A Case of Spuriously Decreased White Blood Cell Count on an Automated Sysmex XN Hematology Analyzer: The Difference Between the WNR and WDF Channels

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Most clinical laboratories utilize automated hematology analyzers for complete blood cell count (CBC). The CBC, including white blood cells (WBCs), is critical to determine patients’ clinical status; therefore, these results should be accurate and precise. We encountered a case that showed spurious WBC count from a Sysmex XN hematology analyzer (Sysmex Corporation, Japan). A 56-year-old man visited our center in October 2018 for follow-up of hepatocellular carcinoma with lung metastases. The initial WBC count measured by the Sysmex XN was 0.02×103/μL and the differential count was not available. A blood smear slide was automatically prepared and reviewed according to the laboratory protocol. The WBC count in the blood smear was within normal limits, much higher than that of CBC result, and there was no WBC aggregation. We also reviewed the raw data from the equipment, and we found the big difference between two channels (WBC and nucleated red blood cells [WNR] and white cell differential [WDF]) of the analyzer. The WBC count of the WNR channel was very low, but that of the WDF channel was 7.16×103/μL, consistent with that of the peripheral blood smear slide. The viscosity of the blood and the neutrophil respiratory burst activity were within normal limits. In conclusion, clinical laboratories operating hematology analyzers should be alert for this phenomenon. If the analyzer shows “Difference between channels”, it is important to check the raw results of the machine, prepare a blood smear and review by microscope, and report accurate results.

Abstract

Most clinical laboratories utilize automated hematology analyzers for complete blood cell count (CBC). The CBC, including white blood cells (WBCs), is critical to determine patients’ clinical status; therefore, these results should be accurate and precise. We encountered a case that showed spurious WBC count from a Sysmex XN hematology analyzer (Sysmex Corporation, Japan). A 56-year-old man visited our center in October 2018 for follow-up of hepatocellular carcinoma with lung metastases. The initial WBC count measured by the Sysmex XN was 0.02×103/μL and the differential count was not available. A blood smear slide was automatically prepared and reviewed according to the laboratory protocol. The WBC count in the blood smear was within normal limits, much higher than that of CBC result, and there was no WBC aggregation. We also reviewed the raw data from the equipment, and we found the big difference between two channels (WBC and nucleated red blood cells [WNR] and white cell differential [WDF]) of the analyzer. The WBC count of the WNR channel was very low, but that of the WDF channel was 7.16×103/μL, consistent with that of the peripheral blood smear slide. The viscosity of the blood and the neutrophil respiratory burst activity were within normal limits. In conclusion, clinical laboratories operating hematology analyzers should be alert for this phenomenon. If the analyzer shows “Difference between channels”, it is important to check the raw results of the machine, prepare a blood smear and review by microscope, and report accurate results.

참고문헌 (6)

  1. Zandecki M, "Spurious counts and spurious results on haematology analysers : a review. Part II : white blood cells, red blood cells, haemoglobin, red cell indices and reticulocytes", Int J Lab Hematol, 29, pp.21~41 (2007)
  2. Zandecki M, "Spurious counts and spurious results on haematology analysers : a review. Part I : platelets", Int J Lab Hematol, 29, pp.4~20 (2007)
  3. Kim H, "Comparison of white blood cell counts by WNR, WDF, and WPC channels in Sysmex XN hematology analyzer", Int J Lab Hematol, 37, pp.869~875 (2015)
  4. Nguyen VT, "Spurious decrease in the WBC count measured by the WNR channel of XN haematology analyser(Sysmex)could be associated with metastatic adenocarcinoma", Int J Lab Hematol, 37, pp.e129~e132 (2015)
  5. Tantanate C, "Spurious white blood cell count from a new automated Sysmex XN hematology analyzer", Int J Lab Hematol, 36, pp.e86~e87 (2014)
  6. de Jonge R, "Automated counting of white blood cells in synovial fluid", Rheumatology (Oxford), 43, pp.170~173 (2004)

관련 논문AI추천

자동혈구분석기 Sysmex XN-2000의 성능 및 Low WBC Mode를 이용한 백혈구 감별계산 정확성 평가김혜란, 송새암, 신정환, 오승환, 이자영, 이정녀, 전경란대한진단검사의학회 · Laboratory Medicine Online2015 · 제5권 제4호Correction of Pseudoreticulocytosis in Leukocytosis Samples Using the Sysmex XE-2100 Analyzer Depends on the Type and Number of White Blood Cells김명신, 김아현, 김용구, 박연준, 박준홍, 오은지, 임지향, 한경자대한진단검사의학회 · Annals of Laboratory Medicine2012 · 제32권 제6호Unreliable Automated Complete Blood Count Results: Causes, Recognition, and ResolutionGong Jerald, Gulati Gene, Uppal Guldeep대한진단검사의학회 · Annals of Laboratory Medicine2022 · 제42권 제5호Performance Evaluation of the Mindray BC-6200 Hematology Analyzer; Comparison with Sysmex XE-2100 and Manual Microscopy김한아, 박미경, 이창훈, 이태환, 허미나대한진단검사의학회 · Laboratory Medicine Online2022 · 제12권 제4호A Clue to Discovering Unstable Hemoglobin Variants via Abnormal WBC Differential Scattergrams Using the Sysmex Automated Hematology Analyzer박설희, 정태동, 허정원, 홍기숙대한진단검사의학회 · Laboratory Medicine Online2019 · 제9권 제2호Validation of Hematology Rapid Reporting System for Complete Blood Cell Count with Differential김지용, 박지영, 윤수영, 이정민대한진단검사의학회 · Laboratory Medicine Online2022 · 제12권 제1호

이 간행물의 다른 논문 (10건)