2D 혈관조영술에서 직접 측정한 혈관 직경과 MR 영상에서 단면적 기반 환산 직경의 비교 분석

Comparative Analysis between Directly Measured Diameter in 2D Angiography and Cross-Sectional Area-Converted Diameter in MR Image

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

This study aimed to quantitatively compare the diameters measured directly from the coronal plane or sagittal plane of 2D digital subtraction angiography (DSA) and the cross-sectional area-converted diameters calculated from contrast-enhanced MR (CE-MR) imaging. A retrospective analysis was conducted on 20 patients who underwent both 2D DSA and CE-MR imaging. Firstly, the venous diameters of the superior sagittal sinus (SSS) and transverse sinus (TS) were directly measured from 2D DSA. Subsequently, the axial planes for SSS diameter and the sagittal plane for TS in CE-MR imaging were utilized to calculate cross-sectional area-based converted diameters. The numerical values obtained from 2D DSA and CE-MR imaging were compared pairwise at each location. For SSS, the diameter measured by 2D DSA was 27% larger than the conversion-based diameter from CE-MR imaging (9.8±1.4 mm vs. 7.1±1.3 mm, P<0.05). Similarly, for the right TS, the difference was 16% (8.8±3.2 mm vs. 7.4±2.0 mm, P<0.05), and for the left TS, the difference was 22% (8.4±2.8 mm vs. 6.6±1.3 mm, P<0.05). In conclusion, the diameter measured directly in conventional 2D DSA may be larger than the diameter converted based on the cross-sectional area. Therefore, when selecting the size of the stent, it is crucial to make precise determinations while keeping this fact in mind.

Abstract

This study aimed to quantitatively compare the diameters measured directly from the coronal plane or sagittal plane of 2D digital subtraction angiography (DSA) and the cross-sectional area-converted diameters calculated from contrast-enhanced MR (CE-MR) imaging. A retrospective analysis was conducted on 20 patients who underwent both 2D DSA and CE-MR imaging. Firstly, the venous diameters of the superior sagittal sinus (SSS) and transverse sinus (TS) were directly measured from 2D DSA. Subsequently, the axial planes for SSS diameter and the sagittal plane for TS in CE-MR imaging were utilized to calculate cross-sectional area-based converted diameters. The numerical values obtained from 2D DSA and CE-MR imaging were compared pairwise at each location. For SSS, the diameter measured by 2D DSA was 27% larger than the conversion-based diameter from CE-MR imaging (9.8±1.4 mm vs. 7.1±1.3 mm, P<0.05). Similarly, for the right TS, the difference was 16% (8.8±3.2 mm vs. 7.4±2.0 mm, P<0.05), and for the left TS, the difference was 22% (8.4±2.8 mm vs. 6.6±1.3 mm, P<0.05). In conclusion, the diameter measured directly in conventional 2D DSA may be larger than the diameter converted based on the cross-sectional area. Therefore, when selecting the size of the stent, it is crucial to make precise determinations while keeping this fact in mind.

참고문헌 (26)

