Effective and Practical Complete Blood Count Delta Check Method and Criteria for the Quality Control of Automated Hematology Analyzers
1
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Background: Delta checks increase patient safety by identifying automated hematology analyzer errors. International standards and guidelines for the complete blood count (CBC) delta check method have not been established. We established an effective, practical CBC delta check method and criteria. Methods: We assessed five delta check methods for nine CBC items (Hb, mean corpuscular volume, platelet count, white blood cell [WBC] count, and five-part WBC differential counts) using 219,804 blood samples from outpatients and inpatients collected over nine months. We adopted the best method and criteria and evaluated them using 42,652 CBC samples collected over two weeks with a new workflow algorithm for identifying test errors and corrections for Hb and platelet count. Results: The median delta check time interval was 1 and 21 days for inpatients and outpatients (range, 1–20 and 1–222 days), respectively. We used delta values at 99.5% as delta check criteria; the criteria varied among the five methods and between outpatients and inpatients. The delta percent change (DPC)/reference range (RR) rate performed best as the delta check for CBC items. Using the new DPC/RR rate method, 1.7% of total test results exceeded the delta check criteria; the retesting and resampling rates were 0.5% and 0.001%, respectively. Conclusions: We developed an effective, practical delta check method, including RRs and delta check time intervals, and delta check criteria for nine CBC items. The criteria differ between outpatients and inpatients. Using the new workflow algorithm, we can identify the causes of criterion exceedance and report correct test results.
Abstract
Background: Delta checks increase patient safety by identifying automated hematology analyzer errors. International standards and guidelines for the complete blood count (CBC) delta check method have not been established. We established an effective, practical CBC delta check method and criteria. Methods: We assessed five delta check methods for nine CBC items (Hb, mean corpuscular volume, platelet count, white blood cell [WBC] count, and five-part WBC differential counts) using 219,804 blood samples from outpatients and inpatients collected over nine months. We adopted the best method and criteria and evaluated them using 42,652 CBC samples collected over two weeks with a new workflow algorithm for identifying test errors and corrections for Hb and platelet count. Results: The median delta check time interval was 1 and 21 days for inpatients and outpatients (range, 1–20 and 1–222 days), respectively. We used delta values at 99.5% as delta check criteria; the criteria varied among the five methods and between outpatients and inpatients. The delta percent change (DPC)/reference range (RR) rate performed best as the delta check for CBC items. Using the new DPC/RR rate method, 1.7% of total test results exceeded the delta check criteria; the retesting and resampling rates were 0.5% and 0.001%, respectively. Conclusions: We developed an effective, practical delta check method, including RRs and delta check time intervals, and delta check criteria for nine CBC items. The criteria differ between outpatients and inpatients. Using the new workflow algorithm, we can identify the causes of criterion exceedance and report correct test results.
참고문헌 (27)
- Howanitz PJ, "Quality assurance measurements in departments of pathology and laboratory medicine", Arch Pathol Lab Med, 114, pp.1131~1135 (1990)
- Lindberg DA, "Collection, evaluation, and transmission of hospital laboratory data", Methods Inf Med, 6, pp.97~107 (1967)
- Klee CG SP, "Advanced laboratory methods in haematology", pp.3~17 (2001), Taylor & Francis
- Fraser CG, "Generation and application of data on biological variation in clinical chemistry", Crit Rev Clin Lab Sci, 27, pp.409~437 (1989)
- Petersen PH, "The index of individuality is often a misinterpreted quantity characteristic", Clin Chem Lab Med, 37, pp.655~661 (1999)
- 박상혁, "New Decision Criteria for Selecting Delta Check Methods Based on the Ratio of the Delta Difference to the Width of the Reference Range Can Be Generally Applicable for Each Clinical Chemistry Test Item", Annals of Laboratory Medicine, 32(5), pp.345~354 (2012), 대한진단검사의학회
- Lee J, "Usefulness of biological variation in the establishment of delta check limits", Clin Chim Acta, 463, pp.18~21 (2016)
- Hong J, "Application and optimization of reference change values for delta checks in clinical laboratory", J Clin Lab Anal, 34, pp.e23550~ (2020)
- Kim JW, "Differential application of rate and delta check on selected clinical chemistry tests", J Korean Med Sci, 5, pp.189~195 (1990)
- Schifman RB, "Delta check practices and outcomes : a Q-probes study involving 49 health care facilities and 6541 delta check alerts", Arch Pathol Lab Med, 141, pp.813~823 (2017)
- Randell EW, "Delta checks in the clinical laboratory", Crit Rev Clin Lab Sci, 56, pp.75~97 (2019)
- Tan RZ, "An approach to optimize delta checks in test panels-the effect of the number of rules included", Ann Clin Biochem, 57, pp.215~222 (2020)
- Ko DH, "Utility of reference change values for delta check limits", Am J Clin Pathol, 148, pp.323~329 (2017)
- Rosenbaum MW, "Using machine learning-based multianalyte delta checks to detect wrong blood in tube errors", Am J Clin Pathol, 150, pp.555~566 (2018)
- Zhou R, "A highly accurate delta check method using deep learning for detection of sample mix-up in the clinical laboratory", Clin Chem Lab Med, 60, pp.1984~1992 (2021)
- Fu Y, "Clinical validation of a delta check model in haematology automated counting improves data validation", Int J Lab Hematol, 42, pp.77~81 (2020)
- Miller I, "Development and evaluation of a logical delta check for identifying erroneous blood count results in a tertiary care hospital", Arch Pathol Lab Med, 139, pp.1042~1047 (2015)
- Park HS, "Computerized processing of test results from Hematologic AutoAnalyzer", Korean J Clin Pathol, 9, pp.49~58 (1989)
- Yang YS, "Experience with abnormal CBC histogram and delta and panic check", Lab Med Qual Assur, 13, pp.201~204 (1991)
- 구본경, "변화치 및 경고치 검색과 Q-flag limit를 이용한 혈구산정검사의 검사소요시간", 대한임상검사과학회지, 44(2), pp.66~74 (2012), 대한임상검사과학회
- Quintó L, "Relationship between haemoglobin and haematocrit in the definition of anaemia", Trop Med Int Health, 11, pp.1295~1302 (2006)
- Helms CC, "Mechanisms of hemolysis-associated platelet activation", J Thromb Haemost, 11, pp.2148~2154 (2013)
- Gulati Gene, "Unreliable Automated Complete Blood Count Results: Causes, Recognition, and Resolution", Annals of Laboratory Medicine, 42(5), pp.515~530 (2022), 대한진단검사의학회[DOI]
- Lippi G, "Quality in laboratory diagnostics: from theory to practice", Biochem Med(Zagreb), 20, pp.126~130 (2010)
- Plebani M, "Mistakes in a stat laboratory : types and frequency", Clin Chem, 43, pp.1348~1351 (1997)
- Carraro P, "Errors in a stat laboratory : types and frequencies 10 years later", Clin Chem, 53, pp.1338~1342 (2007)
- He S, "National survey on delta checks in clinical laboratories in China", Clin Chem Lab Med, 58, pp.569~576 (2020)