치과용 임플란트 식립에 사용되는 핸드피스의 윤활제가 생체내 미치는 영향

Influence of Rotary Instrument Mineral Oil Lubricant on the Handpiece for Dental Implant: In Vitro and In Vivo Study

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Purpose: The purpose of this study was to evaluate the effect of oil lubricant for dental handpiece and to investigate the relationship to the prognosis of dental implants. Materials and Methods: An experiment was conducted on the residual amount of lubricant in the handpiece and the amount of lubricant that will remain in the fixture after drilling. Based on the data, biocompatibility was assessed through the assessment of cytotoxicity with L929 cells and 33 SD rats. This experiment was conducted through an 8 mm-long, 3.5 mm-diameter fixture(Osstem implant Co., Ltd., Korea) and lubricant(KaVo Dental GmbH., Germany) for handpiece. Results: The amount of lubricants in the handpiece decreased with the removal phase, but some of them could be confirmed to remain. In the cytotoxicity assessment, the survival rate was lower over 100 ppm. In experiments with SD rat, there was no significant difference among groups. Conclusion: The surface of the dental implants could be contaminated from some leftover lubricants in the handpiece. Although the lubricant is the biocompatible in cytotoxicity test and SD rat test, it does not mean it has no effect on osseointegration.

Abstract

Purpose: The purpose of this study was to evaluate the effect of oil lubricant for dental handpiece and to investigate the relationship to the prognosis of dental implants. Materials and Methods: An experiment was conducted on the residual amount of lubricant in the handpiece and the amount of lubricant that will remain in the fixture after drilling. Based on the data, biocompatibility was assessed through the assessment of cytotoxicity with L929 cells and 33 SD rats. This experiment was conducted through an 8 mm-long, 3.5 mm-diameter fixture(Osstem implant Co., Ltd., Korea) and lubricant(KaVo Dental GmbH., Germany) for handpiece. Results: The amount of lubricants in the handpiece decreased with the removal phase, but some of them could be confirmed to remain. In the cytotoxicity assessment, the survival rate was lower over 100 ppm. In experiments with SD rat, there was no significant difference among groups. Conclusion: The surface of the dental implants could be contaminated from some leftover lubricants in the handpiece. Although the lubricant is the biocompatible in cytotoxicity test and SD rat test, it does not mean it has no effect on osseointegration.

참고문헌 (35)

