이어폰 및 헤드폰의 정자기장이 인공심장 박동기및 이식형 제세동기에 미치는 영향

Effects of Static Magnetic Fields of Earphones and Headphones on Pacemakers and Implantable Cardioverter Defibrillators

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

In this study, we evaluated the effects of static magnetic fields of earphones and headphones on pacemakersand implantable cardioverter defibrillators(ICDs). Five pacemakers and three ICDs were subjected to in-vitrotest with three headphones which were an in-ear earphone, clip-on headphone, and closed-back headphone. Eachimplantable device was placed in close proximity(within 3 mm) to the ear-pad of each of the earphone and headphonesfor 3 min. As a result, no effects were observed on the pacemakers for the earphone and headphones during the test,but an effect was observed on one ICD for the clip-on and closed-back headphone during the test. When the ICDwas placed in close proximity to the headphones, the ICD temporarily suspended functions of tachyarrhythmia detectionand therapy. The effect was not observed in this study when the headphones were at least 2 cm from the ICD. Based on these findings, patients with ICDs should be advised to keep earphones and headphones at least 2 cm apartfrom their ICDs.

Abstract

In this study, we evaluated the effects of static magnetic fields of earphones and headphones on pacemakersand implantable cardioverter defibrillators(ICDs). Five pacemakers and three ICDs were subjected to in-vitrotest with three headphones which were an in-ear earphone, clip-on headphone, and closed-back headphone. Eachimplantable device was placed in close proximity(within 3 mm) to the ear-pad of each of the earphone and headphonesfor 3 min. As a result, no effects were observed on the pacemakers for the earphone and headphones during the test,but an effect was observed on one ICD for the clip-on and closed-back headphone during the test. When the ICDwas placed in close proximity to the headphones, the ICD temporarily suspended functions of tachyarrhythmia detectionand therapy. The effect was not observed in this study when the headphones were at least 2 cm from the ICD. Based on these findings, patients with ICDs should be advised to keep earphones and headphones at least 2 cm apartfrom their ICDs.

참고문헌 (20)

  1. A. J. Greenspon, "16-year trends in the infection burden for pacemakers and implantable cardioverter-defibrillators in the United States 1993 to 2008", J. Am. Coll. Cardiol., 58(10), pp.1001~1006 (2011)
  2. K. Dyrda, "Implantable rhythm devices and electromagnetic interference: myth or reality?", Expert Rev. Cardiovasc. Ther., 6(6), pp.823~832 (2008)
  3. R. Beinart, "Intermittent, erratic behaviour of an implantable cardioverter defibrillator secondary to a hidden magnetic source of interference", Europace, 13(10), pp.1508~1509 (2011)
  4. 홍하나, "설문지를 통한 고등학생들의 MP3 사용 실태 및 청력에 미치는 영향", 한국음향학회지, 32(1), pp.56~63 (2013), 한국음향학회[DOI]
  5. Korea Communications Commission Annual Report 2013, "Korea Communications Commission", ~ (2014)
  6. S. Ryf, "Interference of neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators: an in vitro study", Technol. Health Care, 16(1), pp.13~18 (2008)
  7. T. Wolber, "Potential interference of small neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators", Heart Rhythm, 4(1), pp.1~4 (2007)
  8. E. H. Lucas, "The effects of electronic article surveillance systems on permanent cardiac pacemakers: an in vitro study", Pacing Clin. Electrophysiol, 17(11), pp.2021~2026 (1994)
  9. A. D. Bernstein, "The revised NASPE", BPEG generic code for antibradycardia, adaptive-rate, and multisite pacing. North American Society of Pacing and Electrophysiology, Pacing Clin. Electrophysiol., pp.2002, 25(2)[DOI]
  10. M. E. McIvor, "Study of pacemaker and implantable cardioverter defibrillator triggering by electronic article surveillance devices(SPICED TEAS)", Pacing Clin. Electrophysiol, 21(10), pp.1847~1861 (1998)
  11. W. J. Groh, "Interactions between electronic article surveillance systems and implantable cardioverter-defibrillators", Circulation, 100(4), pp.387~392 (1999)
  12. "Implantable pacemaker and defibrillator information: Magnets" (2006)[URL]
  13. "Magnet Use For Suspending Medtronic ICD Detection" (2014)[URL]
  14. 김남호, "특집", 심장 돌연사, 2010, pp.53(3), 214~227[DOI]
  15. "Magnet Use with Medtronic Pacemakers" (2014)[URL]
  16. M. Bilitch, "Ventricular fibrillation and competitive pacing", N. Engl. J. Med, 276(11), pp.598~604 (1967)
  17. S. Lee, "Clinically significant magnetic interference of implanted cardiac devices by portable headphones", Heart Rhythm, 6(10), pp.1432~1436 (2009)
  18. R. Luechinger, "Pacemaker reed switch behavior in 0.5, 1.5, and 3.0 Tesla magnetic resonance imaging units: are reed switches always closed in strong magnetic fields?", Pacing Clin. Electrophysiol., 25(10), pp.1419~1423 (2002)
  19. R. Beinart, "Effects of External Electrical and Magnetic Fields on Pacemakers and Defibrillators From Engineering Principles to Clinical Practice", Circulation, 128(25), pp.2799~2809 (2013)
  20. "Electromagnetic Interference(EMI)" (2013)[URL]

관련 논문AI추천

이 간행물의 다른 논문 (4건)