% of the laboratories showed the difference from the assigned values less than 1/10 of the tolerance level of the class B IPRT. This indicates that the participating calibration laboratories performed with satisfactory level that was enough to calibrate IPRTs to significant precision. The sensors used in this work were manufactured and chosen by the criteria of long-term instability less than 4 mK and hysteresis less than 8 mK in the temperature range used in this work. Furthermore, the change in the resistance of the sensors in the calibration temperature range were less than the uncertainty of the calibration, 25 mK (k=2).
Abstract
Calibration capabilities for thermometer calibration by comparison method were assessed using high-precision industrial platinum resistance thermometers (IPRT). It was found in the performance assessment that out of 31 laboratories who participated, 28 laboratories resulted magnitude of En number less than 1 at every calibration points they submitted results in the range from 50 to 500 . The results of about 75% of the laboratories showed the difference from the assigned values less than 1/10 of the tolerance level of the class B IPRT. This indicates that the participating calibration laboratories performed with satisfactory level that was enough to calibrate IPRTs to significant precision. The sensors used in this work were manufactured and chosen by the criteria of long-term instability less than 4 mK and hysteresis less than 8 mK in the temperature range used in this work. Furthermore, the change in the resistance of the sensors in the calibration temperature range were less than the uncertainty of the calibration, 25 mK (k=2).
참고문헌 (9)
- "Korean Laboratory accreditation scheme, KOLASG- 004", ~ (2012)
- "Korean Laboratory accreditation scheme, KOLASR- 001", ~ (2012)
- "Korean Laboratory accreditation scheme, KOLASR- 003", ~ (2012)
- Korean Agency for Technology and Standards, "KS Q ISO", IEC 17043 Conformity assessment - General requirements for proficiency testing, 2011
- 감기술, "백금저항온도센서를 이용한 국가교정기관의 교정능력 평가 및 실험방법", 센서학회지, 14(4), pp.231~236 (2005), 한국센서학회
- K. S. Gam, "Measurement of traceability error for calibration services center using type S thermocouples", J. Sensor Sci. & Tech, 4(1), pp.43~50 (1995)
- K. S. Gam, "Thermal hysteresis in thin-film platinum resistance thermometers", Int. J. Thermophys, 32(11), pp.2388~2396 (2011)
- International Electrotechnical Commission, "IEC 60751 ed2.0, “Industrial platinum resistance thermometers and platinum temperature sensors”", ~ (2008)
- 김용규, "산업용 백금저항온도계를 위한 향상된 내삽식", 센서학회지, 21(2), pp.109~113 (2012), 한국센서학회