  1. Adell R, "A 15-year study of osseointegrated implants in the treatment of the edentulous jaw", Int J Oral Surg, 10(6), pp.387~416 (1981)
  2. Albrektsson T, "The long-term efficacy of currently used dental implants : a review and proposed criteria of success", Int J Oral Maxillofac Implants, 1(1), pp.11~25 (1986)
  3. Zhao G, "High surface energy enhances cell response to titanium substrate microstructure", J Biomed Mater Res A, 74(1), pp.49~58 (2005)
  4. Liu X, "Influence of substratum surface chemistry", energy and topography on the human fetal osteoblastic cell line hFOB 1.19: phenotypic and genotypic responses observed in vitro, 2007, pp.28(31), 4535~4550
  5. Zareidoost A, "The relationship of surface roughness and cell response of chemical surface modification of titanium", J Mater Sci Mater Med, 23(6), pp.1479~1488 (2012)
  6. Carmagnola D, "Bone tissue reaction around implants placed in a compromised jaw", J Clin Periodontol, 26(10), pp.629~635 (1999)
  7. Al-Sayyed A, "Predictable crestal bone remodelling around two porous‐coated titanium alloy dental implant designs. A radiographic study in dogs", Clin Oral Implants Res, 5(3), pp.131~141 (1994)
  8. Lundgren A, "Influence of surface roughness of barrier walls on guided bone augmentation : experimental study in rabbits", Clin Implant Dent Relat Res, 1(1), pp.41~48 (1999)
  9. Att W, "Effect of supramicron roughness characteristics produced by 1-and 2-step acid etching on the osseointegration capability of titanium", Int J Oral Maxillofac Implants, 22(5), pp.719~728 (2007)
  10. Holtorf HL, "Ectopic bone formation in rat marrow stromal cell", titanium fiber mesh scaffold constructs : effect of initial cell phenotype, 2005, pp.26(31), 6208~6216
  11. Khang D, "The role of nanometer and sub-micron surface features on vascular and bone cell adhesion on titanium", Biomaterials, 29(8), pp.970~983 (2008)
  12. Ogawa T, "Different bone integration profiles of turned and acid-etched implants associated with modulated expression of extracellular matrix genes", Int J Oral Maxillofac Implants, 18(2), pp.200~210 (2003)
  13. Wennerberg A, "On implant surfaces : a review of current knowledge and opinions", Int J Oral Maxillofac Implants, 25(1), pp.63~74 (2010)
  14. Kronström M, "Early implant failures in patients treated with Brånemark System titanium dental implants : a retrospective study", Int J Oral Maxillofac Implants, 16(2), pp.201~207 (2001)
  15. Zarb GA, "Towards optimized treatment outcomes for dental implants", The journal of prosthetic dentistry, 80(6), pp.639~640 (1998)
  16. Berglundh T, "A systematic review of the incidence of biological and technical complications in implant dentistry reported in prospective longitudinal studies of at least 5 years", J Clin Periodontol, 29, pp.197~212 (2002)
  17. Palma-Carrió C, "Risk factors associated with early failure of dental implants. A literature review", Med Oral Patol Oral Cir Bucal, 16(4), pp.e514~e517 (2011)
  18. Baqain ZH, "Early dental implant failure : risk factors", Br J Oral Maxillofac Surg, 50(3), pp.239~243 (2012)
  19. Sverzut AT, "The influence of tobacco on early dental implant failure", J Oral Maxillofac Surg, 66(5), pp.1004~1009 (2008)
  20. Friberg B, "Early failures in 4, 641 consecutively placed Brånemark dental implants : a study from stage 1 surgery to the connection of completed prostheses", Int J Oral Maxillofac Implants, 6(2), pp.142~146 (1991)
  21. Olate S, "Influence of diameter and length of implant on early dental implant failure", J Oral Maxillofac Surg, 68(2), pp.414~419 (2010)
  22. Alsaadi G, "Impact of local and systemic factors on the incidence of failures up to abutment connection with modified surface oral implants", J Clin Periodontol, 35(1), pp.51~57 (2008)
  23. Alsaadi G, "Impact of local and systemic factors on the incidence of oral implant failures, up to abutment connection", J Clin Periodontol, 34(7), pp.610~617 (2007)
  24. Leite MF, "Osseointegrated implant failure associated with MMP-1 promotor polymorphisms(-1607 and-519)", Int J Oral Maxillofac Implants, 23(4), pp.653~658 (2008)
  25. Noguerol B, "Prognostic capacity of Periotest® : retrospective study of a large sample", Clin Oral Implants Res, 17(4), pp.459~464 (2006)
  26. Van Steenberghe D, "The relative impact of local and endogenous patient-related factors on implant failure up to the abutment stage", Clin Oral Implants Res, 13(6), pp.617~622 (2002)
  27. Buser D, "Influence of surface characteristics on bone integration of titanium implants. A histomorphometric study in miniature pigs", J Biomed Mater Res, 25(7), pp.889~902 (1991)
  28. Abrahamsson I, "Early bone formation adjacent to rough and turned endosseous implant surfaces : an experimental study in the dog", Clin Oral Implants Res, 15(4), pp.381~392 (2004)
  29. Albertini M, "Advances in surfaces and osseointegration in implantology. Biomimetic surfaces", Med Oral Patol Oral Cir Bucal, 20(3), pp.e316~e25 (2015)
  30. Yuan K, "Comparison of osseointegration on various implant surfaces after bacterial contamination and cleaning : a rabbit study", Int J Oral Maxillofac Implants, 29(1), pp.32~40 (2014)
  31. Morra M, "Adherent endotoxin on dental implant surfaces : a reappraisal", J Oral Implantol, 41(1), pp.10~16 (2015)
  32. Bi Y, "Adherent endotoxin on orthopedic wear particles stimulates cytokine production and osteoclast differentiation", J Bone Miner Res, 16(11), pp.2082~2091 (2001)
  33. Greenfield EM, "Does endotoxin contribute to aseptic loosening of orthopedic implants?", J Biomed Mater Res B Appl Biomater, 72(1), pp.179~185 (2005)
  34. Bonsignore LA, "Surface contaminants inhibit osseointegration in a novel murine model", Bone, 49(5), pp.923~930 (2011)
  35. Bonsignore LA, "Machine oil inhibits the osseointegration of orthopaedic implants by impairing osteoblast attachment and spreading", J Orthop Res, 33(7), pp.979~987 (2015)

관련 논문AI추천

Effect of repeated use of an implant handpiece on an output torque: An in-vitro study김지영, 김진욱, 손영탁, 유원재, 이규복, 이재목, 손큰바다대한치과보철학회 · The Journal of Advanced of Prosthodontics2021 · 제13권 제3호Effects of various treatments for contaminated implants김수관, 문성용, 민영실, 박진성, 서동국, 양왕식, 양정은, 오지수, 유경환, 유재식, 임성철, 조지호치의학연구원 · Oral Biology Research2015 · 제39권 제2호An investigation of reosseointegration according to time course after mechanical loosening of the osseointegrated implant fixtures예선혜, 이청희, 조진현대한턱관절교합학회 · 구강회복응용과학지2015 · 제31권 제3호Dissolution behavior and early bone apposition of calcium phosphate-coated machined implants이은웅, 이인섭, 이중석, 정의원, 최성호, 황지완대한치주과학회 · Journal of Periodontal & Implant Science2013 · 제43권 제6호가토의 대퇴골에 식립한 3종류의 수산화인회석 코팅 임플란트에 대한 제거회전력 비교 연구김상수, 문지원, 박인숙, 손동석, 유석현, 이주형, 이형주대한구강악안면외과학회 · 대한구강악안면외과학회지2011 · 제37권 제1호Effects of a simplified drilling protocol at 50 rpm on heat generation under water-free conditions: an in vitro study김현주, 윤진운, 이주연, 장현지, 주지영대한치주과학회 · Journal of Periodontal & Implant Science2023 · 제53권 제1호

이 간행물의 다른 논문 (4건)