본문 바로가기
KERD
논문KCI 등재

제세동 패들에 가해지는 압력의 모니터링 효과

The effects of monitoring of the pressures applied on the defibrillator paddles- A manikin study -

0

조회

0

다운로드

1

인용

0

참고문헌

키워드

초록

Purpose : The purpose of this study was to determine the changes that occur due to the real-time monitoring of paddle pressures which has an important influence on the defibrillation success rate in defibrillation treatment known as the only treatment for cardiac arrest patients with VF. Methods : 40 people participated in the cardiac arrest simulation training and played the role of the defibrillation operator. Investigators measured the pressure of paddle while defibrillating by using instrument which was developed by the investigator. Results : Through real-time monitoring of the paddle pressures of defibrillator by indicator, the front sternum paddle showed a 77.5% success rate and the apex paddle showed a 40% success rate. While the values without monitoring the paddle pressures, the front sternum paddle showed a 51% success rate and the apex paddle showed a 20% success rate. These experiment revealed statistically significant(p<.001) low success rate. Conclusion : The method of monitoring the paddle pressures during defibrillation showed that the paddle can be precisely gripped. The success rate of paddle pressures is significantly correlated with height, weight and grip strength.

Abstract

Purpose : The purpose of this study was to determine the changes that occur due to the real-time monitoring of paddle pressures which has an important influence on the defibrillation success rate in defibrillation treatment known as the only treatment for cardiac arrest patients with VF. Methods : 40 people participated in the cardiac arrest simulation training and played the role of the defibrillation operator. Investigators measured the pressure of paddle while defibrillating by using instrument which was developed by the investigator. Results : Through real-time monitoring of the paddle pressures of defibrillator by indicator, the front sternum paddle showed a 77.5% success rate and the apex paddle showed a 40% success rate. While the values without monitoring the paddle pressures, the front sternum paddle showed a 51% success rate and the apex paddle showed a 20% success rate. These experiment revealed statistically significant(p<.001) low success rate. Conclusion : The method of monitoring the paddle pressures during defibrillation showed that the paddle can be precisely gripped. The success rate of paddle pressures is significantly correlated with height, weight and grip strength.

참고문헌 (15)

  1. Rea TD, "Increasing use of cardiopulmonary resuscitation during out-of-hospital ventricular fibrillation arrest: survival implications of guideline changes", Circulation, 114(25), pp.2760~2765 (2006)
  2. Hwang SO, "Cardiopulmonary resuscitation and advanced cardiac life support", pp.203~ (2001), Koonja
  3. Steinmetz J, "Improved survival after an out-of-hospital cardiac arrest using new guidelines", Acta Anaesthesiol Scand, 52(7), pp.908~913 (2008)
  4. Hinchey PR, "Improved out-of-hospital cardiac arrest survival after the sequential implementation of 2005 AHA guidelines for compressions, ventilations, and induced hypothermia: the Wake County experience", Ann Emerg Med, 56(4), pp.348~357 (2010)
  5. Kerber RE, "Transthoracic resistance in human defibrillation. Influence of body weight, chest size, serial shocks, paddle size and paddle contact pressure", Circulation, 63(3), pp.676~682 (1981)
  6. Sado DM, "Comparison of the effects of removal of chest hair with not doing so before external defibrillation on transthoracic impedance", Am J Cardiol, 93(1), pp.98~100 (2004)
  7. Bissing JW, "Effect of shaving the chest of hirsute subjects on transthoracic impedance to self-adhesive defibrillation electrode pads", Am J Cardiol, 86(5), pp.587~589 (2000)
  8. Kerber RE, "Advance prediction of transthoracic impedance in human defibrillation and cardioversion: importance of impedance in determining the success of low-energy shocks", Circulation, 70(2), pp.303~308 (1984)
  9. Dalzell GW, "Electrode pad size, transthoracic impedance of external ventricular defibrillation", Am J cardiol, 64(12), pp.741~744 (1989)
  10. Deakin CD, "Determining the optimal paddle force for external defibrillation", Am J Cardiol, 90(7), pp.812~813 (2002)
  11. "Anonymous. Part 6: advanced cardiovascular life support. Section 2: defibrillation. European Resuscitation Council", Resuscitation, 46(1-3), pp.109~113 (2000)
  12. Geddes LA, "The decrease in transthoracic impedance during successive ventricular defibrillation trials", Med Instrum, 9(4), pp.179~180 (1975)
  13. Geddes LA, "The prediction of the impedance of the thorax to defibrillating current", Med Instrum, 10(3), pp.159~162 (1976)
  14. Sado DM, "Are European Resuscitation Council recommendations for paddle force achievable during defibrillation?", Resuscitation, 51(3), pp.287~290 (2001)
  15. 이명훈, "우리나라 응급의료센터의 제세동시 패들-피부 결합물질 사용현황 및 결합물질의 임피던스 측정", 대한응급의학회지, 21(1), pp.125~130 (2010), 대한응급의학회