  1. Li K, "Efficacy of stenting in patients with cerebral venous sinus thrombosis-related cerebral venous sinus stenosis", J Neurointerv Surg, 11(3), pp.307~312 (2019)[DOI]
  2. Levitt MR, "Venous sinus stenting in patients without idiopathic intracranial hypertension", J Neurointerv Surg, 9(5), pp.512~515 (2017)[DOI]
  3. Fargen KM, "Recommendations for the selection and treatment of patients with idiopathic intracranial hypertension for venous sinus stenting", J Neurointerv Surg, 10(12), pp.1203~1208 (2018)[DOI]
  4. Nicholson P, "Venous sinus stenting for idiopathic intracranial hypertension : A systematic review and meta-analysis", J Neurointerv Surg, 11(4), pp.380~385 (2019)[DOI]
  5. Giridharan N, "Understanding the complex pathophysiology of idiopathic intracranial hypertension and the evolving role of venous sinus stenting : A comprehensive review of the literature", Neurosurg Focus, 45(1), pp.E10~ (2018)[DOI]
  6. Peterson KA, "Angiographic cerebral venous sinus calibers and drainage patterns in patients with normal intracranial pressure and idiopathic intracranial hypertension", J Neurointerv Surg, 13(10), pp.958~963 (2021)[DOI]
  7. Saber H, "Stent survival and stent-adjacent stenosis rates following venous sinus stenting for idiopathic intracranial hypertension : A systematic review and meta-analysis", Interv Neurol, 7(6), pp.490~500 (2018)[DOI]
  8. Lee NJ, "Comparison of high-resolution mr imaging and digital subtraction angiography for the characterization and diagnosis of intracranial artery disease", AJNR Am J Neuroradiol, 37(12), pp.2245~2250 (2016)[DOI]
  9. Loiudice P, "Choroidal vascularity index in thyroid-associated ophthalmopathy : A cross-sectional study", Eye Vis(Lond), 8(1), pp.18~ (2021)[DOI]
  10. Kim DO, "Clinical usefulness of 3-D image rendering using DICOM images in cardiovascular system", Journal of the Korean Society of Picture Archiving and Communications System, 7, pp.41~47 (2001)[DOI]
  11. 윤종태, "삼차원 뇌혈관조영술에서 테이블 높이와 확대율 조절에 따른 수정체 선량 감소에 대한 연구", 방사선기술과학, 45(4), pp.313~320 (2022), 대한방사선과학회
  12. 김승기, "뇌혈관 직경측정을 위한 3차원 회전 혈관조영술의 유용성", 방사선기술과학, 46(1), pp.9~14 (2023), 대한방사선과학회
  13. Schneiders JJ, "Intracranial aneurysm neck size overestimation with 3D rotational angiography : The impact on intra-aneurysmal hemodynamics simulated with computational fluid dynamics", AJNR Am J Neuroradiol, 34(1), pp.121~128 (2013)[DOI]
  14. Brinjikji W, "Comparison of 2D digital subtraction angiography and 3D rotational angiography in the evaluation of dome-to-neck ratio", AJNR Am J Neuroradiol, 30(4), pp.831~834 (2009)[DOI]
  15. Timaran CH, "Accuracy and utility of three-dimensional contrast-enhanced magnetic resonance angiography in planning carotid stenting", J Vasc Surg, 46(2), pp.257~263 (2007)[DOI]
  16. Lee J, "Comparison of non-invasive imaging studies in the evaluation of carotid artery stenosis and occlusion : CT angiography, time-of-flight MR angiography and contrast-enh anced MR angiography", J Korean Soc Magn Reson Med, 15, pp.234~241 (2011)[DOI]
  17. Bharatha A, "Comparison of computed tomography angiography with digital subtraction angiography in the assessment of clipped intracranial aneurysms", J Comput Assist Tomogr, 34(3), pp.440~445 (2010)[DOI]
  18. Dinkin MJ, "Venous sinus stenting in idiopathic intracranial hypertension : Results of a prospective trial", J Neuroophthalmol, 37(2), pp.113~121 (2017)[DOI]
  19. Leishangthem L, "Dural venous sinus stenting for idiopathic intracranial hypertension : An updated review", J Neuroradiol, 46(2), pp.148~154 (2019)[DOI]
  20. Boyter E, "Idiopathic intracranial hypertension", JAAPA, 32(5), pp.30~35 (2019)[DOI]
  21. Garner RM, "Quality of life, need for retreatment, and the re-equilibration phenomenon after venous sinus stenting for idiopathic intracranial hypertension", J Neurointerv Surg, 13(1), pp.79~85 (2021)[DOI]
  22. Lee KB, "Feasibility of the quantitative assessment method for CT quality control in phantom image evaluation", Appl. Sci, 11, pp.3570~ (2021)[DOI]
  23. 이기백, "CT영상의 품질관리 시 자동화된 정량적 평가방법에 관한 연구: 공간 분해능과 대조도 분해능 중심으로", 전자공학회논문지, 54(12), pp.186~194 (2017), 대한전자공학회
  24. 이재민, "고해상도 흉부 전산화단층촬영을 이용한 간질성 폐질환을 가진 환자의 자세에 따른 해부학적 구조물 크기 비교", 방사선기술과학, 44(2), pp.91~100 (2021), 대한방사선과학회
  25. Svoboda N, "Diagnostic accuracy of DSA in carotid artery stenosis : A comparison between stenosis measured on carotid endarterectomy specimens and DSA in 644 cases", Acta Neurochir(Wien), 164(12), pp.3197~3202 (2022)[DOI]
  26. 김함겸, "MR Midsagittal 영상을 이용한 정상 뇌량의 크기 측정", 방사선기술과학, 31(3), pp.277~286 (2008), 대한방사선과학회