A Clue to Discovering Unstable Hemoglobin Variants via Abnormal WBC Differential Scattergrams Using the Sysmex Automated Hematology Analyzer

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Incidentally, hemoglobin (Hb) variants can be detected using HbA1c tests in clinical laboratories. We found 38 patients with Hb variants after reviewing a total of 29,398 HbA1c test results from January 2017 to December 2017. While reviewing the complete blood count results of the patients (N=36) using the Sysmex XN-9000 analyzer (Sysmex, Japan), 35 patients were flagged as unremarkable with respect to differential white blood cell (WBC) counts. However, 1 patient with a normal WBC count did not obtain a differential WBC count while being flagged for an abnormal WBC scattergram in the white blood cell differential (WDF) channel. The WBC histogram showed an abnormally low fluorescent signal in the WDF channel; however, the differential WBC count was normal upon microscopic examination. After testing the patient's buffy coat suspended in normal saline and removing red blood cells (RBCs), the WBC scattergram and differential WBC count returned to normal. This finding suggests that the presence of a patient's RBCs may affect WBC scattergrams and Hb variants may interfere with the fluorescent dye in the differential WBC count. Therefore, when an abnormal WBC scattergram with an abnormally low fluorescent signal is encountered on the Sysmex XN-9000 analyzer, the presence of an Hb variant can be suspected.

Abstract

Incidentally, hemoglobin (Hb) variants can be detected using HbA1c tests in clinical laboratories. We found 38 patients with Hb variants after reviewing a total of 29,398 HbA1c test results from January 2017 to December 2017. While reviewing the complete blood count results of the patients (N=36) using the Sysmex XN-9000 analyzer (Sysmex, Japan), 35 patients were flagged as unremarkable with respect to differential white blood cell (WBC) counts. However, 1 patient with a normal WBC count did not obtain a differential WBC count while being flagged for an abnormal WBC scattergram in the white blood cell differential (WDF) channel. The WBC histogram showed an abnormally low fluorescent signal in the WDF channel; however, the differential WBC count was normal upon microscopic examination. After testing the patient's buffy coat suspended in normal saline and removing red blood cells (RBCs), the WBC scattergram and differential WBC count returned to normal. This finding suggests that the presence of a patient's RBCs may affect WBC scattergrams and Hb variants may interfere with the fluorescent dye in the differential WBC count. Therefore, when an abnormal WBC scattergram with an abnormally low fluorescent signal is encountered on the Sysmex XN-9000 analyzer, the presence of an Hb variant can be suspected.

참고문헌 (6)

  1. Thom CS, "Hemoglobin variants : biochemical properties and clinical correlates", Cold Spring Harb Perspect Med, 3, pp.a011858~ (2013)
  2. 박설희, "Diagnostic Utility of Serum Glycated Albumin for Diabetes Mellitus and Its Correlation With Hyperlipidemia", Annals of Laboratory Medicine, 36(4), pp.306~312 (2016), 대한진단검사의학회
  3. Rosetti M, "A rare case of hemoglobin Leiden interfering with the DIFF channel of Sysmex XE-2100", Scand J Clin Lab Invest, 75, pp.436~437 (2015)
  4. Rosetti M, "Serendipitous detection of hemoglobin G-Ferrara variant with Sysmex DIFF channel", Clin Biochem, 49, pp.192~193 (2016)
  5. Mongelli F, "A rare and unstable hemoglobin variant, Hb M Dothan β 25", 26(-GTG), detected by the anomalous cytogram on Sysmex XE-2100, 2016, pp.54, e31~e33
  6. Teixeira C, "Automated detection of unstable hemoglobin variants by Sysmex XE-Series analyzers", Clin Chem Lab Med, 55, pp.e243~e246 (2017)

관련 논문AI추천

A Case of Spuriously Decreased White Blood Cell Count on an Automated Sysmex XN Hematology Analyzer: The Difference Between the WNR and WDF Channels김한주, 남궁승, 박찬정대한진단검사의학회 · Laboratory Medicine Online2021 · 제11권 제3호Hemoglobin Yamagata: Hemoglobin Variant Detected by HbA1c Test권경주, 기창석, 성연아, 이미애, 이승태, 정채림, 정화순, 홍기숙대한진단검사의학회 · Annals of Laboratory Medicine2009 · 제29권 제6호Unreliable Automated Complete Blood Count Results: Causes, Recognition, and ResolutionGong Jerald, Gulati Gene, Uppal Guldeep대한진단검사의학회 · Annals of Laboratory Medicine2022 · 제42권 제5호Correction of Pseudoreticulocytosis in Leukocytosis Samples Using the Sysmex XE-2100 Analyzer Depends on the Type and Number of White Blood Cells김명신, 김아현, 김용구, 박연준, 박준홍, 오은지, 임지향, 한경자대한진단검사의학회 · Annals of Laboratory Medicine2012 · 제32권 제6호The First Study on Nucleotide-level Identification of Hb Koriyama in a Patient with Severe Hemolytic Anemia박성섭, 박승만, 박준은, 성문우, 전용범, 조성임대한진단검사의학회 · Annals of Laboratory Medicine2012 · 제32권 제1호Comparison of Capillary Electrophoresis with Cellulose Acetate Electrophoresis for the Screening of Hemoglobinopathies김보람, 김정만, 김지은, 박주인, 우광숙, 한진영대한진단검사의학회 · Annals of Laboratory Medicine2011 · 제31권 제4호

이 간행물의 다른 논문 (11건)